Evaluation of alignment error due to a speed artifact in stereotactic ultrasound image guidance

被引:9
作者
Salter, Bill J. [1 ]
Wang, Brian [1 ]
Szegedi, Martin W. [1 ]
Rassiah-Szegedi, Prema [1 ]
Shrieve, Dennis C. [1 ]
Cheng, Roger [2 ]
Fuss, Martin [3 ]
机构
[1] Univ Utah, Huntsman Canc Inst, Dept Radiat Oncol, Salt Lake City, UT 84112 USA
[2] Univ Oklahoma, Hlth Sci Ctr, Dept Radiat Oncol, Norman, OK 73019 USA
[3] Oregon Hlth & Sci Univ, Dept Radiat Oncol, Portland, OR 97201 USA
关键词
D O I
10.1088/0031-9155/53/23/N03
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultrasound (US) image guidance systems used in radiotherapy are typically calibrated for soft tissue applications, thus introducing errors in depth-from-transducer representation when used in media with a different speed of sound propagation (e.g. fat). This error is commonly referred to as the speed artifact. In this study we utilized a standard US phantom to demonstrate the existence of the speed artifact when using a commercial US image guidance system to image through layers of simulated body fat, and we compared the results with calculated/predicted values. A general purpose US phantom (speed of sound (SOS) = 1540 m s(-1)) was imaged on a multi-slice CT scanner at a 0.625 mm slice thickness and 0.5 mm x 0.5 mm axial pixel size. Target-simulating wires inside the phantom were contoured and later transferred to the US guidance system. Layers of various thickness (1-8 cm) of commercially manufactured fat-simulating material (SOS = 1435 m s(-1)) were placed on top of the phantom to study the depth- related alignment error. In order to demonstrate that the speed artifact is not caused by adding additional layers on top of the phantom, we repeated these measurements in an identical setup using commercially manufactured tissue-simulating material (SOS = 1540 m s(-1)) for the top layers. For the fat-simulating material used in this study, we observed the magnitude of the depth- related alignment errors resulting from the speed artifact to be 0.7 mm cm(-1) of fat imaged through. The measured alignment errors caused by the speed artifact agreed with the calculated values within one standard deviation for all of the different thicknesses of fat-simulating material studied here. We demonstrated the depth- related alignment error due to the speed artifact when using US image guidance for radiation treatment alignment and note that the presence of fat causes the target to be aliased to a depth greater than it actually is. For typical US guidance systems in use today, this will lead to delivery of the high dose region at a position slightly posterior to the intended region for a supine patient. When possible, care should be taken to avoid imaging through a thick layer of fat for larger patients in US alignments or, if unavoidable, the spatial inaccuracies introduced by the artifact should be considered by the physician during the formulation of the treatment plan.
引用
收藏
页码:N437 / N445
页数:9
相关论文
共 12 条
  • [1] The impact of sound speed errors on medical ultrasound imaging
    Anderson, ME
    McKeag, MS
    Trahey, GE
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 107 (06) : 3540 - 3548
  • [2] Experience of ultrasound-based daily prostate localization
    Chandra, A
    Dong, L
    Huang, E
    Kuban, DA
    O'Neill, L
    Rosen, I
    Pollack, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (02): : 436 - 447
  • [3] Comparing computed tomography localization with daily ultrasound during image-guided radiation therapy for the treatment of prostate cancer: a prospective evaluation
    Feigenberg, Steven J.
    Paskalev, Kamen
    McNeeley, Shawn
    Horwitz, Eric M.
    Konski, Andre
    Wang, Lu
    Ma, Charlie
    Pollack, Alan
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (03): : 99 - 110
  • [4] Daily ultrasound-based image-guided targeting for radiotherapy of upper abdominal malignancies
    Fuss, M
    Salter, BJ
    Cavanaugh, SX
    Fuss, C
    Sadeghi, A
    Fuller, CD
    Ameduri, A
    Hevezi, JM
    Herman, TS
    Thomas, CR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (04): : 1245 - 1256
  • [5] Daily stereotactic ultrasound prostate targeting: Inter-user variability
    Fuss, M
    Cavanaugh, SX
    Fuss, C
    Cheek, DA
    Salter, BJ
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2003, 2 (02) : 161 - 169
  • [6] A comparison of daily CT localization to a daily ultrasound-based system in prostate cancer
    Lattanzi, J
    McNeeley, S
    Pinover, W
    Horwitz, E
    Das, I
    Schultheiss, TE
    Hanks, GE
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (04): : 719 - 725
  • [7] Implementation and utility of a daily ultrasound-based localization system with intensity-modulated radiotherapy for prostate cancer
    Morr, J
    DiPetrillo, T
    Tsai, JS
    Engler, M
    Wazer, DE
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1124 - 1129
  • [8] Sound speed correction in ultrasound imaging
    Napolitano, David
    Chou, Ching-Hua
    McLaughlin, Glen
    Ji, Ting-Lan
    Mo, Larry
    DeBusschere, Derek
    Steins, Robert
    [J]. ULTRASONICS, 2006, 44 (SUPPL.) : E43 - E46
  • [9] Rectal dose sparing with a balloon catheter and ultrasound localization in conformal radiation therapy for prostate cancer
    Patel, RR
    Orton, N
    Tomé, WA
    Chappell, R
    Ritter, MA
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 67 (03) : 285 - 294
  • [10] An assessment of PTV margin definitions for patients undergoing conformal 3D external beam radiation therapy for prostate cancer based on an analysis of 10,327 pretreatment daily ultrasound localizations
    Poli, M. Esmeralda Ramos
    Parker, William
    Patrocinio, Horacio
    Souhami, Luis
    Shenouda, George
    Campos, Leticia Lucente
    Podgorsak, Ervin B.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (05): : 1430 - 1437