Fibre optic glucose sensor

被引:67
|
作者
Binu, S. [1 ]
Pillai, V. R. Mahadevan [1 ]
Pradeepkumar, V. [2 ]
Padhy, B. B. [2 ]
Joseph, C. S. [2 ]
Chandrasekaran, N. [3 ]
机构
[1] Univ Kerala, Dept Optoelect, Thiruvananthapuram 695581, Kerala, India
[2] Deemed Univ, Def Inst Adv Technol, Dept Appl Phys, DRDO, Pune 411025, Maharashtra, India
[3] DRDO, Def Avion Res Estab, Bangalore 560075, Karnataka, India
关键词
Fibre optic glucose sensor; PMMA (polymethyl methacrylate) fibre; Fibre optic probe; REFRACTIVE-INDEX; SYSTEMS;
D O I
10.1016/j.msec.2008.06.007
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper reports the principle of operation. the design aspects, experimentation and performance of a fibre optic probe for sensing the variation of refractive index with concentration of glucose in distilled water. The device consists of fibre optic transmitter, Fibre optic probe, reflecting mirror, temperature controller, photodiode detector and a digital multimeter. The variations in refractive indices of glucose for a range of concentrations varying from distilled water are studied. The measurements are carried out for glucose solutions with concentrations of 5, 10, 15, 20 and 25 g per 100 ml of distilled water. For a concentration change from 0 to 25 g per 100 ml, the refractive index varies from 1.3322 to 1.3617, the displacement corresponding to peak voltage changes by 1000 pin and the variation detected in peak voltage is 182 mV. The results show that the displacement corresponding to maximum intensity observed at different concentrations is a function of the refractive index of the medium and there is a linear relationship between them. From the experimental results it is concluded that as the concentration or density of a solute increases, the refractive index increases proportionately. The stability of the sensor, high sensitivity and the simplicity of the design. low cost of fabrication, make it suitable for chemical, biomedical. pharmaceutical and process control applications. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:183 / 186
页数:4
相关论文
共 50 条
  • [1] Fibre-optic glucose sensor with a pH optrode as the transducer
    Trettnak, W.
    Leiner, M.J.P.
    Wolfbeis, O.S.
    Biosensors, 1988, 4 (01) : 15 - 26
  • [2] FIBRE-OPTIC GLUCOSE SENSOR WITH A PH OPTRODE AS THE TRANSDUCER
    TRETTNAK, W
    LEINER, MJP
    WOLFBEIS, OS
    BIOSENSORS, 1989, 4 (01) : 15 - 26
  • [3] Fibre optic glucose sensor (vol 29, pg 183, 2009)
    Binu, S.
    Pillai, V. P. Mahadevan
    Pradeepkumar, V.
    Padhy, B. B.
    Joseph, C. S.
    Chandrasekaran, N.
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2011, 31 (07): : 1618 - 1618
  • [4] Correction of fibre optic ion sensor readings with use of fibre optic temperature sensor
    Maciejewski, J
    Pozniak, K
    Romaniuk, R
    OPTICAL FIBERS AND THEIR APPLICATIONS VI, 1999, 3731 : 161 - 166
  • [5] Fibre optic sensor technology
    Grattan, KTV
    Sun, T
    PROCEEDINGS OF THE FIRST INTERNATIONAL SYMPOSIUM ON INSTRUMENTATION SCIENCE AND TECHNOLOGY, 1999, : 30 - 39
  • [6] Self aligning fibre for a fibre optic voltage sensor
    Hambley, P.
    Michie, A.
    Bassett, I.
    Henry, P.
    Ingram, J.
    19TH INTERNATIONAL CONFERENCE ON OPTICAL FIBRE SENSORS, PTS 1 AND 2, 2008, 7004
  • [7] Fibre-optic sensor boom
    不详
    PROFESSIONAL ENGINEERING, 1998, 11 (07) : 9 - 9
  • [8] Fibre optic current sensor network
    Goyal, S.
    Irvine-Halliday, D.
    Thomson, R.M.
    Shelley, J.T.
    Canalizo, A.E.
    Journal of Electrical and Electronics Engineering, Australia, 1998, 18 (03): : 197 - 204
  • [9] Novel fibre optic strain sensor
    Dep of Physical Sciences, Glasgow, United Kingdom
    Int J Electron, 4 (545-552):
  • [10] Fibre-optic temperature sensor
    Zhao, Jie
    Fu, Wanshi
    Yadian Yu Shengguang/Piezoelectrics and Acoustooptics, 1994, 16 (02): : 29 - 32