Comparison of two X-ray detection systems used to investigate properties of normal and malignant breast tissues

被引:5
作者
Alaroui, D. F. [1 ]
Johnston, E. M. [1 ]
Farquharson, M. J. [2 ]
机构
[1] McMaster Univ, Radiat Sci Grad Program, 1280 Main St W, Hamilton, ON L8S 4K1, Canada
[2] McMaster Univ, Sch Interdisciplinary Sci, 1280 Main St W, Hamilton, ON L8S 4K1, Canada
关键词
Breast cancer; X-ray fluorescence; X-ray diffraction; Trace elements; Structural components; Breast margin assessment; FROZEN-SECTION ANALYSIS; TRACE-ELEMENTS; SYNCHROTRON-RADIATION; CONSERVING SURGERY; HEALTHY TISSUES; DIFFRACTION; SCATTERING; CANCER; FLUORESCENCE; ZN;
D O I
10.1016/j.nima.2019.02.027
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The present study was implemented using two X-ray detection systems; a monochromatic X-ray system for X-ray Fluorescence (XRF) and Angular Dispersive X-ray Diffraction (ADXRD) techniques, and a combined Polarized Energy Dispersive X-ray Fluorescence (PEDXRF) and Energy Dispersive X-ray Diffraction (EDXRD) system. As both of these systems involve different techniques, the primary objective of this study was to evaluate the performance and accuracy of each system using results achieved from XRF measurements. The assessment of the two systems was carried out by investigating tissues samples containing invasive ductal carcinoma (IDC) of the breast and normal surrounding breast tissues. The results established from the two XRF systems are in a very good agreement with each other. The statistical analysis reveals a significant and measurable increase (p < 0.01) in the concentration of K, Ca, Zn, Rb and Fe (p < 0.05) in the tumorous tissue when compared with the healthy tissue. However, the levels of Cl, Cu and Br attained by both systems have not demonstrated a statistically significant difference between the normal and cancerous tissues. Investigating the structural components of the same breast tissues using each of the X-ray Diffraction (XRD) spectrometers incorporated in both systems indicated a statistically significant difference in the components of normal and malignant samples. Furthermore, the results have shown a significant increase in the fibrous and water contents of the tumour tissue at p < 0.01, and a significant increase in the adipose content of the normal tissue at (p < 0.01). The results acquired from both XRD approaches were shown to be statistically compatible with each other. Overall, the comparisons between the two X-ray detection systems have shown improved results for the combined PEDXRF and EDXRD system for the purpose of classifying normal and tumour breast tissues.
引用
收藏
页数:9
相关论文
共 49 条
  • [1] Quality management and method Validation in EDXRF analysis
    Alvarez, R. Padilla
    Markowicz, A.
    Wegrzynek, D.
    Cano, E. Chinea
    Bamford, S. A.
    Torres, D. Hernandez
    [J]. X-RAY SPECTROMETRY, 2007, 36 (01) : 27 - 34
  • [2] Characterizing fundamental parameter-based analysis for soil-ceramic matrices in polarized energy-dispersive X-ray fluorescence (PEDXRF) spectrometry
    Amin Abuhani, Waleed
    Dasgupta-Schubert, Nabanita
    Villasenor Cendejas, Luis Manuel
    [J]. POWDER DIFFRACTION, 2014, 29 (02) : 159 - 169
  • [3] [Anonymous], 2017, Canadian Cancer Statistics 2017, P1
  • [4] Beckhoff H., 2006, HDB PRACTICAL XRAY F, V49, P6221
  • [5] Trace elements in human cancerous and healthy tissues: A comparative study by EDXRF, TXRF, synchrotron radiation and PIXE
    Carvalho, M. L.
    Magalhaes, T.
    Becker, M.
    von Bohlen, A.
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2007, 62 (09) : 1004 - 1011
  • [6] The optimization of an energy-dispersive X-ray diffraction system for potential clinical application
    Chaparian, A.
    Oghabian, M. A.
    Changizi, V.
    Farquharson, M. J.
    [J]. APPLIED RADIATION AND ISOTOPES, 2010, 68 (12) : 2237 - 2245
  • [7] Preliminary study of human breast tissue using synchrotron radiation combining WAXS and SAXS techniques
    Conceicao, A. L. C.
    Antoniassi, M.
    Poletti, M. E.
    Caldas, L. V. E.
    [J]. APPLIED RADIATION AND ISOTOPES, 2010, 68 (4-5) : 799 - 803
  • [8] Costello LC, 1998, PROSTATE, V35, P285, DOI 10.1002/(SICI)1097-0045(19980601)35:4<285::AID-PROS8>3.0.CO
  • [9] 2-F
  • [10] Discriminant analysis of trace elements in normal, benign and malignant breast tissues measured by total reflection X-ray fluorescence
    da Silva, Marina Piacenti
    Araujo Domingues Zucchi, Orgheda Luiza
    Ribeiro-Silva, Alfredo
    Poletti, Martin E.
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2009, 64 (06) : 587 - 592