Flow-dependent changes in Doppler-derived aortic valve effective orifice area are real and not due to artifact

被引:41
作者
Kadem, L
Rieu, R
Dumesnil, JG
Durand, LG
Pibarot, P
机构
[1] Univ Laval, Quebec Heart Inst, Laval Hosp, Res Ctr, Ste Foy, PQ G1V 4G5, Canada
[2] Univ Mediterranee, CNRS, IRPHE, Cardiovasc Biomech Team, Marseille, France
[3] Inst Rech Clin Montreal, Montreal, PQ H2W 1R7, Canada
基金
加拿大健康研究院;
关键词
D O I
10.1016/j.jacc.2005.05.100
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES We sought to determine whether the flow-dependent changes in Doppler-derived valve effective orifice area (EOA) are real or due to artifact. BACKGROUND It has frequently been reported that the EOA may vary with transvalvular flow in patients with aortic stenosis. However, the explanation of the flow dependence of EOA remains controversial and some studies have suggested that the EOA estimated by Doppler-echocardiography (EOA(Dop)) may underestimate the actual EOA at low flow rates. METHODS One bioprosthetic valve and three rigid orifices were tested in a mock flow circulation model over a wide range of flow rates. The EOA(DOp) was compared with reference values obtained using particle image velocimetry (EOA(PIV)). RESULTS There was excellent agreement between EOA(Dop) and EOA(PIV) (r(2) = 0.94). For rigid orifices of 0.5 and 1.0 cm(2), no significant change in the EOA was observed with increasing flow rate. However, substantial increases of both EOA(Dop) and EOA(PIV) were observed when stroke volume increased from 20 to 70 ml both in the 1.5 cm(2) rigid orifice (+52% for EOA(Dop) and +54% for EOA(PIV)) and the bioprosthetic valve (+62% for EOA(Dop) and +63% for EOA(PIV)); such changes are explained either by the presence of unsteady effects at low flow rates and/or by an increase in valve leaflet opening. CONCLUSIONS The flow-dependent changes in EOA(Dop) are not artifacts but represent real changes in EOA attributable either to unsteady effects at low flow rates and/or to changes in valve leaflet opening. Such changes in EOA(Dop) can be relied on for clinical judgment making.
引用
收藏
页码:131 / 137
页数:7
相关论文
共 47 条
[41]  
SHANDAS R, 1996, P 18 ANN INT C IEEE
[42]   Flow dependence of valve area in aortic stenosis: Relation to valve morphology [J].
Shively, BK ;
Charlton, GA ;
Crawford, MH ;
Chaney, RK .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (03) :654-660
[43]   VARIABLE EFFECTS OF CHANGES IN FLOW-RATE THROUGH THE AORTIC, PULMONARY AND MITRAL-VALVES ON VALVE AREA AND FLOW VELOCITY - IMPACT ON QUANTITATIVE DOPPLER FLOW CALCULATIONS [J].
STEWART, WJ ;
LENG, J ;
MICH, R ;
PANDIAN, N ;
GUERRERO, JL ;
WEYMAN, AE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 6 (03) :653-662
[44]   Extension of PIV to super resolution using PTV [J].
Stitou, A ;
Riethmuller, ML .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2001, 12 (09) :1398-1403
[45]   EFFECTS OF VARIATIONS IN FLOW ON AORTIC-VALVE AREA IN AORTIC-STENOSIS BASED ON IN-VIVO PLANIMETRY OF AORTIC-VALVE AREA BY MULTIPLANE TRANSESOPHAGEAL ECHOCARDIOGRAPHY [J].
TARDIF, JC ;
MILLER, DS ;
PANDIAN, NG ;
SCHWARTZ, SL ;
GORDON, G ;
TIERNEY, R ;
ENGLAND, M ;
PANZA, W .
AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (03) :193-198
[46]   Simultaneous determination of aortic valve area by the gorlin formula and by transesophageal echocardiography under different transvalvular flow conditions - Evidence that anatomic aortic valve area does not change with variations in flow in aortic stenosis [J].
Tardif, JC ;
Rodrigues, AG ;
Hardy, JF ;
Leclerc, Y ;
Petitclerc, R ;
Mongrain, R ;
Mercier, LA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 29 (06) :1296-1302
[47]   COMPARISON OF VALVULAR RESISTANCE, STROKE WORK LOSS, AND GORLIN VALVE AREA FOR QUANTIFICATION OF AORTIC-STENOSIS - AN IN-VITRO STUDY IN A PULSATILE AORTIC FLOW MODEL [J].
VOELKER, W ;
REUL, H ;
NIENHAUS, G ;
STELZER, T ;
SCHMITZ, B ;
STEEGERS, A ;
KARSCH, KR .
CIRCULATION, 1995, 91 (04) :1196-1204