Evaluation of Noninvasive Hemoglobin Monitoring in Surgical Critical Care Patients

被引:13
作者
Xu, Tim [1 ]
Yang, Ting [2 ]
Kim, Joseph B. [1 ]
Romig, Mark C. [1 ,2 ]
Sapirstein, Adam [1 ,2 ]
Winters, Bradford D. [1 ,2 ]
机构
[1] Johns Hopkins Univ, Sch Med, Dept Anesthesiol & Crit Care Med, Baltimore, MD 21205 USA
[2] Johns Hopkins Univ, Sch Med, Armstrong Inst Patient Safety & Qual, Baltimore, MD USA
关键词
hematology; medical devices; noninvasive hemoglobin monitoring; PULSE CO-OXIMETRY; CARDIAC-SURGERY; ACCURACY; HEMODILUTION; TRANSFUSION; THRESHOLD;
D O I
10.1097/CCM.0000000000001634
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To assess the clinical utility of noninvasive hemoglobin monitoring based on pulse cooximetry in the ICU setting. Design and Setting: A total of 358 surgical patients from a large urban, academic hospital had the noninvasive hemoglobin monitoring pulse cooximeter placed at admission to the ICU. Core and stat laboratory hemoglobin measurements were taken at the discretion of the clinicians, who were blinded to noninvasive hemoglobin monitoring values. Measurement and Main Results: There was a poor correlation between the 2,465 time-matched noninvasive hemoglobin monitoring and laboratory hemoglobin measurements (r(2) = 0.29). Bland-Altman analysis showed a positive bias of 1.0 g/dL and limits of agreement of -2.5 to 4.6 g/dL. Accuracy was best at laboratory values of 10.5-14.5 g/dL and least at laboratory values of 6.5-8 g/dL. At hemoglobin values that would ordinarily identify a patient as requiring a transfusion (< 8 g/dL), noninvasive hemoglobin monitoring consistently overestimated the patient's true hemoglobin. When sequential laboratory values declined below 8 g/dL (n = 102) and 7 g/dL (n = 13), the sensitivity and specificity of noninvasive hemoglobin monitoring at identifying these events were 27% and 7%, respectively. At a threshold of 8 g/dL, continuous noninvasive hemoglobin monitoring values reached the threshold before the labs in 45 of 102 instances (44%) and at 7 g/dL, noninvasive hemoglobin monitoring did so in three of 13 instances (23%). Noninvasive hemoglobin monitoring minus laboratory hemoglobin differences showed an intraclass correlation coefficient of 0.47 within individual patients. Longer length of stay and higher All Patient Refined Diagnostic-Related Groups severity of illness were associated with poor noninvasive hemoglobin monitoring accuracy. Conclusions: Although noninvasive hemoglobin monitoring technology holds promise, it is not yet an acceptable substitute for laboratory hemoglobin measurements. Noninvasive hemoglobin monitoring performs most poorly in the lower hemoglobin ranges that include commonly used transfusion trigger thresholds and is not consistent within individual patients. Further refinement of the signal acquisition and analysis algorithms and clinical reevaluation are needed.
引用
收藏
页码:E344 / E352
页数:9
相关论文
共 17 条
[1]   Complications related to blood transfusion in surgical patients:: data from the French national survey on anesthesia-related deaths [J].
Auroy, Yves ;
Lienhart, Andre ;
Pequignot, Francoise ;
Benhamou, Dan .
TRANSFUSION, 2007, 47 (02) :184S-189S
[2]   The Accuracy of Noninvasive Hemoglobin Measurement by Multiwavelength Pulse Oximetry After Cardiac Surgery [J].
Ba-Vinh Nguyen ;
Vincent, Jean-Louis ;
Nowak, Emmanuel ;
Coat, Michelle ;
Paleiron, Nicolas ;
Gouny, Pierre ;
Ould-Ahmed, Mehdi ;
Guillouet, Maite ;
Arvieux, Charles Christian ;
Gueret, Gildas .
ANESTHESIA AND ANALGESIA, 2011, 113 (05) :1052-1057
[3]   Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station [J].
Causey, Marlin Wayne ;
Miller, Seth ;
Foster, Andrew ;
Beekley, Alec ;
Zenger, David ;
Martin, Matthew .
AMERICAN JOURNAL OF SURGERY, 2011, 201 (05) :590-596
[4]   Precision of noninvasive hemoglobin-level measurement by pulse co-oximetry in patients admitted to intensive care units for severe gastrointestinal bleeds [J].
Coquin, Julien ;
Dewitte, Antoine ;
Le Manach, Yannick ;
Caujolle, Marie ;
Joannes-Boyau, Olivier ;
Fleureau, Catherine ;
Janvier, Gerard ;
Ouattara, Alexandre .
CRITICAL CARE MEDICINE, 2012, 40 (09) :2576-2582
[5]  
DeFrances Carol J, 2008, Natl Health Stat Report, P1
[6]   Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients [J].
Frasca, Denis ;
Dahyot-Fizelier, Claire ;
Catherine, Karen ;
Levrat, Quentin ;
Debaene, Bertrand ;
Mimoz, Olivier .
CRITICAL CARE MEDICINE, 2011, 39 (10) :2277-2282
[7]   Hematocrit Distribution and Tissue Oxygenation in Large Microcirculatory Networks [J].
Gould, Ian G. ;
Linninger, Andreas A. .
MICROCIRCULATION, 2015, 22 (01) :1-18
[8]   Transfusion Threshold of 7 g per Deciliter - The New Normal [J].
Hebert, Paul C. ;
Carson, Jeffrey L. .
NEW ENGLAND JOURNAL OF MEDICINE, 2014, 371 (15) :1459-1461
[9]   Lower versus Higher Hemoglobin Threshold for Transfusion in Septic Shock [J].
Holst, Lars B. ;
Haase, Nicolai ;
Wetterslev, Jorn ;
Wernerman, Jan ;
Guttormsen, Anne B. ;
Karlsson, Sari ;
Johansson, Par I. ;
Aneman, Anders ;
Vang, Marianne L. ;
Winding, Robert ;
Nebrich, Lars ;
Nibro, Helle L. ;
Rasmussen, Bodil S. ;
Lauridsen, Johnny R. M. ;
Nielsen, Jane S. ;
Oldner, Anders ;
Pettila, Ville ;
Cronhjort, Maria B. ;
Andersen, Lasse H. ;
Pedersen, Ulf G. ;
Reiter, Nanna ;
Wiis, Jorgen ;
White, Jonathan O. ;
Russell, Lene ;
Thornberg, Klaus J. ;
Hjortrup, Peter B. ;
Muller, Rasmus G. ;
Moller, Morten H. ;
Steensen, Morten ;
Tjader, Inga ;
Kilsand, Kristina ;
Odeberg-Wernerman, Suzanne ;
Sjobo, Brit ;
Bundgaard, Helle ;
Thyo, Maria A. ;
Lodahl, David ;
Maerkedahl, Rikke ;
Albeck, Carsten ;
Illum, Dorte ;
Kruse, Mary ;
Winkel, Per ;
Perner, Anders .
NEW ENGLAND JOURNAL OF MEDICINE, 2014, 371 (15) :1381-1391
[10]   Accuracy of Continuous Noninvasive Hemoglobin Monitoring: A Systematic Review and Meta-Analysis [J].
Kim, Sang-Hyun ;
Lilot, Marc ;
Murphy, Linda Suk-Ling ;
Sidhu, Kulraj S. ;
Yu, Zhaoxia ;
Rinehart, Joseph ;
Cannesson, Maxime .
ANESTHESIA AND ANALGESIA, 2014, 119 (02) :332-346