Investigation of PVA cryogel Young's modulus stability with time, controlled by a simple reliable technique

被引:36
作者
Duboeuf, Francois [1 ]
Basarab, Adrian [1 ]
Liebgott, Herve [1 ]
Brusseau, Elisabeth [1 ]
Delachartre, Philippe [1 ]
Vray, Didier [1 ]
机构
[1] Univ Lyon, UCB Lyon 1, INSA Lyon, CREATIS LRMN,CNRS,INSERM,U630,UMR 5220, F-69621 Villeurbanne, France
关键词
biological tissues; biomedical ultrasonics; biomimetics; melting; phantoms; polymer gels; Young's modulus; POLYVINYL-ALCOHOL CRYOGEL; ELASTICITY; ULTRASOUND; ELASTOGRAPHY; PHANTOMS; TISSUES; MR;
D O I
10.1118/1.3065031
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We describe a quasistatic method for mechanical characterization of tissue-mimicking material used in elastography. We demonstrate that it is possible to assess the elasticity modulus with a reasonable reproducibility using simple and easy tools and methods. Possessing a simple relevant technique with evaluated relative error to assess Young's modulus of these phantoms could deeply improve the quality of the research in the field of elastography. The method was tested and validated with four samples of polyvinyl alcohol (PVA) cryogel with different elasticity values corresponding to those of stiffer soft biological tissues. Young's moduli, varying from 70 to 180 kPa depending on the number of freeze-thaw cycles (two to five), were measured within strict measurement conditions and found to have a reproducibility varying from 4% to 8%. Relative error, estimated as the ratio between observed and reference values, varied from 16% to 32%. Besides, measurement stability over 4 months was evaluated. The method demonstrated good feasibility and acceptable reproducibility for mechanically characterizing and controlled over time phantoms used for validating new potential ultrasound imaging techniques in the field of elastography. Nevertheless, in this study, investigation was performed on gel possessing young's modulus values ranging from 80 to 215 kPa. Some tissue values of Young'modulus were reported to be lower, ranging from 0.6 to 28 kPa as liver or glandular values. Consequently, further validation of this static method for mechanical characterization of phantom gels should be performed using softer PVA cryogel.
引用
收藏
页码:656 / 661
页数:6
相关论文
共 11 条
  • [1] Polyvinyl alcohol cryogel: An ideal phantom material for MR studies of arterial flow and elasticity
    Chu, KC
    Rutt, BK
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (02) : 314 - 319
  • [2] Characterization of PVA cryogel for intravascular ultrasound elasticity imaging
    Fromageau, J
    Brusseau, E
    Vray, D
    Gimenez, G
    Delachartre, P
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2003, 50 (10) : 1318 - 1324
  • [3] Estimation of polyvinyl alcohol cryogel mechanical properties with four ultrasound elastography methods and comparison with gold standard testings
    Fromageau, Jeremie
    Gennisson, Jean-Luc
    Schmitt, Cedric
    Maurice, Roch L.
    Mongrain, Rosaire
    Cloutier, Guy
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (03) : 498 - 509
  • [4] Elastic moduli of breast and prostate tissues under compression
    Krouskop, TA
    Wheeler, TM
    Kallel, F
    Garra, BS
    Hall, T
    [J]. ULTRASONIC IMAGING, 1998, 20 (04) : 260 - 274
  • [5] Tissue-mimicking agar/gelatin materials for use in heterogeneous elastography phantoms
    Madsen, EL
    Hobson, MA
    Shi, HR
    Varghese, T
    Frank, GR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (23) : 5597 - 5618
  • [6] NEW POLYVINYL-ALCOHOL GEL MATERIAL FOR MRI PHANTOMS
    MANO, I
    GOSHIMA, H
    NAMBU, M
    IIO, M
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (06) : 921 - 926
  • [7] ELASTOGRAPHY - A QUANTITATIVE METHOD FOR IMAGING THE ELASTICITY OF BIOLOGICAL TISSUES
    OPHIR, J
    CESPEDES, I
    PONNEKANTI, H
    YAZDI, Y
    LI, X
    [J]. ULTRASONIC IMAGING, 1991, 13 (02) : 111 - 134
  • [8] SAMANI A, 2003, PHYS MED BIOL, V14, P83
  • [9] Improved common carotid elasticity and intima-media thickness measurements from computer analysis of sequential ultrasound frames
    Selzer, RH
    Mack, WJ
    Lee, PL
    Kwong-Fu, H
    Hodis, HN
    [J]. ATHEROSCLEROSIS, 2001, 154 (01) : 185 - 193
  • [10] Poly(vinyl alcohol) cryogel phantoms for use in ultrasound and MR imaging
    Surry, KJM
    Austin, HJB
    Fenster, A
    Peters, TM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (24) : 5529 - 5546