OPTIMISATION OF PATIENT DOSE AND IMAGE QUALITY IN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY: A PHANTOM-BASED EVALUATION

被引:5
作者
Hadjiconstanti, Anastasia C. [1 ]
Messaris, Gerasimos A. T. [1 ]
Thomopoulos, Konstantinos C. [2 ]
Solomou, Aikaterini G. [3 ]
Panayiotakis, George S. [1 ]
机构
[1] Univ Patras, Sch Med, Dept Med Phys, Patras 26504, Greece
[2] Univ Patras, Sch Med, Dept Gastroenterol, Patras 26504, Greece
[3] Univ Patras, Sch Med, Dept Radiol, Patras 26504, Greece
关键词
PEDIATRIC INTERVENTIONAL CARDIOLOGY; FLAT DETECTOR; NATIONAL-SURVEY; RADIATION; INTENSIFIER; SYSTEM; ERCP; CHOLANGIOGRAPHY; EXPOSURE; ADULT;
D O I
10.1093/rpd/ncw276
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A phantom-based study is presented aiming to optimise patient dose and image quality (IQ) in endoscopic retrograde cholangiopancreatography procedures, utilising a fluoroscopy system equipped with a flat panel detector. The patient thickness was simulated with various polymethyl methacrylate slabs, whilst IQ was evaluated using the Leeds test object. The main factors evaluated were phantom thickness, distance between phantom and detector, field of view and pulse rate. For all these factors, the dosemetric indices, entrance surface air kerma (ESAK) rate and ESAK per pulse, as well as the IQ parameters, signal-to-noise ratio and high contrast spatial resolution, were measured. Based on these measurements, the figure of merit (FOM) was estimated. The FOM and ESAK rate values indicated the optimum combination of the factors evaluated which could provide adequate clinical information, assuring minimum patient dose.
引用
收藏
页码:118 / 123
页数:6
相关论文
共 36 条
[1]  
Alzimami K, 2013, GASTROENT RES PRACT, V2013, P1
[2]  
[Anonymous], 2005, 74 ICRU
[3]  
[Anonymous], 2007, The Recommendations of the International Commission on Radiological Protection, V103
[4]   Radiation dose to patients during endoscopic retrograde cholangiopancreatography [J].
Boix, Jaume ;
Lorenzo-Zuniga, Vicente .
WORLD JOURNAL OF GASTROINTESTINAL ENDOSCOPY, 2011, 3 (07) :140-144
[5]   Medical progress: Pancreatic and biliary endoscopy [J].
Brugge, WR ;
Van Dam, J .
NEW ENGLAND JOURNAL OF MEDICINE, 1999, 341 (24) :1808-1816
[6]   Patient and staff exposure during endoscopic retrograde cholangiopancreatography [J].
Buls, N ;
Pages, J ;
Mana, F ;
Osteaux, M .
BRITISH JOURNAL OF RADIOLOGY, 2002, 75 (893) :435-443
[7]   Overview of the role of ERCP in the management of diseases of the biliary tract and the pancreas [J].
Carr-Locke, DL .
GASTROINTESTINAL ENDOSCOPY, 2002, 56 (06) :S157-S160
[8]   Radiation protection in digestive endoscopy: European Society of Digestive Endoscopy (ESGE) Guideline [J].
Dumonceau, J. -M. ;
Garcia-Fernandez, F. J. ;
Verdun, F. R. ;
Carinou, E. ;
Donadille, L. ;
Damilakis, J. ;
Mouzas, I. ;
Paraskeva, K. ;
Ruiz-Lopez, N. ;
Struelens, L. ;
Tsapaki, V. ;
Vanhavere, F. ;
Valatas, V. ;
Sans-Merce, M. .
ENDOSCOPY, 2012, 44 (04) :408-421
[9]   PATIENT DOSE DURING THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE [J].
Hadjiconstanti, Anastasia C. ;
Messaris, Gerasimos A. T. ;
Thomopoulos, Konstantinos C. ;
Panayiotakis, George S. .
RADIATION PROTECTION DOSIMETRY, 2017, 173 (04) :380-382
[10]  
International Atomic Energy Agency, 2007, Dosimetry In Diagnostic Radiology: An International Code Of PractICE