Development of a low noise readout ASIC for CZT detectors for gamma-ray spectroscopy applications

被引:6
|
作者
Luo, J. [1 ,2 ]
Deng, Z. [1 ,2 ]
Wang, G. [3 ]
Li, H. [3 ]
Liu, Y. [1 ,2 ]
机构
[1] Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China
[2] Tsinghua Univ, Key Lab Particle & Radiat Imaging, Minist Educ, Beijing 100084, Peoples R China
[3] Beijing Hamamatsu Photon Tech Co Ltd, Beijing 100070, Peoples R China
来源
JOURNAL OF INSTRUMENTATION | 2012年 / 7卷
关键词
VLSI circuits; Analogue electronic circuits; Front-end electronics for detector readout;
D O I
10.1088/1748-0221/7/08/P08030
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A multi-channel readout ASIC for pixelated CZT detectors has been developed for gamma-ray spectroscopy applications. Each channel consists of a low noise dual-stage charge sensitive amplifier (CSA), a CR-(RC)(4) semi-Gaussian shaper and a class-AB output buffer. The equivalent noise charge (ENC) of input PMOS transistor is optimized for 5 pF input capacitance and 1 mu s peaking time using g(m)/I-D design methodology. The gain can be adjusted from 100 mV/fC to 400 mV/fC and the peaking time can be adjusted from 1 mu s to 4 mu s. A 16-channel chip has been designed and fabricated in 0.35 mu m m 2P4M CMOS technology. The test results show that the chip works well and fully satisfies the design specifications. The ENC was measured to be 72 e + 26 e/pF at 1 mu s peaking time and 86 e + 20 e/pF at 4 mu s peaking time. The non-uniformity of the channel gain and ENC was less than +/- 12% and +/- 11% respectively for 16 channels in one chip. The chip was also tested with a pixelated CZT detector at room temperature. The measured energy resolution at 59.5 keV photopeak of Am-241 and (122)keV photopeak of Co-57 were 4.5% FWHM and 2.8% FWHM for the central area pixels, respectively.
引用
收藏
页数:12
相关论文
共 13 条
  • [1] Development of a pixel readout ASIC for CZT detectors for spectral X-ray photon-counting imaging applications
    Wang, X.
    Deng, Z.
    Liu, W.
    Liu, Y.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (01):
  • [2] Development of a low-noise readout ASIC for Silicon Drift Detectors in high energy resolution X-ray spectrometry
    Atkin, E.
    Levin, V.
    Malankin, E.
    Shumikhin, V.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [3] Development of a detector based on Silicon Drift Detectors for gamma-ray spectroscopy and imaging applications
    Busca, P.
    Butt, A. D.
    Fiorini, C.
    Marone, A.
    Occhipinti, M.
    Peloso, R.
    Quaglia, R.
    Bombelli, L.
    Giacomini, G.
    Piemonte, C.
    Camera, F.
    Giaz, A.
    Million, B.
    Nelms, N.
    Shortt, B.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [4] A low-power CMOS ASIC for X-ray Silicon Drift Detectors low-noise pulse processing
    Ahangarianabhari, M.
    Bertuccio, G.
    Macera, D.
    Malcovati, P.
    Grassi, M.
    Rashevsky, A.
    Rashevskaya, I.
    Vacchi, A.
    Zampa, G.
    Zampa, N.
    Fuschino, F.
    Evangelista, Y.
    Campana, R.
    Labanti, C.
    Feroci, M.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [5] Measurements of low noise 64 channel counting ASIC for Si and CdTe strip detectors
    Kachel, M.
    Grybos, P.
    Szczygiel, R.
    Takeyoshi, T.
    JOURNAL OF INSTRUMENTATION, 2011, 6
  • [6] Characterisation of the MUSIC ASIC for large-area silicon photomultipliers for gamma-ray astronomy
    De Angelis, Nicolas
    Gascon, David
    Gomez, Sergio
    Heller, Matthieu
    Montaruli, Teresa
    Nagai, Andrii
    JOURNAL OF INSTRUMENTATION, 2023, 18 (01)
  • [7] Low-noise analog readout channel for SDD in X-ray spectrometry
    Atkin, E.
    Gusev, A.
    Krivchenko, A.
    Levin, V.
    Malankin, E.
    Normanov, D.
    Rotin, A.
    Sagdiev, I.
    Samsonov, V.
    JOURNAL OF INSTRUMENTATION, 2016, 11
  • [8] Design of MIRA, a low-noise pixelated ASIC for the readout of micro-channel plates
    Fabbrica, E.
    Carminati, M.
    Butta, D.
    Uslenghi, M.
    Fiorini, M.
    Corso, A. J.
    Pelizzo, M. G.
    Fiorini, C.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (01):
  • [9] Monolithic arrays of SDDs and low noise CMOS readout for X-ray spectroscopy measurements in nuclear physics experiments
    Quaglia, R.
    Fiorini, C.
    Bombelli, L.
    Butt, A. D.
    Giacomini, G.
    Ficorella, F.
    Picciotto, A.
    Piemonte, C.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [10] Design and implementation of a compact analog Constant Fraction Discriminator for high-resolution timing in gamma-ray spectroscopy
    Wiebusch, Michael
    JOURNAL OF INSTRUMENTATION, 2024, 19 (12):