The Charnley versus the Spectron hip prosthesis -: Radiographic evaluation of a randomized, prospective study of 2 different hip implants

被引:45
作者
Garellick, G [1 ]
Malchau, H [1 ]
Regnér, H [1 ]
Herberts, P [1 ]
机构
[1] Gothenburg Univ, Sahlgrens Hosp, Dept Orthopaed, Inst Surg Sci, S-41345 Gothenburg, Sweden
关键词
hip prosthesis; cemented; radiographic evaluation; Charnley; Spectron;
D O I
10.1016/S0883-5403(99)90096-7
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A total of 410 hips were randomized to treatment with either a Charnley (206 hips) or a Spectron (204 hips) total hip arthroplasty. The patients were operated on by a standardized procedure using a contemporary cementing technique and were followed after 1, 3, 5 to 6, and 10 years. The postoperative radiographs showed a significantly increased rate of malalignment and consequently low grade of cement mantle quality of the Charnley stem compared to the Spectron. No differences concerning cement mantle quality or positioning were found between the Charnley ogee cup and the metal-backed Spectron. Evaluation of the follow-up radiographs revealed 10 loose Charnley stems and I loose Spectron stem and 4 loose Charnley ogee cups and 23 loose Spectron metal-backed cups. The differences of revision rate for the femoral and acetabular components of the 2 prostheses were significant (P = .03, Charnley femoral component more frequent: P = .03, Spectron acetabular component more frequent). The radiographic evaluation strengthened this disparity Poor wear characteristics of the metal-backed Spectron cup are perhaps the main reason for the highly significant difference in mechanical failure rate between the 2 cups. We therefore propose that metal-backing of cemented cups should be avoided, at least when combined with larger femoral heads. The difficulty in positioning the Charnley stem with an adequate cement mantle, especially in the absence of trochanteric ostectomy, might explain the inferior Charnley stem longevity in this study.
引用
收藏
页码:414 / 425
页数:12
相关论文
共 56 条
[1]  
Ahlb_ack S., 1968, ACTA RADIOL SCAND S, P7, DOI 10.1148/23.1.134
[2]  
[Anonymous], 1979, CLIN ORTHOP RELAT RE
[3]   IMPROVED CEMENTING TECHNIQUES AND FEMORAL COMPONENT LOOSENING IN YOUNG-PATIENTS WITH HIP-ARTHROPLASTY - A 12-YEAR RADIOGRAPHIC REVIEW [J].
BARRACK, RL ;
MULROY, RD ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :385-389
[4]  
BARTEL DL, 1983, P 11 OP SCI M HIP SO, P229
[5]  
CALLAGHAN JJ, 1995, CLIN ORTHOP RELAT R, V317, P14
[6]   A LARGE COLLAR INCREASES NECK RESORPTION IN TOTAL HIP-REPLACEMENT - 204 HIPS EVALUATED DURING 5 YEARS [J].
CARLSSON, AS ;
RYDBERG, J ;
ONNERFALT, R .
ACTA ORTHOPAEDICA SCANDINAVICA, 1995, 66 (04) :339-342
[7]  
CARTER DR, 1983, P 11 OP SCI M HIP SO, P216
[8]   POLYETHYLENE WEAR IN CEMENTED METAL-BACKED ACETABULAR CUPS [J].
CATES, HE ;
FARIS, PM ;
KEATING, EM ;
RITTER, MA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02) :249-253
[9]   OPTIMUM SIZE OF PROSTHETIC HEADS IN RELATION TO WEAR OF PLASTIC SOCKETS IN TOTAL REPLACEMENT OF HIP [J].
CHARNLEY, J ;
KAMANGAR, A ;
LONGFIELD, MD .
MEDICAL & BIOLOGICAL ENGINEERING, 1969, 7 (01) :31-+
[10]  
Charnley J., 1979, LOW FRICTION ARTHROP