Long-term follow-up of transvenous defibrillation leads - High incidence of fracture in coaxial polyurethane lead

被引:18
作者
Kitamura, S
Satomi, K
Kurita, T
Shimizu, W
Suyama, K
Aihara, N
Niwaya, K
Kobayashi, J
Kamakura, S
机构
[1] Natl Cardiovasc Ctr, Dept Cardiovasc Med, Suita, Osaka 5658565, Japan
[2] Natl Cardiovasc Ctr, Dept Cardiovasc Surg, Suita, Osaka 5658565, Japan
关键词
follow-up; implantable cardioverter-defibrillator; lead failure; polyurethane;
D O I
10.1253/circj.70.273
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background As a result of longer follow-up after implantation of cardioverter defibrillators (ICD), fatigue of the leads has become a concern. The aim of this study was to determine the incidence and clinical presentation of ICD lead failures. Methods and Results The Study Population consisted of 241 patients with 249 ICD leads Who underwent implantation of an ICD with a transvenous lead system. After device implantation, the patients were routinely followed up every 4 months. Five lead failures (2.0%) Occurred as an oversensing of artifact during the follow-up period (2.6 +/- 2.1 years); 4 of those 5 patients received inappropriate shocks and I case of lead failure was identified in a patient with frequent episodes of non-sustained ventricular fibrillation. In particular, the right ventricular polyurethane transvenous lead in the Medtronic model 6936 failed in 4 (13%) of 31 cases. Percutaneous lead extraction was not available in all cases, so an additional ICD lead was inserted through the same site of the subclavian vein. Conclusions Lead failures may occur 5 years after ICD implantation and polyurethane leads have an especially high incidence of failure. However, there were no follow-up parameters observed that predicted lead failures.
引用
收藏
页码:273 / 277
页数:5
相关论文
共 11 条
[1]   Insulation lead failure: Is it a matter of insulation coating, venous approach, or both? [J].
Antonelli, D ;
Rosenfeld, T ;
Freedberg, NA ;
Palma, E ;
Gross, JN ;
Furman, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (02) :418-421
[2]   Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure [J].
Bardy, GH ;
Lee, KL ;
Mark, DB ;
Poole, JE ;
Packer, DL ;
Boineau, R ;
Domanski, M ;
Troutman, C ;
Anderson, J ;
Johnson, G ;
McNulty, SE ;
Clapp-Channing, N ;
Davidson-Ray, LD ;
Fraulo, ES ;
Fishbein, DP ;
Luceri, RM ;
Ip, JH .
NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (03) :225-237
[3]   Transvenous defibrillation leads: High incidence of failure during long-term follow-up [J].
Dorwarth, U ;
Frey, B ;
Dugas, M ;
Matis, T ;
Fiek, M ;
Schmoeckel, M ;
Remp, T ;
Durchlaub, I ;
Gerth, A ;
Steinbeck, G ;
Hoffmann, E .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2003, 14 (01) :38-43
[4]   Detection and management of an implantable cardioverter defibrillator lead failure - Incidence and clinical implications [J].
Ellenbogen, KA ;
Wood, MA ;
Shepard, RK ;
Clemo, HF ;
Vaughn, T ;
Holloman, K ;
Dow, M ;
Leffler, J ;
Abeyratne, A ;
Verness, D .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (01) :73-80
[5]   Long-term structural failure of coaxial polyurethane implantable cardioverter defibrillator leads [J].
Hauser, RG ;
Cannom, D ;
Hayes, DL ;
Parsonnet, V ;
Hayes, J ;
Ratliff, N ;
Tyers, GFO ;
Epstein, AE ;
Vlay, SC ;
Furman, S ;
Gross, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (06) :879-882
[6]   MULTICENTER EXPERIENCE WITH A BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD [J].
HAYES, DL ;
GRAHAM, KJ ;
IRWIN, M ;
VIDAILLET, H ;
DISLER, G ;
SWEESY, M ;
KINCAID, D ;
OSBORN, MJ ;
SUMAN, VJ ;
NEUBAUER, SA ;
SEEBANDT, M ;
KALLINEN, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05) :999-1004
[7]  
McAnulty J, 1997, NEW ENGL J MED, V337, P1576
[8]   Fracture of the subcutaneous patch electrode in a patient with an implanted cardioverter-defibrillator [J].
Milasinovic, G ;
Jelic, V ;
Savic, D ;
Pavlovic, SU ;
Velinovic, M .
CIRCULATION JOURNAL, 2005, 69 (01) :116-118
[9]   Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction [J].
Moss, AJ ;
Zareba, W ;
Hall, WJ ;
Klein, H ;
Wilber, DJ ;
Cannom, DS ;
Daubert, JP ;
Higgins, SL ;
Brown, MW ;
Andrews, ML .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 346 (12) :877-883
[10]   High incidence of device-related and lead-related complications in the dual-chamber implantable cardioverter defibrillator compared with the single-chamber version [J].
Takahashi, T ;
Bhandari, AK ;
Watanuki, M ;
Cannom, DS ;
Sakurada, H ;
Hiraoka, M .
CIRCULATION JOURNAL, 2002, 66 (08) :746-750