Deformation of press-fitted metallic resurfacing cups. Part 1: experimental simulation

被引:85
|
作者
Jin, ZM [1 ]
Meakins, S
Morlock, MM
Parsons, P
Hardaker, C
Flett, M
Isaac, G
机构
[1] Univ Leeds, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[2] DePuy Int Ltd, Leeds, W Yorkshire, England
[3] Tech Univ Hamburg, Biomech Sect, D-2100 Hamburg, Germany
关键词
metal-on-metal articulations; hip resurface; cup deformation;
D O I
10.1243/095441105X69150
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The interference press fit of a metallic one-piece acetabular cup employed for metal-on-metal hip resurfacing procedures was investigated experimentally under laboratory conditions in the present study, in particular regarding the cup deformation. Tests were carried out in cadavers as well as polyurethane foams of various grades with different elastic moduli to represent different cancellous bone qualities. The cadaver test was used to establish the most suitable configuration of the foam model representing realistic support and geometrical conditions at the pelvis. It was found that a spherical cavity, with two identical areas relieved on opposite sides, was capable of creating a two-point pinching action of the ischeal and ilial columns on the cup as the worst-case scenario. Furthermore, the cup deformation produced from such a two-point loading model with a grade 30 foam was similar to that measured from the cadaver test. Therefore, such a protocol was employed in subsequent experimental tests. For a given size of the outside diameter of the cup of 60 mm, the cup deflection was shown to be dependent largely on the cup wall thickness and the diametral interference between cup and prepared cavity at implantation. For a relatively thin cup with a wall thickness between 2.3 mm (equator) and 4 mm (pole) and with a modest nominal diametral interference of 1 mm, which corresponds to an actual interference of approximately 0.5 mm, the maximum diametral cup deflection (at the rim) was around 60 mu m, compared with a diametral clearance of 80-120 mu m between the femoral head and the acetabular cup, generally required for fluid-film lubrication and tribological performances. Stiffening of the cup, by both thickening and lateralizing by 1 mm, reduced the cup deformation to between 30 and 50 mu m with actual diametral interferences between 0.5 and 1 mm.
引用
收藏
页码:299 / 309
页数:11
相关论文
共 50 条
  • [1] Deformation of press-fitted metallic resurfacing cups. Part 2: finite element simulation
    Yew, A
    Jin, ZM
    Donn, A
    Morlock, MM
    Isaac, G
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 311 - 319
  • [2] Simulation Research on the Fretting Wear of Press-fitted Hollow Axle
    Gong Y.
    Lu L.
    Zhang Y.
    Zhang J.
    Zeng D.
    Lu, Liantao (luliantao@swjtu.cn), 1600, Chinese Mechanical Engineering Society (53): : 123 - 130
  • [3] Stress concentration of transition groove induced by a press-fitted part in railway axles
    Yamamoto, M.
    Ishiduka, H.
    INTERNATIONAL JOURNAL OF FATIGUE, 2017, 97 : 48 - 55
  • [4] Numerical Simulation of Fretting Wear in Press-fitted Shaft considering the Running-in Period
    Lee, Dong-Hyong
    Kwon, Seok-Jin
    You, Won-Hee
    ADVANCES IN FRACTURE AND DAMAGE MECHANICS VIII, 2010, 417-418 : 717 - 720
  • [5] Simulation research on fretting wear of train axles with interference fit based on press-fitted specimen
    Shu, Yiliang
    Yang, Guangxue
    Liu, Zhiming
    WEAR, 2023, 523
  • [6] Damage Analysis and Finite Element Simulation of Fretting Wear on Press-Fitted Surface of Railway Axle
    Yuan C.
    Li Y.
    Yang K.
    Lu L.
    Zou L.
    Li W.
    Zeng D.
    Mocaxue Xuebao/Tribology, 2020, 40 (04): : 520 - 530
  • [7] Acetabular revision arthroplasty using press-fitted jumbo cups: an average 10-year follow-up study
    Moon, Jun-Ki
    Ryu, Jaejoon
    Kim, Yeesuk
    Yang, Jae-Hyuk
    Hwang, Kyu-Tae
    Kim, Young-Ho
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2019, 139 (08) : 1149 - 1160
  • [8] Experimental and numerical study on the fretting wear-fatigue interaction evolution in press-fitted axles
    Zou, Lang
    Zeng, Dongfang
    Dong, Yihui
    Li, Jun
    Chen, Xiong
    Zhao, Hai
    Lu, Liantao
    INTERNATIONAL JOURNAL OF FATIGUE, 2023, 175
  • [9] Acetabular revision arthroplasty using press-fitted jumbo cups: an average 10-year follow-up study
    Jun-Ki Moon
    Jaejoon Ryu
    Yeesuk Kim
    Jae-Hyuk Yang
    Kyu-Tae Hwang
    Young-Ho Kim
    Archives of Orthopaedic and Trauma Surgery, 2019, 139 : 1149 - 1160
  • [10] Fretting fatigue crack growth simulation and residual life assessment of railway press-fitted axle
    Xu, Tian
    Lu, Liantao
    Zeng, Dongfang
    Zou, Lang
    ENGINEERING FRACTURE MECHANICS, 2023, 286