Voltage pulses change neural interface properties and improve unit recordings with chronically implanted microelectrodes

被引:97
作者
Otto, KJ
Johnson, MD
Kipke, DR
机构
[1] Univ Michigan, Kresge Hearing Res Inst, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Biomed Engn, Ann Arbor, MI 48109 USA
基金
美国国家科学基金会;
关键词
brain-machine interface; chronic recording; iridium; neuroprosthesis; silicon;
D O I
10.1109/TBME.2005.862530
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Current neuroprosthetic systems based on electrophysiological recording have an extended, yet finite working lifetime. Some posited lifetime-extension solutions involve improving device biocompatibility or suppressing host immune responses. Our objective was to test an alternative solution comprised of applying a voltage pulse to a microelectrode site, herein termed "rejuvenation:'' Previously, investigators have reported preliminary electrophysiological results by utilizing a similar voltage pulse. In this study we sought to further explore this phenomenon via two methods: 1) electrophysiology; 2) an equivalent circuit model applied to impedance spectroscopy data. The experiments were conducted via chronically implanted silicon-substrate iridium microelectrode arrays in the rat cortex. Rejuvenation voltages resulted in increased unit recording signal-to-noise ratios (10% +/- 2%), with a maximal increase of 195% from 3.74 to 11.02. Rejuvenation also reduced the electrode site impedances at 1 kHz (67% +/- 2%). Neither the impedance nor recording properties of the electrodes changed on neighboring microelectrode sites that were not rejuvenated. In the equivalent circuit model, we found a transient increase in conductivity, the majority of which corresponded to a decrease in the tissue resistance component (44% +/- 7%). These findings suggest that rejuvenation may be an intervention strategy to prolong the functional lifetime of chronically implanted microelectrodes.
引用
收藏
页码:333 / 340
页数:8
相关论文
共 36 条
[21]   A METHOD FOR PNEUMATICALLY INSERTING AN ARRAY OF PENETRATING ELECTRODES INTO CORTICAL TISSUE [J].
ROUSCHE, PJ ;
NORMANN, RA .
ANNALS OF BIOMEDICAL ENGINEERING, 1992, 20 (04) :413-422
[22]   ORGANIZATION OF AUDITORY-CORTEX IN THE ALBINO-RAT - SOUND FREQUENCY [J].
SALLY, SL ;
KELLY, JB .
JOURNAL OF NEUROPHYSIOLOGY, 1988, 59 (05) :1627-1638
[23]  
SANES JN, 1990, EXP BRAIN RES, V79, P479
[24]  
Schmidt E. M., 1994, Society for Neuroscience Abstracts, V20, P982
[25]   Controlling cellular reactive responses around neural prosthetic devices using peripheral and local intervention strategies [J].
Shain, W ;
Spataro, L ;
Dilgen, J ;
Haverstick, K ;
Retterer, S ;
Isaacson, M ;
Saltzman, M ;
Turner, JN .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2003, 11 (02) :186-188
[26]  
SNELLINGS AE, 2003, SOC NEUR ABSTR, V29
[27]   HISTOPATHOLOGICAL EVALUATION OF MATERIALS IMPLANTED IN CEREBRAL-CORTEX [J].
STENSAAS, SS ;
STENSAAS, LJ .
ACTA NEUROPATHOLOGICA, 1978, 41 (02) :145-155
[28]  
STENSAAS SS, 1976, ACTA NEUROPATHOL, V35, P187
[29]   Brain responses to micro-machined silicon devices [J].
Szarowski, DH ;
Andersen, MD ;
Retterer, S ;
Spence, AJ ;
Isaacson, M ;
Craighead, HG ;
Turner, JN ;
Shain, W .
BRAIN RESEARCH, 2003, 983 (1-2) :23-35
[30]   Cerebral astrocyte response to micromachined silicon implants [J].
Turner, JN ;
Shain, W ;
Szarowski, DH ;
Andersen, M ;
Martins, S ;
Isaacson, M ;
Craighead, H .
EXPERIMENTAL NEUROLOGY, 1999, 156 (01) :33-49