Calibrating capillary gas flow meters by means of the transient method

被引:2
作者
Kudasik, Mateusz [1 ]
机构
[1] Polish Acad Sci, Strata Mech Research Inst, Reymonta 27, PL-30059 Krakow, Poland
关键词
gas flow meter; capillary tube; calibration; CO2/CH4 EXCHANGE SORPTION; THERMAL FLOW; FLOWMETER;
D O I
10.1088/1361-6501/ab3dcb
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The present article discusses the transient method of calibrating capillary gas flow meters which involves the application of a type of manometric flow meters, which measure the release of a substance by registering changes in the differential pressure on a measuring element. The measurement methodology presented in the article is but a concept proposing actions and calculations aimed at determining the calibration coefficient of a section of a measurement capillary in an outstandingly simple manner. This method is dedicated to capillary flow meters with very small gas flow rates. Calibrating flow meters by means of the transient method involves registering changes in the differential pressure Delta p(t) on both ends of a capillary tube while gas flows out of a measuring burette, through a calibrated capillary tube, to be released into the atmosphere. On the basis of a combination of the Hagen-Poiseuille equation and the ideal gas law, a relationship describing the phenomenon of a gas flow through a capillary tube was formulated, which, in turn, served as a basis for determining the calibration coefficients of the two capillary sections measuring 10 cm and 40 cm, with an internal radius of 0.015 cm. The calibration processes of both capillary sections were conducted for 21 different temperature values, using methane. The obtained values of calibration coefficients were juxtaposed with their model values obtained via calculation. The relative differences between the calibration-obtained values and model values did not exceed 7.65% and 4.83% in the case of the 10 cm capillary tube and the 40 cm capillary tube, respectively.
引用
收藏
页数:12
相关论文
共 27 条
  • [1] [Anonymous], 1998, PVT PHASE BEHAV PETR
  • [2] Flowmeter for very low gas flows of inert gases
    Avdiaj, Sefer
    Setina, Janez
    [J]. MEASUREMENT, 2017, 111 : 384 - 388
  • [3] Micro-flow facility for traceability in steady and pulsating flow
    Bissig, Hugo
    Tschannen, Martin
    de Huu, Marc
    [J]. FLOW MEASUREMENT AND INSTRUMENTATION, 2015, 44 : 34 - 42
  • [4] MEASUREMENT OF SMALL VALUES OF HYDROSTATIC PRESSURE DIFFERENCE
    Broda, Krzysztof
    Filipek, Wiktor
    [J]. ARCHIVES OF MINING SCIENCES, 2012, 57 (01) : 157 - 167
  • [5] Study on the measurement accuracy of an improved cemented carbide orifice flowmeter in natural gas pipeline
    Dong, Jingya
    Jing, Chengjun
    Peng, Yang
    Liu, Yang
    Ren, Hongyu
    Liu, Xianxu
    [J]. FLOW MEASUREMENT AND INSTRUMENTATION, 2018, 59 : 52 - 62
  • [6] Balance of CO2/CH4 exchange sorption in a coal briquette
    Dutka, Barbara
    Kudasik, Mateusz
    Pokryszka, Zbigniew
    Skoczylas, Norbert
    Topolnicki, Juliusz
    Wierzbicki, Miroslaw
    [J]. FUEL PROCESSING TECHNOLOGY, 2013, 106 : 95 - 101
  • [7] Pore pressure changes accompanying exchange sorption of CO2/CH4 in a coal briquette
    Dutka, Barbara
    Kudasik, Mateusz
    Topolnicki, Juliusz
    [J]. FUEL PROCESSING TECHNOLOGY, 2012, 100 : 30 - 34
  • [8] Natural gas viscosity estimation through corresponding states based models
    Heidaryan, Ehsan
    Esmaeilzadeh, Feridun
    Moghadasi, Jamshid
    [J]. FLUID PHASE EQUILIBRIA, 2013, 354 : 80 - 88
  • [9] Henderson MA, 2017, ANAEST INTENS CARE M, V18, P180, DOI 10.1016/j.mpaic.2017.01.009
  • [10] A simple correlation to estimate natural gas viscosity
    Jarrahian, Azad
    Heidaryan, Ehsan
    [J]. JOURNAL OF NATURAL GAS SCIENCE AND ENGINEERING, 2014, 20 : 50 - 57