Full characterization of a compact 90Sr/90Y beta source for TID radiation testing

被引:5
|
作者
Di Capua, F. [1 ]
Campajola, L. [1 ]
Casolaro, P. [1 ]
Campajola, M. [1 ]
Aloisio, A. [1 ]
Lucaroni, A. [2 ]
Furano, G. [3 ]
Menicucci, A. [4 ]
Di Mascio, S. [4 ]
Malatesta, F. [5 ]
Ottavi, M. [5 ]
机构
[1] Univ Naples Federico II, Dept Phys E Pancini, Naples, Italy
[2] ARTE Srl, Taverne Di Corciano, PG, Italy
[3] European Space Agcy, ESTEC, Postbus 299, NL-2200 AG Noordwijk, Netherlands
[4] Delft Univ Technol, Fac Aerosp Engn, Space Syst Engn, Delft, Netherlands
[5] Univ Roma Tor Vergata, Dept Elect Engn, Rome, Italy
关键词
Total Ionizing Dose; Radiation hardness assurance; Sr-90/Y-90 beta source; COTS;
D O I
10.1016/j.asr.2019.01.037
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
A new methodology for Total Ionizing Dose (TID) tests is proposed. It is based on the employment of an on-chip Sr-90/Y-90 beta source as alternative to standard methods such as Co-60 gamma rays and electrons from LINAC. The use of a compact beta source for TID tests has several advantages. In particular, the irradiation of devices with more than one radiation source results in a better representation of the complex space radiation environment composed of several types, energies and dose-rates. In addition, the use of an easy handling beta source allows the irradiation of electronic devices without any damage to other auxiliary circuit. In this work, Sr-90/Y-90 beta source dosimetry and related radiation field characteristics are discussed in depth. In order to validate the proposed source for TID tests, a rather complex device such as the "SPC56EL70L5" microcontroller from ST-Microelectronics was exposed to Sr-90/Y-90 beta rays. The results of this test were compared to that of a previous test of another sample from the same lot with a standard gamma Co-60 source. The electronic performances following the two irradiations have been found to be in excellent agreement, by demonstrating therefore the validity of the proposed beta source for TID tests. (C) 2019 COSPAR. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:3249 / 3257
页数:9
相关论文
共 50 条
  • [1] TID testing of plastic encapsulated microcircuits with 90Sr/90Y source
    Thuesen, GG
    Jorgensen, JL
    Guldager, PB
    PROCEEDINGS OF THE 7TH EUROPEAN CONFERENCE ON RADIATION AND ITS EFFECTS ON COMPONENTS AND SYSTEMS, 2004, 536 : 127 - 129
  • [2] Optimization of a 90Sr/90Y radiation source train stepping for intravascular brachytherapy
    Miften, MM
    Das, SK
    Shafman, TD
    Marks, LB
    MEDICAL PHYSICS, 2002, 29 (12) : 2891 - 2896
  • [3] Dosimetric characterization of a new 90Sr/90Y source with balloon for intravascular brachytherapy
    Wang, R
    Li, XA
    Lobdell, J
    MEDICAL PHYSICS, 2002, 29 (06) : 1318 - 1319
  • [4] Calibration and validation of a quality assurance system for 90Sr/90Y radiation source trains
    Rosenthal, P
    Weber, W
    Förster, A
    Orth, O
    Köhler, B
    Seiler, F
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (05): : 573 - 585
  • [5] Implementation of energy reduced 90Sr/90Y radiation fields, or: Propagation of beta radiation, a case study
    Behrens, R.
    JOURNAL OF INSTRUMENTATION, 2024, 19 (07):
  • [6] Energy-reduced beta radiation fields from 90Sr/90Y for the BSS 2
    Behrens, R.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (05)
  • [7] A NOVEL ELECTROCHEMICAL 90Sr/90Y GENERATOR
    Comor, J.
    Beyer, G.
    JOURNAL OF LABELLED COMPOUNDS & RADIOPHARMACEUTICALS, 2009, 52 : S251 - S251
  • [8] Coprecipitation of 90Sr and 90Y with Lead Phosphate
    V. P. Medvedev
    A. O. Merkushkin
    Radiochemistry, 2002, 44 (6) : 571 - 575
  • [9] Monte Carlo dose characterization of a new 90Sr/90Y source with balloon for intravascular brachytherapy
    Wang, RQ
    Li, XA
    Lobdell, J
    MEDICAL PHYSICS, 2003, 30 (01) : 27 - 33
  • [10] A mini X-ray generator as an alternative to a 90Sr/90Y beta source in luminescence dating
    Andersen, CE
    Botter-Jensen, L
    Murray, AS
    RADIATION MEASUREMENTS, 2003, 37 (4-5) : 557 - 561