Central Corneal Thickness Measurements With Fourier-Domain Optical Coherence Tomography Versus Ultrasonic Pachymetry and Rotating Scheimpflug Camera

被引:92
作者
Ishibazawa, Akihiro [1 ]
Igarashi, Sho [1 ]
Hanada, Kazuomi [1 ]
Nagaoka, Taiji [1 ]
Ishiko, Satoshi [1 ]
Ito, Haruna [1 ]
Yoshida, Akitoshi [1 ]
机构
[1] Asahikawa Med Univ, Dept Ophthalmol, Asahikawa, Hokkaido 0788510, Japan
关键词
central corneal thickness; Fourier-domain optical coherence tomography; ultrasonic pachymetry; rotating Scheimpflug camera; ANTERIOR-CHAMBER DEPTH; OCULAR HYPERTENSION; ORBSCAN-II; REPRODUCIBILITY; AGREEMENT; EYES;
D O I
10.1097/ICO.0b013e3181d00800
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare the accuracy and repeatability of Fourier-domain optical coherence tomography (FD-OCT) with ultrasonic pachymetry (USP) and a rotating Scheimpflug camera for measuring the central corneal thickness (CCT). Methods: The CCT was measured in 30 subjects (30 normal corneas) by the same examiner using RTVue-100 FD-OCT with an anterior segment adaptor, Pentacam rotating Scheimpflug camera, and SP-2000 USP. Two examiners obtained one FD-OCT measurement from 10 eyes of 5 subjects to assess interexaminer reproducibility. Results: The mean CCT (+/- SD) measured by FD-OCT, USP, and the Pentacam were 530 +/- 33, 544 +/- 34, and 552 +/- 35 mu m, respectively. Significant correlations were found between FD-OCT and USP (r = 0.97; P < 0.0001), FD-OCT and Pentacam (r = 0.97; P < 0.0001), and USP and Pentacam (r = 0.96; P < 0.0001). Pairwise comparisons showed that the FD-OCT CCT measurement was significantly thinner than those of the other 2 methods (P < 0.001 for all comparisons). Regarding intraexaminer repeatability, the intraclass correlation coefficients ranged between 0.97 and 0.98. There was high repeatability of the CCT measurements with all methods. FD-OCT also had high interexaminer reproducibility (intraclass correlation coefficient = 0.98). Conclusions: RTVue-100 FD-OCT may be a useful alternative for measuring the CCT; however, it significantly underestimates the CCT compared with the USP and the Pentacam with slight differences. Although highly correlated, the measurements are not directly interchangeable in clinical practice.
引用
收藏
页码:615 / 619
页数:5
相关论文
共 26 条
[21]   Intra-observer and inter-observer reliability and reproducibility of slit-lamp-adapted optical coherence tomography for evaluation of anterior chamber depth and central corneal thickness [J].
Sandler, Shlomit F. ;
Zelefsky, Joseph R. ;
Dorairaj, Syril ;
Arthur, Stella N. ;
Ritch, Robert ;
Liebmann, Jeffrey M. .
OPHTHALMIC SURGERY LASERS & IMAGING, 2008, 39 (04) :299-303
[22]   Corneal thickness and relating factors in a popu lation-based study in japan: The tajimi study [J].
Tomidokoro, Atsuo ;
Araie, Makoto ;
Iwase, Aiko .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 144 (01) :152-154
[23]  
WARING GO, 1982, OPHTHALMOLOGY, V89, P531
[24]   Noncontact corneal pachymetry with slit lamp-adapted optical coherence tomography [J].
Wirbelauer, C ;
Scholz, C ;
Hoerauf, H ;
Pham, DT ;
Laqua, H ;
Birngruber, R .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2002, 133 (04) :444-450
[25]   Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography [J].
Wylegala, Edward ;
Teper, Slawornir ;
Nowinska, Anna K. ;
Milka, Michal ;
Dobrowolski, Dariusz .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (08) :1410-1414
[26]   Comparison of central corneal thickness measurements by visante anterior segment optical coherence tomography with ultrasound pachymetry [J].
Zhao, Paul S. ;
Wong, Tien Yin ;
Wong, Wan-Ling ;
Saw, Seang-Mei ;
Aung, Tin .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (06) :1047-1049