Numerical and Experimental Flow Analysis of the Wang-Zwische Double-Lumen Cannula

被引:25
作者
De Bartolo, Carmine [2 ]
Nigro, Alessandra [2 ]
Fragomeni, Gionata [3 ]
Colacino, Francesco M. [3 ]
Wang, Dongfang [1 ]
Jones, Cameron C.
Zwischenberger, Joseph
机构
[1] Univ Kentucky, Coll Med, Div Cardiothorac Surg, Chandler Med Ctr,Dept Surg, Lexington, KY 40536 USA
[2] Univ Calabria, Dept Mech Engn, I-87036 Arcavacata Di Rende, Italy
[3] Magna Graecia Univ Catanzaro, Dept Expt & Clin Med, Catanzaro, Italy
关键词
SHEAR-STRESS; BLOOD DAMAGE; HEMOLYSIS;
D O I
10.1097/MAT.0b013e31821c08bc
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An experimental and numerical analysis was performed for the Wang-Zwische double-lumen cannula (DLC) (Avalon Elite). The aim of this work was to provide insight for future improvement by characterizing the fluid dynamic behavior of the novel catheter with metrics often associated with blood trauma. Pressure and flow distributions were measured on a steady-flow rig using a 50% glycerol-water mixture by imposing a 2 L/min flow rate across the drainage and infusion lumens. The fluid was modeled as Newtonian with density of 1050 kg/m(3) and dynamic viscosity of 0.0035 kg/m.s. Reynolds numbers typical for transitional flow (Re = 2000-2500) were computed within the lumens because of the changing cross-sections of the cannula geometry. Numerical computations were performed using the steady three-dimensional Reynolds-averaged Navier-Stokes (RANS) equations and the low-Reynolds k-omega turbulence model. Discretization of governing equations was based on a cell-centered finite volume method. Numerical results correlated well with global performance of the cannula, allowing evaluation of the geometry toward potential blood trauma. Peak wall shear stress (WSS) in the drainage lumen was higher than that of infusion lumen, mainly due to the presence of side holes. Furthermore, recirculation regions were predicted in transition tubing to connectors of both the drainage and the infusion lumens because of adverse pressure gradients caused by the sudden enlargement of the cannula geometry. In this three-dimensional computational fluid dynamics (CFD) study, we observed higher peak WSS values for the drainage lumen, which may potentially cause blood trauma. Furthermore, recirculation regions were predicted in the proximity of the exit sections of both the infusion and drainage lumens, which may contribute to thrombosis formation. This study provides insight for future DLC modifications in minimizing cannula-induced blood trauma and thrombogenicity in long-term applications. ASAIO Journal 2011; 57:318-327.
引用
收藏
页码:318 / 327
页数:10
相关论文
共 14 条
[1]  
AHMED S, 1983, J BIOMECH, V16, P509
[2]  
COLACINO FM, 2009, ASAIO J, V55, P174
[3]   Numerical calculation of hemolysis levels in peripheral hemodialysis cannulas [J].
De Wachter, D ;
Verdonck, P .
ARTIFICIAL ORGANS, 2002, 26 (07) :576-582
[4]   Low Reynolds number turbulence modeling of blood flow in arterial stenoses [J].
Ghalichi, F ;
Deng, XY ;
De Champlain, A ;
Douville, Y ;
King, M ;
Guidoin, R .
BIORHEOLOGY, 1998, 35 (4-5) :281-294
[5]   ESTIMATION OF SHEAR STRESS-RELATED BLOOD DAMAGE IN HEART-VALVE PROSTHESES - INVITRO COMPARISON OF 25 AORTIC VALVES [J].
GIERSIEPEN, M ;
WURZINGER, LJ ;
OPITZ, R ;
REUL, H .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1990, 13 (05) :300-306
[6]   Numerical estimation of blood damage in artificial organs [J].
Goubergrits, L ;
Affeld, K .
ARTIFICIAL ORGANS, 2004, 28 (05) :499-507
[7]  
Grigioni M, 1999, J BIOMECH, V32, P1107, DOI 10.1016/S0021-9290(99)00063-9
[8]   Evaluation of computational models for hemolysis estimation [J].
Gu, L ;
Smith, WA .
ASAIO JOURNAL, 2005, 51 (03) :202-207
[9]   Using computational fluid dynamics to evaluate a novel venous cannula (Smart canula®) for use in cardiopulmonary bypass operating procedures [J].
Jegger, D. ;
Sundaram, S. ;
Shah, K. ;
Mallabiabarrena, I. ;
Mucciolo, G. ;
von Segesser, L. K. .
PERFUSION-UK, 2007, 22 (04) :257-265
[10]   RED BLOOD-CELL DAMAGE BY SHEAR-STRESS [J].
LEVERETT, LB ;
LYNCH, EC ;
ALFREY, CP ;
HELLUMS, JD .
BIOPHYSICAL JOURNAL, 1972, 12 (03) :257-&