STARR: A Speech Test for Evaluation of the Effectiveness of Auditory Prostheses Under Realistic Conditions

被引:27
作者
Boyle, Patrick J. [1 ,2 ]
Nunn, Terry B. [3 ]
O'Connor, Alec Fitzgerald [3 ]
Moore, Brian C. J. [1 ]
机构
[1] Univ Cambridge, Dept Expt Psychol, Cambridge CB2 3EB, England
[2] Adv Bion European Res Ctr, Res & Technol Dept, Hannover, Germany
[3] Guys & St Thomas Natl Hlth Serv Trust, Auditory Implant Programme, London, England
基金
英国医学研究理事会;
关键词
COCHLEAR IMPLANT RECIPIENTS; INPUT DYNAMIC-RANGE; WORD RECOGNITION; HEARING-LOSS; NOISE; PERCEPTION; INTELLIGIBILITY; LISTENERS; RECEPTION; FREQUENCY;
D O I
10.1097/AUD.0b013e31826a8e82
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objectives: The objective was to develop and evaluate a new sentence test, the Sentence Test with Adaptive Randomized Roving levels, intended to emulate everyday listening experience, using both normal-hearing (NH) and cochlear implant (CI) groups, examining practicality, learning, test-retest variability, and interlist variability. Design: In experiment 1, each of 25 NH adults was tested using five lists, each comprising 30 sentences. One male and one female speaker each spoke 15 sentences. Ten sentences were presented at each of three presentation levels: 50, 65, and 80 dB SPL. The relative level of a speech-shaped noise was varied adaptively to estimate the speech reception threshold (SRT). Counterbalance for list order was achieved by staggering the allocation of lists to participants. To allow assessment of learning effects, no practice was given. The variability of mean SRTs across lists was small, but correction factors were derived for each list so that, after correction, all lists gave the same mean SRT. Test-retest variability was estimated by examining the corrected SRTs for each subject's five lists. In experiment 2, 25 CI users each received one test list after a small amount of practice. Experiment 3 examined the effect of speech rate using time-compressed speech, for age-matched NH and CI users. Results: The mean SRT for the NH participants was approximately -6 dB and was similar for the male and female speakers. There was a small but significant improvement in SRTs between the first and later lists administered, but no further improvement for subsequent lists. On the basis of the variability of the corrected SRTs within each participant, a 2.2 dB difference in SRT is meaningful for comparisons using one test list per condition, for a single participant. The percentage of key words correct varied with presentation level over a 13% range, being best at 65 dB SPL. Only 40% of the CI group achieved an SRT lower than 20 dB for both speakers. There was large individual variability in the SRTs, and SRTs were higher for the female than for the male speaker. For the CI participants, the percentage of key words correct varied markedly with level, from 19% at the lowest level to 57% at the medium level. Time compression had a small effect for NH participants but a very large effect for CI participants. Conclusions: The Sentence Test with Adaptive Randomized Roving levels seems practical to administer and is reasonably sensitive. For NH participants, a 2.2 dB difference in SRT is meaningful for a single list per condition and a single participant. Although learning effects were small for NH participants, it seems prudent to provide some practice sentences when testing hearing-impaired or CI participants. The very large effect of time compression for the CI group has implications for live voice testing of children, because speech rate is only poorly controlled in such testing. (Ear & Hearing 2013;34;203-212)
引用
收藏
页码:203 / 212
页数:10
相关论文
共 46 条
  • [1] [Anonymous], 2009, PRAAT DOING PHONETIC
  • [2] Bench J, 1979, Br J Audiol, V13, P108, DOI 10.3109/03005367909078884
  • [3] FACTORS PREDICTING POSTOPERATIVE SENTENCE SCORES IN POSTLINGUISTICALLY DEAF ADULT COCHLEAR IMPLANT PATIENTS
    BLAMEY, PJ
    PYMAN, BC
    CLARK, GM
    DOWELL, RC
    GORDON, M
    BROWN, AM
    HOLLOW, RD
    [J]. ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1992, 101 (04) : 342 - 348
  • [4] VOICE FUNDAMENTAL-FREQUENCY AS AN AUDITORY SUPPLEMENT TO THE SPEECHREADING OF SENTENCES
    BOOTHROYD, A
    HNATHCHISOLM, T
    HANIN, L
    KISHONRABIN, L
    [J]. EAR AND HEARING, 1988, 9 (06) : 306 - 312
  • [5] BOOTHROYD A, 2008, J ACAD REHABILITATIV, V41, P31
  • [6] Comparison of dual-time-constant and fast-acting automatic gain control (AGC) systems in cochlear implants
    Boyle, Patrick J.
    Buechner, Andreas
    Stone, Michael A.
    Lenarz, Thomas
    Moore, Brian C. J.
    [J]. INTERNATIONAL JOURNAL OF AUDIOLOGY, 2009, 48 (04) : 211 - 221
  • [7] The Effect of Instantaneous Input Dynamic Range Setting on the Speech Perception of Children with the Nucleus 24 Implant
    Davidson, Lisa S.
    Skinner, Margaret W.
    Holstad, Beth A.
    Fears, Beverly T.
    Richter, Marie K.
    Matusofsky, Margaret
    Brenner, Christine
    Holden, Timothy
    Birath, Amy
    Kettel, Jerrica L.
    Scollie, Susan
    [J]. EAR AND HEARING, 2009, 30 (03) : 340 - 349
  • [8] Effects of presentation level on phoneme and sentence recognition in quiet by cochlear implant listeners
    Donaldson, GS
    Allen, SL
    [J]. EAR AND HEARING, 2003, 24 (05) : 392 - 405
  • [10] ANALYSES OF PHONETIC COMPOSITION AND WORD FAMILIARITY ATTRIBUTES OF CNC INTELLIGIBILITY WORD LISTS
    ELKINS, EF
    [J]. JOURNAL OF SPEECH AND HEARING DISORDERS, 1970, 35 (02): : 156 - 160