Microwave liver ablation: Influence of hepatic vein size on heat-sink effect in a porcine model

被引:179
作者
Yu, Nam C. [1 ]
Raman, Steven S.
Kim, Young Jun [2 ]
Lassman, Charles
Chang, Xinlian
Lu, David S. K.
机构
[1] Univ Calif Los Angeles, Med Imaging Informat Grp, Biomed Informat Ctr, CHS BR 158, Los Angeles, CA 90095 USA
[2] Konkuk Univ Hosp, Dept Radiol, Seoul, South Korea
关键词
D O I
10.1016/j.jvir.2008.03.023
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To determine influence of hepatic vein size on perfusion-mediated attenuation in adjacent microwave thermal ablation. MATERIALS AND METHODS: With approval of the institutional animal research committee, seven Yorkshire pigs underwent percutaneous (n = 2) or open (n = 5) microwave liver ablation under general anesthesia. In each, multiple ultrasound-guided, nonoverlapping thermal lesions were created within 1 cm of hepatic veins in a 5-10-minute ablation at 45 W. After euthanasia, the liver was harvested and sectioned at 0.5-cm intervals and the degree of perivascular coagulation attenuation was graded on histopathologic analysis. Correlation between venous size (small, <= 3 mm; medium, 3-6 mm; and large, > 6 mm) and attenuation grade was performed with use of the Spearman rank test. RESULTS: In 63 of 103 sections (61%)-29 of 37 (78%) small, 27 of 48 (56%) medium, and seven of 18 (39%) large veins-the thermal injury extended to the vein wall around the entire circumference of the coagulation front without distortion of the ablation margin. In 40 of 103 sections (38.9%), varying degrees of concave distortion of perivenous ablation margins were noted, with significant correlation between vein size and heat-sink extent (P <.01). However, thermal injury extended to the vascular wall in all sections without complete circumferential sparing of liver tissue. Around two thrombosed veins, thermal lesions encased the vessels, producing paradoxically convex ablation margins. CONCLUSIONS: Although the heat-sink effect was significantly dependent on hepatic vein size, the majority of pathologic section's exhibited no or minimal effect. Further study is required to assess clinical implications.
引用
收藏
页码:1087 / 1092
页数:6
相关论文
共 23 条
[1]   A field-focusing device to increase power output of ThermoRod™ implants for thermal ablation of tissue [J].
Berthusen, AH ;
Tucker, RD ;
Park, JB .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 63 (05) :650-656
[2]  
Goldberg SN, 1996, ACAD RADIOL, V3, P636
[3]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[4]   Microwave ablation: Results with a 2.45-GHz applicator in ex vivo bovine and in vivo porcine liver [J].
Hines-Peralta, AU ;
Pirani, N ;
Clegg, P ;
Cronin, N ;
Ryan, TP ;
Liu, ZJ ;
Goldberg, SN .
RADIOLOGY, 2006, 239 (01) :94-102
[5]   In vitro and in vivo evaluation of the thermal patterns and lesions of catheter ablation with a microwave monopole antenna [J].
Keane, D ;
Ruskin, J ;
Norris, N ;
Chapelon, PA ;
Bérubé, D .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2004, 10 (02) :111-119
[6]   Percutaneous radiofrequency ablation of hepatic colorectal metastases - Technique, indications, results, and new promises [J].
Lencioni, R ;
Crocetti, L ;
Cioni, D ;
Della Pina, C ;
Bartolozzi, C .
INVESTIGATIVE RADIOLOGY, 2004, 39 (11) :689-697
[7]   Early-stage hepatocellular carcinoma in patients with cirrhosis: Long-term results of percutaneous image-guided radiofrequency ablation [J].
Lencioni, R ;
Cioni, D ;
Crocetti, L ;
Franchini, C ;
Della Pina, C ;
Lera, J ;
Bartolozzi, C .
RADIOLOGY, 2005, 234 (03) :961-967
[8]   Studies on microwaves in medicine and biology: From snails to humans [J].
Lin, JC .
BIOELECTROMAGNETICS, 2004, 25 (03) :146-159
[9]   Effect of vessel size on creation of hepatic radiofrequency lesions in Pigs: Assessment of the "heat sink" effect [J].
Lu, DSK ;
Raman, SS ;
Vodopich, DJ ;
Wang, M ;
Sayre, J ;
Lassman, C .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 178 (01) :47-51
[10]   Influence of larze peritumoral vessels on outcome of radiofrequency ablation of liver tumors [J].
Lu, DSK ;
Raman, SS ;
Limanond, P ;
Aziz, D ;
Economou, J ;
Busuttil, R ;
Sayre, J .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (10) :1267-1274