Artifact reduction strategies for prosthetic heart valve CT imaging

被引:16
作者
Habets, Jesse [1 ]
Symersky, Petr [2 ]
Leiner, Tim [1 ]
de Mol, Bas A. J. M. [2 ]
Mali, Willem P. Th M. [1 ]
Budde, Ricardo P. J. [1 ,3 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiol, NL-3508 GA Utrecht, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Cardiothorac Surg, NL-1105 AZ Amsterdam, Netherlands
[3] Gelre Hosp, Dept Radiol, Apeldoorn, Netherlands
关键词
Computed tomography; Prosthetic heart valve; Metal artifact reduction; Iterative reconstruction; Tube voltage; COMPUTED-TOMOGRAPHY; ECHOCARDIOGRAPHY;
D O I
10.1007/s10554-012-0041-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Multislice CT evaluation of prosthetic heart valves (PHV) is limited by PHV-related artifacts. We assessed the influence of different kV settings, a metal artifact reduction filter (MARF) and an iterative reconstruction algorithm (IR) on PHV-induced artifacts in an in vitro model. A Medtronic-Hall tilting disc and St Jude bileafet PHV were imaged using a 64-slice scanner with 100 kV/165 mAs, 120 kV/100 mAs, 140 kV/67 mAs at an equal CTDIvol. Images were reconstructed with (1) filtered back projection (FBP), (2) IR, (3) MARF and (4) MARF and IR. Hypo- and hyperdense artifacts volumes (mean mm(3) +/- SD) were quantified with 2 thresholds (<=-50 and >= 175 Hounsfield Units). Image noise was measured and the presence of secondary artifacts was scored by 2 observers independently. Mean hypodense artifacts for the Medtronic-Hall/St Jude valve (FBP) were 966 +/- 23/1,738 +/- 21 at 100 kV, 610 +/- 13/991 +/- 12 at 120 kV, and 420 +/- 9/634 +/- 9 at 140 kV. Compared to FBP, hypodense artifact reductions for IR were 9/8 %, 10/7 % and 12/6 % respectively, for MARF 92 %/84 %, 89/81 % and 86/77 % respectively; for MARF + IR 94/85 %, 92/82 %, and 90/79 % respectively. Mean hyperdense artifacts for the Medtronic-Hall/St Jude valve were 5,530 +/- 48/6,940 +/- 70 at 100 kV, 5,120 +/- 42/6,250 +/- 53 at 120 kV, and 5,011 +/- 52/6,000 +/- 0 at 140 kV. Reductions for IR were 2/2 %, 2/3 % and 3/4 % respectively, for MARF were 9/30 %, 0/25 %, 5/22 % respectively, MARF + IR 12/32 %, 4/27 % and 7/25 % respectively. Secondary artifacts were found in all MARF images. Image noise was reduced in the IR images. In vitro PHV-related artifacts can be reduced by increasing kV despite maintaining identical CTDIvol. Although MARF is more effective than IR, it induces secondary artifacts.
引用
收藏
页码:2099 / 2108
页数:10
相关论文
共 21 条
  • [1] Metal artifact reduction in CT using tissue-class modeling and adaptive prefiltering
    Bal, Matthieu
    Spies, Lothar
    [J]. MEDICAL PHYSICS, 2006, 33 (08) : 2852 - 2859
  • [2] Artifacts in CT: Recognition and avoidance
    Barrett, JF
    Keat, N
    [J]. RADIOGRAPHICS, 2004, 24 (06) : 1679 - 1691
  • [3] Evaluation of Two Iterative Techniques for Reducing Metal Artifacts in Computed Tomography
    Boas, F. Edward
    Fleischmann, Dominik
    [J]. RADIOLOGY, 2011, 259 (03) : 894 - 902
  • [4] Evaluation of Anatomic Valve Opening and Leaflet Morphology in Aortic Valve Bioprosthesis by Using Multidetector CT: Comparison with Transthoracic Echocardiography
    Chenot, Fabien
    Montant, Patrick
    Goffinet, Celine
    Pasquet, Agnes
    Vancraeynest, David
    Coche, Emmanuel
    Vanoverschelde, Jean-Louis
    Gerber, Bernhard L.
    [J]. RADIOLOGY, 2010, 255 (02) : 377 - 385
  • [5] Multislice Computed Tomography in Infective Endocarditis Comparison With Transesophageal Echocardiography and Intraoperative Findings
    Feuchtner, Gudrun M.
    Stolzmann, Paul
    Dichtl, Wolfgang
    Schertler, Thomas
    Bonatti, Johannes
    Scheffel, Hans
    Mueller, Silvana
    Plass, Andre
    Mueller, Ludwig
    Bartel, Thomas
    Wolf, Florian
    Alkadhi, Hatem
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 53 (05) : 436 - 444
  • [6] Computed tomography-old ideas and new technology
    Fleischmann, Dominik
    Boas, F. Edward
    [J]. EUROPEAN RADIOLOGY, 2011, 21 (03) : 510 - 517
  • [7] Reoperation for prosthetic aortic valve obstruction in the era of echocardiography: Trends in diagnostic testing and comparison with surgical findings
    Girard, SE
    Miller, FA
    Orszulak, TA
    Mullany, CJ
    Montgomery, S
    Edwards, WD
    Tazelaar, HD
    Malouf, JF
    Tajik, AJ
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) : 579 - 584
  • [8] A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves
    Habets, Jesse
    Symersky, Petr
    de Mol, Bas A. J. M.
    Mali, Willem P. Th. M.
    Leiner, Tim
    Budde, Ricardo P. J.
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 (06) : 1567 - 1575
  • [9] Diagnostic evaluation of left-sided prosthetic heart valve dysfunction
    Habets, Jesse
    Budde, Ricardo P.
    Symersky, Petr
    van den Brink, Renee B.
    de Mol, Bas A.
    Mali, Willem P.
    van Herwerden, Lex A.
    Chamuleau, Steven A.
    [J]. NATURE REVIEWS CARDIOLOGY, 2011, 8 (08) : 466 - 478
  • [10] Prosthetic heart valve assessment with multidetector-row CT: imaging characteristics of 91 valves in 83 patients
    Habets, Jesse
    Symersky, Petr
    van Herwerden, Lex A.
    de Mol, Bas A. J. M.
    Spijkerboer, Anje M.
    Mali, Willem P. Th. M.
    Budde, Ricardo P. J.
    [J]. EUROPEAN RADIOLOGY, 2011, 21 (07) : 1390 - 1396