Preconcentration, Speciation, and Determination of Mercury by Solid Phase Extraction with Cold Vapor Atomic Absorption Spectrometry

被引:27
作者
Turker, Ali Rehber [1 ]
Cabuk, Dilek [1 ]
Yalcinkaya, Ozcan [1 ]
机构
[1] Gazi Univ, Fac Sci, Dept Chem, TR-06500 Ankara, Turkey
关键词
Cold vapor atomic absorption spectrometry; Mercury; Methyl mercury; Solid phase extraction; Speciation; PERFORMANCE LIQUID-CHROMATOGRAPHY; METHYL MERCURY; INORGANIC MERCURY; ONLINE SPECIATION; WATER SAMPLES; FISH; FLUORESCENCE; AAS; METHYLMERCURY; DIGESTION;
D O I
10.1080/00032719.2012.753608
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study, inorganic mercury [iHg (Hg2+)] and monomethylmercury [MeHg (CH3Hg+)] ions have been preconcentrated and determined by a column solid phase extraction method. Cold vapor atomic absorption spectrometry was used for the determination of mercury species. Optimum retention and elution parameters were determined for both mercury species separately. For this purpose, the effect of pH, flow rate and volume of sample solution, type, and concentration of eluent on the recovery of mercury species were investigated. Effects of possible interfering ions were also studied. Optimum values of flow rate, pH, and volume of sample solution were found to be 3mL min1, 4.0 and 50mL, respectively, for both iHg and MeHg ions. Although the both species were retained at pH 4, the elution could be performed only with different eluents. While MeHg ions were eluted with 10mL of 0.1mol/L HCl solution, iHg ions were eluted with 10mL of 0.2molL1 thioureasolution in 3molL1HCl. Limits of detection for iHg and MeHg were 0.44 mu gL1 and 0.56 mu gL1, respectively. To validate the proposed speciation and/or preconcentration method, a certified reference material (ERM-CE464) was analyzed and MeHg was determined with a relative error of 7.3%. The accuracy was also checked by analyzing spiked water and fish samples for both species.
引用
收藏
页码:1155 / 1170
页数:16
相关论文
共 50 条
  • [1] Immobilized Activated Carbon as Sorbent in Solid Phase Extraction with Cold Vapor Atomic Absorption Spectrometry for the Preconcentration and Determination of Mercury Species in Water and Freshwater Fish Samples
    Thongsaw, Anion
    Sananmuang, Ratana
    Udnan, Yuthapong
    Ampiah-Bonney, Richmond J.
    Chaiyasith, Wipharat Chuachuad
    ANALYTICAL SCIENCES, 2019, 35 (11) : 1195 - 1202
  • [2] Preconcentration, speciation and determination of ultra trace amounts of mercury by modified octadecyl silica membrane disk/electron beam irradiation and cold vapor atomic absorption spectrometry
    Ashkenani, Hamid
    Dadfarnia, Shayessteh
    Shabani, Ali Mohammad Haji
    Jaffari, Abbas Ali
    Behjat, Abbas
    JOURNAL OF HAZARDOUS MATERIALS, 2009, 161 (01) : 276 - 280
  • [3] Automated on-line preconcentration of trace aqueous mercury with gold trap focusing for cold vapor atomic absorption spectrometry
    Puanngam, Mahitti
    Dasgupta, Purnendu K.
    Unob, Fuangfa
    TALANTA, 2012, 99 : 1040 - 1045
  • [4] Immobilized Activated Carbon as Sorbent in Solid Phase Extraction with Cold Vapor Atomic Absorption Spectrometry for the Preconcentration and Determination of Mercury Species in Water and Freshwater Fish Samples
    Arnon Thongsaw
    Ratana Sananmuang
    Yuthapong Udnan
    Richmond J. Ampiah-Bonney
    Wipharat Chuachuad Chaiyasith
    Analytical Sciences, 2019, 35 : 1195 - 1202
  • [5] Preconcentration and determination of mercury (II) and methylmercury in waters by immobilized 1,5-diphenylcarbazone and cold vapor atomic absorption spectrometry
    Dadfarnia, S
    Salmanzadeh, AM
    Shabani, AMH
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2002, 23 (12) : 1719 - 1723
  • [6] Mercury speciation in sea food by flow injection cold vapor atomic absorption spectrometry using selective solid phase extraction
    Alonso, E. Vereda
    Cordero, M. T. Siles
    de Torres, A. Garcia
    Rudner, P. Canada
    Pavon, J. M. Cano
    TALANTA, 2008, 77 (01) : 53 - 59
  • [7] Speciation of mercury in fish samples by flow injection catalytic cold vapour atomic absorption spectrometry
    Zhang, Yanlin
    Adeloju, Samuel B.
    ANALYTICA CHIMICA ACTA, 2012, 721 : 22 - 27
  • [8] A Preconcentration Procedure for Determination of Ultra-Trace Mercury (II) in Environmental Samples Employing Continuous-Flow Cold Vapor Atomic Absorption Spectrometry
    Adlnasab, Laleh
    Ebrahimzadeh, Homeira
    Asgharinezhad, Ali Akbar
    Aghdam, Mahnaz Nasiri
    Dehghani, Ali
    Esmaeilpour, Sousan
    FOOD ANALYTICAL METHODS, 2014, 7 (03) : 616 - 628
  • [9] Suitable extraction of soils and sediments for mercury species and determination combined with the cold vapor generation atomic absorption spectrometry technique
    Almeida, Lone L. S.
    Oliveira, Monica D. R.
    Silva, Jose B. B.
    Coelho, Nivia M. M.
    MICROCHEMICAL JOURNAL, 2016, 124 : 326 - 330
  • [10] Modification of a Reaction Vessel for the Determination of Total Mercury in Hair by Cold Vapor Atomic Absorption Spectrometry
    Pereira, Eduardo F.
    SouzaDe, Jurandir Rodrigues
    BIOLOGICAL TRACE ELEMENT RESEARCH, 2009, 130 (03) : 210 - 217