Correlation between Temperature-dependent Fatigue Resistance and Differential Scanning Calorimetry Analysis for 2 Contemporary Rotary Instruments

被引:31
作者
Arias, Ana [1 ,2 ]
Macorra, Jose C. [1 ]
Govindjee, Sanjay [3 ]
Peters, Ove A. [2 ]
机构
[1] Univ Complutense Madrid, Dept Conservat Dent, Sch Dent, Ciudad Univ,Plaza Ramon y Cajal S-N, Madrid 28040, Spain
[2] Univ Pacific, Dept Endodont, Arthur A Dugoni Sch Dent, San Francisco, CA USA
[3] Univ Calif Berkeley, Dept Civil & Environm Engn, Berkeley, CA 94720 USA
关键词
NICKEL-TITANIUM INSTRUMENTS; ROOT-CANAL PREPARATION; CYCLIC FATIGUE; ENDODONTIC INSTRUMENTS; THERMOMECHANICAL TREATMENT; FILES; FRACTURE; UNIVERSAL; RECIPROC;
D O I
10.1016/j.joen.2017.11.022
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The aim of this study was to assess differences in cyclic fatigue (CF) life of contemporary heat-treated nickel-titanium rotary instruments at room and body temperatures and to document corresponding phase transformations. Methods: Forty Hy-flex EDM (H-EDM) files (Coltene, Cuyahoga Falls, OH [#25/.08, manufactured by electrical discharge machining]) and 40 TRUShape (TS) files (Dentsply Tulsa Dental Specialties, Tulsa, OK [#25/.06v, manufactured by grinding and shape setting]) were divided into 2 groups (n = 20) for CF resistance tests in a water bath either at room (22 degrees C +/- 0.5 degrees C) or body temperature (37 degrees C +/- 0.5 degrees C). Instruments were rotated in a simulated canal (angle = 60, radius = 3 mm, and center of the curvature 5 mm from the tip) until fracture occurred. The motor was controlled by an electric circuit that was interrupted after instrument fracture. The mean half-life and beta and eta Weibull parameters were determined and compared. Two instruments of each brand were subjected to differential scanning calorimetry (DSC). Results: While TS instruments lasted significantly longer at room temperature (mean life = 234.7 seconds; 95% confidence interval [CI], 209-263.6) than at body temperature (mean life = 83.2 seconds; 95% CI, 76-91.1), temperature did not affect H-EDM behavior (room temperature mean life = 725.4 seconds; 95% CI, 658.8-798.8 and body temperature mean life = 717.9 seconds; 95% CI, 636.8-809.3). H-EDM instruments significantly outlasted TS instruments at both temperatures. At body temperature, TS was predominantly austenitic, whereas H-EDM was martensitic or in R-phase. TS was in a mixed austenitic/martensitic phase at 22 degrees C, whereas H-EDM was in the same state as at 37 degrees C. Conclusions: H-EDM had a longer fatigue life than TS, which showed a marked decrease in fatigue life at body temperature; neither the life span nor the state of the microstructure in the DSC differed for H-EDM between room or body temperature.
引用
收藏
页码:630 / 634
页数:5
相关论文
共 35 条
[1]  
Abernethy R. B., 2000, NEW WEIBULL HDB RELI
[2]  
Arias A, 2017, AUST ENDOD J, V18, P269
[3]   Differences in Cyclic Fatigue Resistance at Apical and Coronal Levels of Reciproc and Wave One New Files [J].
Arias, Ana ;
Perez-Higueras, Juan J. ;
de la Macorra, Jose C. .
JOURNAL OF ENDODONTICS, 2012, 38 (09) :1244-1248
[4]  
Bergmans L, 2001, AM J DENT, V14, P324
[5]   Fatigue testing of a NiTi rotary instrument. Part 1: Strain-life relationship [J].
Cheung, G. S. P. ;
Darvell, B. W. .
INTERNATIONAL ENDODONTIC JOURNAL, 2007, 40 (08) :612-618
[6]  
Coltene Whaledent, NEW FIL GEN
[7]   Evidence for Reduced Fatigue Resistance of Contemporary Rotary Instruments Exposed to Body Temperature [J].
de Vasconcelos, Rafaela Andrade ;
Murphy, Sarah ;
Talge Carvalho, Claudio Antonio ;
Govindjee, Rajiv G. ;
Govindjee, Sanjay ;
Peters, Ove A. .
JOURNAL OF ENDODONTICS, 2016, 42 (05) :782-787
[8]   Blue Thermomechanical Treatment Optimizes Fatigue Resistance and Flexibility of the Reciproc Files [J].
De-Deus, Gustavo ;
Nogueira Leal Silva, Emmanuel Joao ;
Leal Vieira, Victor Talarico ;
Belladonna, Felipe Goncalves ;
Elias, Carlos Nelson ;
Plotino, Gianluca ;
Grande, Nicola Maria .
JOURNAL OF ENDODONTICS, 2017, 43 (03) :462-466
[9]   The Effect of Temperature on Cyclic Fatigue of Nickel-titanium Rotary Endodontic Instruments [J].
Dosanjh, Amandeep ;
Paurazas, Susan ;
Askar, Mazin .
JOURNAL OF ENDODONTICS, 2017, 43 (05) :823-826
[10]   Cyclic fatigue analysis of twisted file rotary NiTi instruments used in reciprocating motion [J].
Gambarini, G. ;
Gergi, R. ;
Naaman, A. ;
Osta, N. ;
Al Sudani, D. .
INTERNATIONAL ENDODONTIC JOURNAL, 2012, 45 (09) :802-806