Pediatric Radiation Exposure and Effective Dose Reduction during Voiding Cystourethrography

被引:77
作者
Ward, Valerie L. [1 ]
Strauss, Keith J. [1 ]
Barnewolt, Carol E. [1 ]
Zurakowski, David [1 ,2 ]
Venkatakrishnan, Vaidehi [1 ]
Fahey, Frederic H. [3 ]
Lebowitz, Robert L. [1 ]
Taylor, George A. [1 ]
机构
[1] Harvard Univ, Sch Med, Childrens Hosp Boston, Dept Radiol, Boston, MA 02115 USA
[2] Harvard Univ, Sch Med, Childrens Hosp Boston, Dept Orthopaed, Boston, MA 02115 USA
[3] Harvard Univ, Sch Med, Childrens Hosp Boston, Dept Nucl Med, Boston, MA 02115 USA
关键词
D O I
10.1148/radiol.2492062066
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare radiation exposure and effective dose in children who underwent voiding cystourethrography (VCUG) performed with grid-controlled variable-rate pulsed fluoroscopy (GCPFL) with radiation exposure and effective dose in children who underwent VCUG performed with continuous fluoroscopy (CFL) and to compare these effective doses with those estimated with radionuclide cystography (RNC). Materials and Methods: Institutional review board approval was obtained, and the informed consent requirement was waived for this HIPAA-compliant retrospective study. Radiation exposure and fluoroscopy time during VCUG were reviewed in 145 children (75 girls, 70 boys; age range, 3 days to 8 years) who underwent GCPFL or CFL between 2001 and 2002. Children were grouped on the basis of the fluoroscopy unit used and their supine anteroposterior abdominal diameter (group 1, 8.08.5-cm diameter; group 2, 10-11-cm diameter; group 3, 12-13-cm diameter). Analysis of variance was used to compare radiation exposure and fluoroscopy time between fluoroscopy units and patient diameter groups. Effective doses were calculated and compared for both fluoroscopes and for estimated RNC dose values. Results: GCPFL resulted in a significant reduction in total radiation exposure, which was at least eight times lower than that with CFL in all three groups (P < .001 for all). There was no significant difference in fluoroscopy time (P > .50). Effective radiation doses from GCPFL were approximately one order of magnitude lower than those from CFL but one order of magnitude higher than those from RNC. Conclusion: In children, VCUG can be performed with a GCPFL unit that delivers radiation exposures that are at least eight times lower than those delivered by a conventional CFL unit. (c) RSNA, 2008
引用
收藏
页码:1002 / 1009
页数:8
相关论文
共 43 条
[31]   Helical CT of the body: Are settings adjusted for pediatric patients? [J].
Paterson, A ;
Frush, DP ;
Donnelly, LF .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :297-301
[32]   Fluoroscopy-controlled voiding cystourethrography in infants and children: Are the radiation risks trivial? [J].
Perisinakis, K ;
Raissaki, M ;
Damilakis, J ;
Stratakis, J ;
Neratzoulakis, J ;
Gourtsoyiannis, N .
EUROPEAN RADIOLOGY, 2006, 16 (04) :846-851
[33]   Dose and image quality in the comparison of analogue and digital techniques in paediatric urology examinations [J].
Persliden, J ;
Helmrot, E ;
Hjort, P ;
Resjö, M .
EUROPEAN RADIOLOGY, 2004, 14 (04) :638-644
[34]   Radiation dose and image quality in pediatric CT: Effect of technical factors and phantom size and shape [J].
Siegel, MJ ;
Schmidt, B ;
Bradley, D ;
Suess, C ;
Hildebolt, C .
RADIOLOGY, 2004, 233 (02) :515-522
[35]   Pediatric interventional radiography equipment: safety considerations [J].
Strauss, Keith J. .
PEDIATRIC RADIOLOGY, 2006, 36 (Suppl 2) :126-135
[36]   The ALARA (as low as reasonably achievable) concept in pediatric interventional and fluoroscopic imaging: striving to keep radiation doses as low as possible during fluoroscopy of pediatric patients - a white paper executive summary [J].
Strauss, Keith J. ;
Kaste, Sue C. .
PEDIATRIC RADIOLOGY, 2006, 36 (Suppl 2) :110-112
[37]  
STRAUSS KJ, 1995, SYLLABUS CATEGORICAL, P171
[38]  
TREVES ST, 1995, PEDIAT NUCL MED, P411
[39]   Pediatric phantoms [J].
Varchena, V .
PEDIATRIC RADIOLOGY, 2002, 32 (04) :280-284
[40]   Minimizing radiation injury and neoplastic effects during pediatric fluoroscopy: what should we know? [J].
Wagner, Louis Kenneth .
PEDIATRIC RADIOLOGY, 2006, 36 (Suppl 2) :141-145