Dose distributions using kilovoltage x-rays and dose enhancement from iodine contrast agents

被引:95
作者
Mesa, AV [1 ]
Norman, A
Solberg, TD
Demarco, JJ
Smathers, JB
机构
[1] Univ Calif Los Angeles, Dept Radiat Oncol, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Dept Radiol Sci, Los Angeles, CA 90095 USA
[3] Univ Calif Los Angeles, Jonsson Comprehens Canc Ctr, Los Angeles, CA 90095 USA
关键词
D O I
10.1088/0031-9155/44/8/308
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In x-ray phototherapy of brain tumours, the tumour is loaded with iodine and exposed to kilovoltage x-rays. Due to the high photoelectric cross sections of iodine, substantial photoelectric interactions occur. The flux of photoelectrons, characteristic x-rays and Auger electrons produce a localized dose enhancement. A modified computed tomography scanner, CTRx. can be used both for tumour localization and delivery of the dose enhancement therapy. Monte Carlo methods were employed to simulate the treatment of iodinated brain tumours with a CTRx. The calculated results reveal the effect of tumour iodine concentration on dose distribution, the degree of skull bone sparing with the application of multiple arcs, and the homogeneity of tumour dose distribution versus iodine concentration. A comparison with 10 MV stereotactic radiosurgery treatment shows the potential of CTRx treatment relative to conventional treatment modalities.
引用
收藏
页码:1955 / 1968
页数:14
相关论文
共 19 条
[1]  
ALFIDI RJ, 1975, RADIOLOGY, V124, P199
[2]  
[Anonymous], INT J RAD ONCOLOGY B
[3]  
BARTH RF, 1992, CANCER-AM CANCER SOC, V70, P2995
[4]  
BREISMEISTER JF, 1993, LA12625
[5]   A comparison of three stereotactic radiotherapy techniques; arcs vs. noncoplanar fixed fields vs. intensity modulation [J].
Cardinale, RM ;
Benedict, SH ;
Wu, QW ;
Zwicker, RD ;
Gaballa, HE ;
Mohan, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (02) :431-436
[6]   CT-based Monte Carlo simulation tool for dosimetry planning and analysis [J].
DeMarco, JJ ;
Solberg, TD ;
Smathers, JB .
MEDICAL PHYSICS, 1998, 25 (01) :1-11
[7]  
*ICRP, 1989, 44 ICRU
[8]  
International Commission on Radiological Protection, 1975, REPORT TASK GROUP RE
[9]   DIAGNOSIS AND TREATMENT OF SPONTANEOUS CANINE BRAIN-TUMORS WITH A CT SCANNER [J].
IWAMOTO, KS ;
NORMAN, A ;
FRESHWATER, DB ;
INGRAM, M ;
SKILLEN, RG .
RADIOTHERAPY AND ONCOLOGY, 1993, 26 (01) :76-78
[10]   RADIATION-DOSE ENHANCEMENT THERAPY WITH IODINE IN RABBIT VX-2 BRAIN-TUMORS [J].
IWAMOTO, KS ;
COCHRAN, ST ;
WINTER, J ;
HOLBURT, E ;
HIGASHIDA, RT ;
NORMAN, A .
RADIOTHERAPY AND ONCOLOGY, 1987, 8 (02) :161-170