A simplified method of four-dimensional dose accumulation using the mean patient density representation

被引:50
作者
Glide-Hurst, Carri K. [1 ]
Hugo, Geoffrey D. [1 ]
Liang, Jian [1 ]
Yan, Di [1 ]
机构
[1] William Beaumont Hosp, Royal Oak, MI 48073 USA
关键词
cancer; computerised tomography; dosimetry; image motion analysis; image registration; lung; medical image processing; phantoms; pneumodynamics; radiation therapy; tumours;
D O I
10.1118/1.3002304
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this work was to demonstrate, both in phantom and patient, the feasibility of using an average 4DCT image set (AVG-CT) for 4D cumulative dose estimation. A series of 4DCT numerical phantoms and corresponding AVG-CTs were generated. For full 4D dose summation, static dose was calculated on each phase and cumulative dose was determined by combining each phase's static dose distribution with known tumor displacement. The AVG-CT cumulative dose was calculated similarly, although the same AVG-CT static dose distribution was used for all phases (i.e., tumor displacements). Four lung cancer cases were also evaluated for stereotactic body radiotherapy and conformal treatments; however, deformable image registration of the 4DCTs was used to generate the displacement vector fields (DVFs) describing patient-specific motion. Dose discrepancy between full 4D summation and AVG-CT approach was calculated and compared. For all phantoms, AVG-CT approximation yielded slightly higher cumulative doses compared to full 4D summation, with dose discrepancy increasing with increased tumor excursion. In vivo, using the AVG-CT coupled with deformable registration yielded clinically insignificant differences for all GTV parameters including the minimum, mean, maximum, dose to 99% of target, and dose to 1% of target. Furthermore, analysis of the spinal cord, esophagus, and heart revealed negligible differences in major dosimetric indices and dose coverage between the two dose calculation techniques. Simplifying 4D dose accumulation via the AVG-CT, while fully accounting for tumor deformation due to respiratory motion, has been validated, thereby, introducing the potential to streamline the use of 4D dose calculations in clinical practice, particularly for adaptive planning purposes.
引用
收藏
页码:5269 / 5277
页数:9
相关论文
共 30 条
  • [1] Dose calculations accounting for breathing motion in stereotactic lung radiotherapy based on 4D-CT and the internal target volume
    Admiraal, Marjan A.
    Schuring, Danny
    Hurkmans, Coen W.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) : 55 - 60
  • [2] Evaluating the influence of setup uncertainties on treatment planning for focal liver tumors
    Balter, JM
    Brock, KK
    Lam, KL
    Tatro, D
    Dawson, LA
    McShan, D
    Ten Haken, RK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : 610 - 614
  • [3] A fluence-convolution method to calculate radiation therapy dose distributions that incorporate random set-up error
    Beckham, WA
    Keall, PJ
    Siebers, JV
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (19) : 3465 - 3473
  • [4] Assessment of gross tumor volume regression and motion changes during radiotherapy for non-small-cell lung cancer as measured by four-dimensional computed tomography
    Britton, Keith R.
    Starkschall, George
    Tucker, Susan L.
    Pan, Tinsu
    Nelson, Christopher
    Chang, Joe Y.
    Cox, James D.
    Mohan, Radhe
    Komaki, Ritsuko
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (04): : 1036 - 1046
  • [5] A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study
    Chetty, IJ
    Rosu, M
    Tyagi, N
    Marsh, LH
    McShan, DL
    Balter, JM
    Fraass, BA
    Ten Haken, RK
    [J]. MEDICAL PHYSICS, 2003, 30 (07) : 1776 - 1780
  • [6] Tumour delineation and cumulative dose computation in radiotherapy based on deformable registration of respiratory correlated CT images of lung cancer patients
    de Xivry, Jonathan Orban
    Janssens, Guillaume
    Bosmans, Geert
    De Craene, Mathieu
    Dekker, Andre
    Buijsen, Jeroen
    van Baardwijk, Angela
    De Ruysscher, Dirk
    Macq, Benoit
    Lambin, Philippe
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 85 (02) : 232 - 238
  • [7] Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer
    Erridge, SC
    Seppenwoolde, Y
    Muller, SH
    van Herk, M
    De Jaeger, K
    Belderbos, JSA
    Boersma, LJ
    Lebesque, JV
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 66 (01) : 75 - 85
  • [8] Estimation of the delivered patient dose in lung IMRT treatment based on deformable registration of 4D-CT data and Monte Carlo simulations
    Flampouri, Stella
    Jiang, Steve B.
    Sharp, Greg C.
    Wolfgang, John
    Patel, Abhijit A.
    Choi, Noah C.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (11) : 2763 - 2779
  • [9] Respiration-correlated spiral CT: A method of measuring respiratory-induced anatomic motion for radiation treatment planning
    Ford, EC
    Mageras, GS
    Yorke, E
    Ling, CC
    [J]. MEDICAL PHYSICS, 2003, 30 (01) : 88 - 97
  • [10] Four-dimensional treatment planning for stereotactic body radiotherapy
    Guckenberger, Matthias
    Wilbert, Juergen
    Krieger, Thomas
    Richter, Anne
    Baier, Kurt
    Meyer, Juergen
    Flentje, Michael
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (01): : 276 - 285