Simultaneous SPECT and CT with an opposed dual head gamma camera system and conventional parallel hole collimators

被引:4
作者
Kimiaei, S
Larsson, SA
机构
[1] Dept. of Hospital Physics, Karolinska Hospital
关键词
D O I
10.1109/23.531888
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
All examinations aiming at quantitative assessment of activity concentrations suffer from non-uniform attenuation due to inhomogeneous tissue density in various body regions. In such applications, a combined transmission tomography with SPECT has demonstrated great clinical advantages since it allows a pixel by pixel correction for attenuation based on measured data. We have developed a method for transmission studies that is applicable to dual head detector systems with conventional parallel-hole collimators. A scanning line-source (Am-241) is mounted close to one camera head with its beam directed towards the other head. Two energy discrimination windows are used for each camera head. The camera-head near to the transmission source registrates SPECT-data in one window and the spill-over of scattered emission photons into the Am-window. The opposed camera registrates SPECT-data in one window and the transmission data from the Am-source in another window. Post acquisition corrections are made for differences in camera uniformity and for loss of SPECT-data due to shadowing of the field-of-view during the transverse movement of the transmission device (about 5%). The height of the collimated slit source is only 30 mm to allow a setting of the camera head with the transmission device close to the patient.
引用
收藏
页码:2239 / 2243
页数:5
相关论文
共 9 条
[1]  
BAILEY DL, 1987, J NUCL MED, V28, P844
[3]  
FICARO EP, 1994, J NUCL MED, V35, P652
[4]  
FREY EC, 1992, J NUCL MED, V33, P2238
[5]  
Hubbell J.H., 1995, 5632 NISTIR US DEP C
[6]  
JASZCZAK RJ, 1993, J NUCL MED, V34, P1577
[7]   SIMULTANEOUS SPECT AND CT WITH SHUTTER CONTROLLED RADIONUCLIDE SOURCES AND PARALLEL COLLIMATOR GEOMETRY [J].
LARSSON, SA ;
KIMIAEI, S ;
RIBBE, T .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) :1117-1122
[8]  
TAN P, 1993, J NUCL MED, V34, P1752
[9]   NONUNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY [J].
TUNG, CH ;
GULLBERG, GT ;
ZENG, GL ;
CHRISTIAN, PE ;
DATZ, FL ;
MORGAN, HT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) :1134-1143