Reproducibility of SD-OCT-Based Ganglion Cell-Layer Thickness in Glaucoma Using Two Different Segmentation Algorithms

被引:17
作者
Garvin, Mona K. [1 ,2 ]
Lee, Kyungmoo [2 ]
Burns, Trudy L. [3 ]
Abramoff, Michael D. [1 ,2 ,4 ,5 ]
Sonka, Milan [2 ,4 ]
Kwon, Young H. [4 ]
机构
[1] Iowa City VA Hlth Care Syst, Ctr Prevent & Treatment Visual Loss, Iowa City, IA USA
[2] Univ Iowa, Dept Elect & Comp Engn, Iowa City, IA 52242 USA
[3] Univ Iowa, Dept Epidemiol, Iowa City, IA 52242 USA
[4] Univ Iowa, Dept Ophthalmol & Visual Sci, Iowa City, IA 52242 USA
[5] Univ Iowa, Dept Biomed Engn, Iowa City, IA 52242 USA
基金
美国国家卫生研究院;
关键词
ganglion cell layer; glaucoma; image segmentation; reproducibility; OPTICAL-COHERENCE-TOMOGRAPHY; NERVE-FIBER LAYER; DAMAGE; HEAD; EYES;
D O I
10.1167/iovs.13-12131
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To compare the reproducibility of spectral-domain optical coherence tomography (SD-OCT)-based ganglion cell-layer-plus-inner plexiform-layer (GCL+IPL) thickness measurements for glaucoma patients obtained using both a publicly available and a commercially available algorithm. METHODS. Macula SD-OCT volumes (200 x 200 x 1024 voxels, 6 x 6 x 2 mm(3)) were obtained prospectively from both eyes of patients with open-angle glaucoma or with suspected glaucoma on two separate visits within 4 months. The combined GCL+IPL thickness was computed for each SD-OCT volume within an elliptical annulus centered at the fovea, based on two algorithms: (1) a previously published graph-theoretical layer segmentation approach developed at the University of Iowa, and (2) a ganglion cell analysis module of version 6 of Cirrus software. The mean overall thickness of the elliptical annulus was computed as was the thickness within six sectors. For statistical analyses, eyes with an SD-OCT volume with low signal strength (<6), image acquisition errors, or errors in performing the commercial GCL+IPL analysis in at least one of the repeated acquisitions were excluded. RESULTS. Using 104 eyes (from 56 patients) with repeated measurements, we found the intraclass correlation coefficient for the overall elliptical annular GCL+IPL thickness to be 0.98 (95% confidence interval [CI]: 0.97-0.99) with the Iowa algorithm and 0.95 (95% CI: 0.93-0.97) with the Cirrus algorithm; the intervisit SDs were 1.55 mu m (Iowa) and 2.45 mu m (Cirrus); and the coefficients of variation were 2.2% (Iowa) and 3.5% (Cirrus), P < 0.0001. CONCLUSIONS. SD-OCT-based GCL+IPL thickness measurements in patients with early glaucoma are highly reproducible.
引用
收藏
页码:6998 / 7004
页数:7
相关论文
共 24 条
[21]   Three-Dimensional Analysis of Retinal Layer Texture: Identification of Fluid-Filled Regions in SD-OCT of the Macula [J].
Quellec, Gwenole ;
Lee, Kyungmoo ;
Dolejsi, Martin ;
Garvin, Mona K. ;
Abramoff, Michael D. ;
Sonka, Milan .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (06) :1321-1330
[22]   Detection of Macular Ganglion Cell Loss in Glaucoma by Fourier-Domain Optical Coherence Tomography [J].
Tan, Ou ;
Chopra, Vikas ;
Lu, Ake Tzu-Hui ;
Schuman, Joel S. ;
Ishikawa, Hiroshi ;
Wollstein, Gadi ;
Varma, Rohit ;
Huang, David .
OPHTHALMOLOGY, 2009, 116 (12) :2305-2314
[23]  
Zar JH, 1974, BIOSTAT ANAL, P103
[24]   Reproducibility of Macular, Retinal Nerve Fiber Layer, and ONH Measurements by OCT in Rhesus Monkeys: The Beijing Intracranial and Intraocular Pressure (iCOP) Study [J].
Zhang, Zheng ;
Yang, Diya ;
Sang, Jinhong ;
Hou, Ruowu ;
Liu, Kegao ;
Li, Zhen ;
Xie, Xiaobin ;
Jonas, Jost B. ;
Wang, Ningli .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2012, 53 (08) :4505-4509