Patient and staff dosimetry during radiographic procedures in an intensive care unit

被引:9
作者
Fernandez, Rosario [1 ]
Moreno-Torres, Miguel [2 ]
Contreras, Antonia M. [2 ]
Nunez, Maria I. [3 ]
Guirado, Damian [2 ]
Penas, Luis [1 ]
机构
[1] Hosp Univ San Cecilio, Unidad Cuidados Intens, Granada, Spain
[2] Hosp Univ San Cecilio, Unidad Radiofis Hosp, Granada, Spain
[3] Univ Granada, Dept Radiol & Med Fis, E-18071 Granada, Spain
关键词
intensive care; radiological protection; radiation dose; radiology; CHEST RADIOGRAPHY; RADIATION-EXPOSURE; INFANTS; UK;
D O I
10.1088/0952-4746/35/3/727
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The performance of radiography in the Intensive Care Unit (ICU) may be associated with a certain level of radiation exposure for staff and patients in the unit. Little evidence on exposure levels is available in the literature. However, healthcare professionals in the ICUs at our centre tend to leave the room during radiographic examinations, potentially compromising patient care. The objectives of this study were to quantify dose levels within the ICU and to evaluate the performance of ICU x-ray studies according to patient dose measurements. This study was conducted in the 18-bed ICU of a third-level hospital. The scattering radiation due to mobile x-ray examinations was measured by using four personal thermoluminiscent dosimeters (TLDs). The dose area product (DAP) was measured at each examination using a transmission chamber installed on the diaphragm of the x-ray equipment. Based on the TLD readings and taking account of the error margin, the annual dose to patients and staff was less than 0.6 mSv. The value given by the DAP meter for chest x-rays was 94 +/- 17 mGy cm(2); this value is well below the lower limit recommended by different agencies and committees. Exposure levels were found to be extremely low and pose no apparent risk to staff or to those in beds adjacent to the patients undergoing x-ray examinations, which were correctly performed in the unit.
引用
收藏
页码:727 / 732
页数:6
相关论文
共 28 条
  • [1] [Anonymous], 2011, ICRU Report 85a
  • [2] Radiology in pediatric intensive care units
    Beer, M.
    Wirth, C.
    Neubauer, H.
    Wirbelauer, J.
    [J]. MEDIZINISCHE KLINIK, 2011, 106 (02) : 103 - 110
  • [3] Utility of routine chest radiographs in a medical-surgical intensive care unit: a quality assurance survey
    Chahine-Malus, N
    Stewart, T
    Lapinsky, SE
    Marras, T
    Dancey, D
    Leung, R
    Mehta, S
    [J]. CRITICAL CARE, 2001, 5 (05): : 271 - 275
  • [4] Cho Jeong-Yeon, 2012, Restor Dent Endod, V37, P160
  • [5] Exposure to Low-Dose Ionizing Radiation from Medical Imaging Procedures.
    Fazel, Reza
    Krumholz, Harlan M.
    Wang, Yongfei
    Ross, Joseph S.
    Chen, Jersey
    Ting, Henry H.
    Shah, Nilay D.
    Nasir, Khurram
    Einstein, Andrew J.
    Nallamothu, Brahmajee K.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2009, 361 (09) : 849 - 857
  • [6] Fukui R, 2013, YONAGO ACTA MED, V56, P85
  • [7] The risk of radiation in the intensive care unit
    Galimany Masclans, J.
    Blanca Yela, I.
    [J]. ENFERMERIA INTENSIVA, 2013, 24 (02): : 49 - 50
  • [8] Chest radiography practice in critically ill patients: A postal survey in the Netherlands
    Graat M.E.
    Hendrikse K.A.
    Spronk P.E.
    Korevaar J.C.
    Stoker J.
    Schultz M.J.
    [J]. BMC Medical Imaging, 6 (1)
  • [9] The UK national patient dose database: now and in the future
    Hart, D
    Wall, BF
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (906) : 361 - 365
  • [10] National reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK
    Hart, D.
    Hillier, M. C.
    Wall, B. F.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2009, 82 (973) : 1 - 12