Impact of radiation dose and iterative reconstruction on pulmonary nodule measurements at chest CT: a phantom study

被引:21
作者
Kim, Hyungjin [1 ,2 ]
Park, Chang Min [1 ,3 ]
Chae, Hee-Dong [1 ]
Lee, Sang Min [1 ]
Goo, Jin Mo [1 ,3 ]
机构
[1] Seoul Natl Univ, Coll Med, Dept Radiol, Seoul, South Korea
[2] Air Force Educ & Training Command, Aerosp Med Grp, Jinju, South Korea
[3] Seoul Natl Univ, Canc Res Inst, Seoul, South Korea
关键词
LUNG-CANCER; VOLUMETRY; DETECTABILITY; MANAGEMENT; REDUCTION; ACCURACY; SCANS;
D O I
10.5152/dir.2015.14541
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE We aimed to identify the impact of radiation dose and iterative reconstruction (IR) on measurement of pulmonary nodules by chest computed tomography (CT). METHODS CT scans were performed on a chest phantom containing various nodules (diameters of 3, 5, 8, 10, and 12 mm; +100, -630 and -800 HU for each diameter) at 80, 100, 120 kVp and 10, 20, 50, 100 mAs (a total of 12 radiation dose settings). Each CT was reconstructed using filtered back projection, iDose(4), and iterative model reconstruction (IMR). Thereafter, two radiologists measured the diameter and attenuation of the nodules. Noise, contrast-to-noise ratio and signal-to-noise ratio of CT images were also obtained. Influence of radiation dose and reconstruction algorithm on measurement error and objective image quality metrics was analyzed using generalized estimating equations. RESULTS The 80 kVp, 10 mAs CT scan was not feasible for the measurement of 3 mm sized simulated ground-glass nodule (GGN); otherwise, diameter measurement error was not significantly influenced by radiation dose (P > 0.05). IR did not have a significant impact on diameter measurement error for simulated solid nodules (P > 0.05). However, for simulated GGNs, IMR was associated with significantly decreased relative diameter measurement error (P < 0.001). Attenuation measurement error was not significantly influenced by either radiation dose or reconstruction algorithm (P > 0.05). Objective image quality was significantly better with IMR (P < 0.05). CONCLUSION Nodule measurements were not affected by radiation dose except for 3 mm simulated GGN on 80 kVp, 10 mAs dose setting. However, for GGNs, IMR may help reduce diameter measurement error while improving image quality.
引用
收藏
页码:459 / 465
页数:7
相关论文
共 23 条
  • [1] Reduced Lung-Cancer Mortality with Low-Dose Computed Tomographic Screening
    Aberle, Denise R.
    Adams, Amanda M.
    Berg, Christine D.
    Black, William C.
    Clapp, Jonathan D.
    Fagerstrom, Richard M.
    Gareen, Ilana F.
    Gatsonis, Constantine
    Marcus, Pamela M.
    Sicks, JoRean D.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2011, 365 (05) : 395 - 409
  • [2] Benefits and Harms of CT Screening for Lung Cancer A Systematic Review
    Bach, Peter B.
    Mirkin, Joshua N.
    Oliver, Thomas K.
    Azzoli, Christopher G.
    Berry, Donald A.
    Brawley, Otis W.
    Byers, Tim
    Colditz, Graham A.
    Gould, Michael K.
    Jett, James R.
    Sabichi, Anita L.
    Smith-Bindman, Rebecca
    Wood, Douglas E.
    Qaseem, Amir
    Detterbeck, Frank C.
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2012, 307 (22): : 2418 - 2429
  • [3] Measuring agreement in method comparison studies
    Bland, JM
    Altman, DG
    [J]. STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) : 135 - 160
  • [4] CT Screening and Follow-Up of Lung Nodules: Effects of Tube Current-Time Setting and Nodule Size and Density on Detectability and of Tube Current-Time Setting on Apparent Size
    Christe, Andreas
    Torrente, Jaled Charimo
    Lin, Margaret
    Yen, Andrew
    Hallett, Rich
    Roychoudhury, Kingshuk
    Schmitzberger, Florian
    Vock, Peter
    Roos, Justus
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (03) : 623 - 630
  • [5] Computer-assisted solid lung nodule 3D volumetry on CT: influence of scan mode and iterative reconstruction: a CT phantom study
    Coenen, Adriaan
    Honda, Osamu
    van der Jagt, Eric J.
    Tomiyama, Noriyuki
    [J]. JAPANESE JOURNAL OF RADIOLOGY, 2013, 31 (10) : 677 - 684
  • [6] Multisection CT Protocols: Sex- and Age-specific Conversion Factors Used to Determine Effective Dose from Dose-Length Product
    Deak, Paul D.
    Smal, Yulia
    Kalender, Willi A.
    [J]. RADIOLOGY, 2010, 257 (01) : 158 - 166
  • [7] Combination of a Low-Tube-Voltage Technique With Hybrid Iterative Reconstruction (iDose) Algorithm at Coronary Computed Tomographic Angiography
    Funama, Yoshinori
    Taguchi, Katsuyuki
    Utsunomiya, Daisuke
    Oda, Seitaro
    Yanaga, Yumi
    Yamashita, Yasuyuki
    Awai, Kazuo
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2011, 35 (04) : 480 - 485
  • [8] Linear and Volume Measurements of Pulmonary Nodules at Different CT Dose Levels - Intrascan and Interscan Analysis
    Hein, P. A.
    Romano, V. C.
    Rogalla, P.
    Klessen, C.
    Lembcke, A.
    Dickenz, V.
    Bornemannz, L.
    Bauknecht, H. -C.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2009, 181 (01): : 24 - 31
  • [9] Model-Based Iterative Reconstruction Technique for Ultralow-Dose Chest CT Comparison of Pulmonary Nodule Detectability With the Adaptive Statistical Iterative Reconstruction Technique
    Katsura, Masaki
    Matsuda, Izuru
    Akahane, Masaaki
    Yasaka, Koichiro
    Hanaoka, Shohei
    Akai, Hiroyuki
    Sato, Jiro
    Kunimatsu, Akira
    Ohtomo, Kuni
    [J]. INVESTIGATIVE RADIOLOGY, 2013, 48 (04) : 206 - 212
  • [10] A Comparison of Two Commercial Volumetry Software Programs in the Analysis of Pulmonary Ground-Glass Nodules: Segmentation Capability and Measurement Accuracy
    Kim, Hyungjin
    Park, Chang Min
    Lee, Sang Min
    Lee, Hyun-Ju
    Goo, Jin Mo
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2013, 14 (04) : 683 - 691