Application of a combination of neutral red and amido black staining for rapid, reliable cytotoxicity testing of biomaterials

被引:82
作者
Ciapetti, G
Granchi, D
Verri, E
Savarino, L
Cavedagna, D
Pizzoferrato, A
机构
[1] Lab. Biocompatibility Res. I., Istituti Ortopedici Rizzoli, 40136, Bologna
关键词
cells; viability; neutral red; amido black; cytotoxicity;
D O I
10.1016/0142-9612(96)88670-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Cell viability and growth for cytotoxicity evaluation of materials for prosthetic devices has been tested using Various methods. The aim of this study was to extend the choice of reliable methods to quantify cytotoxicity of materials in vitro. By measuring both viability and growth of cells exposed to biomaterials in vitro, two different parameters are analysed and quantified upon reading of the absorbance of coloured solutions in a spectrophotometer. Neutral red uptake and amido black staining of cells have been used for cell viability and cell number measurement, respectively: they have been found to be well correlated with the number of surviving cells. These methods have been adjusted to a 96-well microplate cell culture system and re-evaluated as simple and reliable methods for the quantitative assessment of biomaterial effect on cells. Copyright (C) 1996 Elsevier Science Limited
引用
收藏
页码:1259 / 1264
页数:6
相关论文
共 20 条
[1]  
Allen Matthew, 1994, Clinical Materials, V16, P189
[2]  
[Anonymous], 1992, 10993 ISO
[3]   CORROSION AND CELL-CULTURE EVALUATIONS OF NICKEL-CHROMIUM DENTAL CASTING ALLOYS [J].
BUMGARDNER, JD ;
LUCAS, LC .
JOURNAL OF APPLIED BIOMATERIALS, 1994, 5 (03) :203-213
[4]   INVITRO EVALUATION OF CELL BIOMATERIAL INTERACTION BY MTT ASSAY [J].
CIAPETTI, G ;
CENNI, E ;
PRATELLI, L ;
PIZZOFERRATO, A .
BIOMATERIALS, 1993, 14 (05) :359-364
[5]   CELL-DAMAGE IN-VITRO FOLLOWING DIRECT-CONTACT WITH FINE PARTICLES OF TITANIUM, TITANIUM-ALLOY AND COBALT-CHROME-MOLYBDENUM ALLOY [J].
EVANS, EJ .
BIOMATERIALS, 1994, 15 (09) :713-717
[6]   SPECTROPHOTOMETRIC QUANTITATION OF ANCHORAGE-DEPENDENT CELL NUMBERS USING EXTRACTION OF NAPHTHOL BLUE-BLACK-STAINED CELLULAR PROTEIN [J].
EVERITT, E ;
WOHLFART, C .
ANALYTICAL BIOCHEMISTRY, 1987, 162 (01) :122-129
[8]   RE-EXAMINATION AND FURTHER DEVELOPMENT OF A PRECISE AND RAPID DYE METHOD FOR MEASURING CELL-GROWTH CELL KILL [J].
HANSEN, MB ;
NIELSEN, SE ;
BERG, K .
JOURNAL OF IMMUNOLOGICAL METHODS, 1989, 119 (02) :203-210
[9]  
LOWRY OH, 1951, J BIOL CHEM, V193, P265