IDENTIFYING ELDERLY WOMEN WITH OSTEOPOROSIS BY SPINAL DUAL X-RAY ABSORPTIOMETRY, CALCANEAL QUANTITATIVE ULTRASOUND AND SPINAL QUANTITATIVE COMPUTED TOMOGRAPHY: A COMPARATIVE STUDY

被引:21
作者
Edelmann-Schaefer, Bianca [1 ]
Berthold, Lars Daniel [2 ]
Stracke, Hilmar [3 ,4 ]
Luehrmann, Petra Maria [1 ,5 ]
Neuhaeuser-Berthold, Monika [1 ]
机构
[1] Univ Giessen, Inst Nutr Sci, D-35390 Giessen, Germany
[2] Hannover Med Sch, Inst Radiol, D-3000 Hannover, Germany
[3] Univ Hosp Giessen, Med Clin & Policlin 3, Giessen, Germany
[4] Marburg GmbH, Giessen, Germany
[5] Univ Educ, Dept Nutr Consumpt & Fash, Schwabisch Gmund, Germany
关键词
Osteoporosis; Diagnosis; Dual x-ray absorptiometry; Quantitative ultrasound; Quantitative computed tomography; T-score; BONE-MINERAL DENSITY; T-SCORE; RISK; BMD; THRESHOLDS; IDENTIFICATION; DISCORDANCE; PREVALENCE; PREDICTION; MANAGEMENT;
D O I
10.1016/j.ultrasmedbio.2010.10.003
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The ability of spinal dual x-ray absorptiometry (DXA), calcaneal quantitative ultrasound (QUS) and spinal quantitative computed tomography (QCT) to identify women with osteoporosis within the GISELA study was evaluated in 43 women, aged 62-87 years. Osteoporosis was defined as a T-score below or equal to -2.5 using DXA (femoral neck). To determine the performance of each method, the sensitivity, specificity and area under the curve (by means of a receiver operating characteristic [ROC] analysis) were calculated. The median T-scores from the measurements differed significantly (p < 0.0001). DXA (spine) identified 75% of women with osteoporosis; QUS and QCT identified 100%. The specificity was 89% for DXA (spine), 66% for QUS and 29% for QCT. ROC analysis showed that all three methods are qualified to identify women with osteoporosis; however, the different sensitivities and specificities of the methods, as well as the thresholds used for diagnosing osteoporosis have to be considered. (E-mail: monika.neuhaeuser-berthold@ernaehrung.uni-giessen.de) (C) 2011 Published by Elsevier Inc. on behalf of World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:29 / 36
页数:8
相关论文
共 49 条
[1]  
[Anonymous], WHO TECHNICAL REPORT
[2]   Apparent pre- and postmenopausal bone loss evaluated by DXA at different skeletal sites in women: The OFELY cohort [J].
Arlot, ME ;
SornayRendu, E ;
Garnero, P ;
VeyMarty, B ;
Delmas, PD .
JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (04) :683-690
[3]   Recalculation of the NHANES database SD improves T-score agreement and reduces osteoporosis prevalence [J].
Binkley, N ;
Kiebzak, GM ;
Lewiecki, EM ;
Krueger, D ;
Gangnon, RE ;
Miller, PD ;
Shepherd, JA ;
Drezner, MK .
JOURNAL OF BONE AND MINERAL RESEARCH, 2005, 20 (02) :195-201
[4]   A list of device-specific thresholds for the clinical interpretation of peripheral x-ray absorptiometry examinations [J].
Blake, GM ;
Chinn, DJ ;
Steel, SA ;
Patel, R ;
Panayiotou, E ;
Thorpe, J ;
Fordham, JN .
OSTEOPOROSIS INTERNATIONAL, 2005, 16 (12) :2149-2156
[5]  
Blake GM, 2005, CALCIFIED TISSUE INT, V76, P249, DOI 10.1007/s00223-004-0175-0
[6]   Dual X-ray absorptiometry - Clinical evaluation of a new cone-beam system [J].
Blake, GM ;
Knapp, KM ;
Fogelman, I .
CALCIFIED TISSUE INTERNATIONAL, 2005, 76 (02) :113-120
[7]   Technical principles of dual energy x-ray absorptiometry [J].
Blake, GM ;
Fogelman, I .
SEMINARS IN NUCLEAR MEDICINE, 1997, 27 (03) :210-228
[8]   DXA in vivo BMD methodology:: An erroneous and misleading research and clinical gauge of bone mineral status, bone fragility, and bone remodelling [J].
Bolotin, H. H. .
BONE, 2007, 41 (01) :138-154
[9]   Evaluation of error bounds on calcaneal speed of sound caused by surrounding soft tissue [J].
Chappard, C ;
Camus, E ;
Lefebvre, F ;
Guillot, G ;
Bittoun, J ;
Berger, G ;
Laugier, P .
JOURNAL OF CLINICAL DENSITOMETRY, 2000, 3 (02) :121-131
[10]   Influence of region of interest and bone size on calcaneal BMD: implications for the accuracy of quantitative ultrasound assessments at the calcaneus [J].
Cheng, S ;
Njeh, CF ;
Fan, B ;
Cheng, X ;
Hans, D ;
Wang, L ;
Fuerst, T ;
Genant, HK .