Personalized 3D-printed anthropomorphic phantoms for dosimetry in charged particle fields

被引:6
作者
Halloran, Andrew [1 ,3 ]
Newhauser, Wayne [1 ,2 ]
Chu, Connel [2 ]
Donahue, William [1 ,4 ]
机构
[1] Louisiana State Univ, Dept Phys & Astron, Baton Rouge, LA 70803 USA
[2] Mary Bird Perkins Canc Ctr, Dept Radiat Phys, Baton Rouge, LA 70809 USA
[3] Univ Colorado, Environm Hlth & Safety, Anschutz Med Campus, Denver, CO 80202 USA
[4] Yale Sch Med, Dept Therapeut Radiol, New Haven, CT USA
关键词
anthropomorphic phantom; 3D printing; dosimetry; electron beam; 3D; FABRICATION; DESIGN; ACCURACY;
D O I
10.1088/1361-6560/ac3047
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Anthropomorphic phantoms used for radiation dose measurements are designed to mimic human tissue in shape, size, and tissue composition. Reference phantoms are widely available and are sufficiently similar to many, but not all, human subjects. 3D printing has the potential to overcome some of these shortcomings by enabling rapid fabrication of personalized phantoms for individual human subjects based on radiographic imaging data. Objective. The objective of this study was to test the efficacy of personalized 3D printed phantoms for charged particle therapy. To accomplish this, we measured dose distributions from 6 to 20 MeV electron beams, incident on printed and molded slices of phantoms. Approach. Specifically, we determined the radiological properties of 3D printed phantoms, including beam penetration range. Additionally, we designed and printed a personalized head phantom to compare results obtained with a commercial, reference head phantom for quality assurance of therapeutic electron beam dose calculations. Main Results. For regions of soft tissue, gamma index analyses revealed a 3D printed slice was able to adequately model the same electron beam penetration ranges as the molded reference slice. The printed, personalized phantom provided superior dosimetric accuracy compared to the molded reference phantom for electron beam dose calculations at all electron beam energies. However, current limitations in the ability to print high-density structures, such as bone, limited pass rates of 60% or better at 16 and 20 MeV electron beam energies. Significance. This study showed that creating personalized phantoms using 3D printing techniques is a feasible way to substantially improve the accuracy of dose measurements of therapeutic electron beams, but further improvements in printing techniques are necessary in order to increase the printable density in phantoms.
引用
收藏
页数:9
相关论文
共 23 条
[1]  
Avelino SR, 2012, IEEE ENG MED BIO, P5718, DOI 10.1109/EMBC.2012.6347293
[2]   Investigating the accuracy of microstereotactic-body-radiotherapy utilizing anatomically accurate 3D printed rodent-morphic dosimeters [J].
Bache, Steven T. ;
Juang, Titania ;
Belley, Matthew D. ;
Koontz, Bridget F. ;
Adamovics, John ;
Yoshizumi, Terry T. ;
Kirsch, David G. ;
Oldham, Mark .
MEDICAL PHYSICS, 2015, 42 (02) :846-855
[3]   Preparation and fabrication of a full-scale, sagittal-sliced, 3D-printed, patient-specific radiotherapy phantom [J].
Craft, Daniel F. ;
Howell, Rebecca M. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05) :285-292
[4]  
Dewerd L.A., 2014, PHANTOMS MED HLTH PH, DOI DOI 10.1007/978-1-4614-8304-5
[5]   Patient specific 3D printed phantom for IMRT quality assurance [J].
Ehler, Eric D. ;
Barney, Brett M. ;
Higgins, Patrick D. ;
Dusenbery, Kathryn E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (19) :5763-5773
[6]   Recent advances on the development of phantoms using 3D printing for imaging with CT, MRI, PET, SPECT, and ultrasound [J].
Filippou, Valeria ;
Tsoumpas, Charalampos .
MEDICAL PHYSICS, 2018, 45 (09) :E740-E760
[7]   Development of patient-specific molecular imaging phantoms using a 3D printer [J].
Gear, J. I. ;
Long, C. ;
Rushforth, D. ;
Chittenden, S. J. ;
Cummings, C. ;
Flux, G. D. .
MEDICAL PHYSICS, 2014, 41 (08) :525-527
[8]  
Halloran AM., 2015, DOSIMETRIC ADVANTAGE
[9]   Design of a head phantom produced on a 3D rapid prototyping printer and comparison with a RANDO and 3M lucite head phantom in eye dosimetry applications [J].
Homolka, Peter ;
Figl, Michael ;
Wartak, Andreas ;
Glanzer, Mathias ;
Duenkelmeyer, Martina ;
Hojreh, Azadeh ;
Hummel, Johann .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08) :3158-3174
[10]   New Technique for Developing a Proton Range Compensator With Use of a 3-Dimensional Printer [J].
Ju, Sang Gyu ;
Kim, Min Kyu ;
Hong, Chae-Seon ;
Kim, Jin Sung ;
Han, Youngyih ;
Choi, Doo Ho ;
Shin, Dongho ;
Lee, Se Byeong .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 88 (02) :453-458