Study of electron avalanche multiplication in gaseous argon detectors at low pressure and operating at cryogenic temperature

被引:3
作者
Brunetti, R
Calligarich, E
Cambiaghi, M
De Vecchi, C
Dolfini, R
Menegolli, A
Prata, M
Prata, MC
Raselli, GL
Rossella, M
Vignoli, C
机构
[1] Ist Nazl Fis Nucl, Sez Pavia, I-27100 Pavia, Italy
[2] Univ Pavia, Dipartimento Fis Nucl & Teor, I-27100 Pavia, Italy
关键词
photodetectors; gas avalanche multiplication; cryogenic detectors; liquid argon TPC;
D O I
10.1016/j.nima.2005.12.133
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We are developing a photodetector sensitive to liquid noble gas scintillation light which is able to operate down to liquid nitrogen temperature (77 K). It has to be a simple, cheap and compact device, with a photocathode surface comparable to that of photomultipliers and it has to be able to produce fast signals for timing and triggering purposes. We studied the possibility of coupling a standard photocathode to a suitable electron multiplication system in gas; this would offer a cheap alternative to the use of photomultipliers. Gas avalanche multiplication at cryogenic temperatures and low pressure was observed in a small prototype chamber. Gains of the order of 10(3) were obtained in pure argon, while gains up to 10(5) were achieved using argon/ methane mixture in cryogenic environment. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:511 / 515
页数:5
相关论文
共 12 条
  • [11] Sauli F., 1977, Principles of operation of multiwire proportional and drift chambers
  • [12] SHARMA A, 1998, SLACJOURNALICFA163