Biomechanical influences of magnetic resonance imaging on the SOUNDTEC Direct System implant

被引:7
作者
Dyer, RK
Dormer, KJ
Hough, JVD
Nakmali, U
Wickersham, R
机构
[1] Hough Ear Inst, Oklahoma City, OK 73112 USA
[2] Univ Oklahoma, Hlth Sci Ctr, Oklahoma City, OK USA
[3] SOUNDTEC Inc, Oklahoma City, OK USA
关键词
D O I
10.1067/mhn.2002.129895
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
OBJECTIVE. The purpose of this study was to measure the forces experienced by the SOUNDTEC Direct System magnetic implant during 0.3-T MRI. STUDY DESIGN: Torsional and linear forces imposed on 8 implants were measured by using calibrated neurologic Von Frey hairs and were compared with finite-element analysis predictions and the forces required to separate the incudostapedial joints of unpreserved temporal bones. An implanted embalmed autopsy specimen was also examined before and after 1.5-T MRI. RESULTS: Peak linear force at the orifice of the MRI core measured 0.51g, (+/-0.2 SD). Maximum torque occurred at the MRI core center and measured 11.4g-cm (+/-1.2 SD). The mean torque required to separate the incudostapedial joints of 12 unpreserved temporal bones was 33.8g-cm (+/-20.4 SD). The autopsy specimen sustained a 1.5-T MRI scan without disruption of the ossicular chain or explantation. CONCLUSIONS., Physical and mechanical testing of the SOUNDTEC implant indicates that the structural integrity of the ossicles will be maintained during 0.3-T MRI of the human head.
引用
收藏
页码:520 / 530
页数:11
相关论文
共 16 条
[1]   FURTHER-STUDIES ON THE EFFECTS OF MAGNETIC-RESONANCE-IMAGING FIELDS ON MIDDLE-EAR IMPLANTS [J].
APPLEBAUM, EL ;
VALVASSORI, GE .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1990, 99 (10) :801-804
[2]  
CHENG D, 1989, FIELD WAVE ELECTROMA, P286
[3]  
DORMER KJ, 1981, LARYNGOSCOPE, V91, P1812
[4]   Three-dimensional modeling of middle ear biomechanics and its applications [J].
Gan, RZ ;
Sun, QL ;
Dyer, RK ;
Chang, KH ;
Dormer, KJ .
OTOLOGY & NEUROTOLOGY, 2002, 23 (03) :271-280
[5]   IMPLANTABLE BONE CONDUCTION HEARING DEVICE - AUDIANT BONE CONDUCTOR - UPDATE ON OUR EXPERIENCES [J].
HOUGH, J ;
HIMELICK, T ;
JOHNSON, B .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1986, 95 (05) :498-504
[6]   Semi-implantable electromagnetic middle ear hearing device for moderate to severe sensorineural hearing loss [J].
Hough, JVD ;
Dyer, RK ;
Matthews, P ;
Wood, MW .
OTOLARYNGOLOGIC CLINICS OF NORTH AMERICA, 2001, 34 (02) :401-+
[7]   Early clinical results: SOUNDTEC implantable hearing device phase II study [J].
Hough, JVD ;
Dyer, RK ;
Matthews, P ;
Wood, MW .
LARYNGOSCOPE, 2001, 111 (01) :1-8
[8]  
Hunyadi S, 1997, AM J OTOL, V18, P328
[9]   COCHLEAR IMPLANTS AS A CONTRAINDICATION TO MAGNETIC-RESONANCE-IMAGING [J].
PORTNOY, WM ;
MATTUCCI, K .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1991, 100 (03) :195-197
[10]   MR IMAGING AND BIOMEDICAL IMPLANTS, MATERIALS, AND DEVICES - AN UPDATED REVIEW [J].
SHELLOCK, FG ;
CURTIS, JS .
RADIOLOGY, 1991, 180 (02) :541-550