Mapping motion from 4D-MRI to 3D-CT for use in 4D dose calculations: A technical feasibility study

被引:104
作者
Boye, Dirk [1 ,2 ]
Lomax, Tony [1 ,3 ]
Knopf, Antje [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, CH-5232 Villigen, Switzerland
[2] ETH, Comp Vis Lab, CH-8092 Zurich, Switzerland
[3] ETH, CH-8092 Zurich, Switzerland
基金
瑞士国家科学基金会;
关键词
4D-MRI; 4D dose calculations; DEFORMABLE REGISTRATION; COMPUTED-TOMOGRAPHY; RESPIRATORY MOTION; TUMOR TRACKING; CT; RADIOTHERAPY; APPROXIMATION; INTERPOLATION; SEGMENTATION; STRATEGIES;
D O I
10.1118/1.4801914
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Target sites affected by organ motion require a time resolved (4D) dose calculation. Typical 4D dose calculations use 4D-CT as a basis. Unfortunately, 4D-CT images have the disadvantage of being a "snap-shot" of the motion during acquisition and of assuming regularity of breathing. In addition, 4D-CT acquisitions involve a substantial additional dose burden to the patient making many, repeated 4D-CT acquisitions undesirable. Here the authors test the feasibility of an alternative approach to generate patient specific 4D-CT data sets. Methods: In this approach motion information is extracted from 4D-MRI. Simulated 4D-CT data sets [which the authors call 4D-CT(MRI)] are created by warping extracted deformation fields to a static 3D-CT data set. The employment of 4D-MRI sequences for this has the advantage that no assumptions on breathing regularity are made, irregularities in breathing can be studied and, if necessary, many repeat imaging studies (and consequently simulated 4D-CT data sets) can be performed on patients and/or volunteers. The accuracy of 4D-CT(MRI)s has been validated by 4D proton dose calculations. Our 4D dose algorithm takes into account displacements as well as deformations on the originating 4D-CT/4D-CT(MRI) by calculating the dose of each pencil beam based on an individual time stamp of when that pencil beam is applied. According to corresponding displacement and density-variation-maps the position and the water equivalent range of the dose grid points is adjusted at each time instance. Results: 4D dose distributions, using 4D-CT( MRI) data sets as input were compared to results based on a reference conventional 4D-CT data set capturing similar motion characteristics. Almost identical 4D dose distributions could be achieved, even though scanned proton beams are very sensitive to small differences in the patient geometry. In addition, 4D dose calculations have been performed on the same patient, but using 4D-CT(MRI) data sets based on variable breathing patterns to show the effect of possible irregular breathing on active scanned proton therapy. Using a 4D-CT(MRI), including motion irregularities, resulted in significantly different proton dose distributions. Conclusions: The authors have demonstrated that motion information from 4D-MRI can be used to generate realistic 4D-CT data sets on the basis of a single static 3D-CT data set. 4D-CT(MRI) presents a novel approach to test the robustness of treatment plans in the circumstance of patient motion. (c) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:11
相关论文
共 43 条
  • [1] Experimental verification of IMPT treatment plans in an anthropomorphic phantom in the presence of delivery uncertainties
    Albertini, F.
    Casiraghi, M.
    Lorentini, S.
    Rombi, B.
    Lomax, A. J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (14) : 4415 - 4431
  • [2] Integrated radiotherapy imaging system (IRIS): design considerations of tumour tracking with linac gantry-mounted diagnostic x-ray systems with flat-panel detectors
    Berbeco, RI
    Jiang, SB
    Sharp, GC
    Chen, GTY
    Mostafavi, H
    Shirato, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (02) : 243 - 255
  • [3] Quantification of interplay effects of scanned particle beams and moving targets
    Bert, Christoph
    Groezinger, Sven O.
    Rietzel, Eike
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (09) : 2253 - 2265
  • [4] Consistent image registration
    Christensen, GE
    Johnson, HJ
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (07) : 568 - 582
  • [5] De Laney T.F., 2007, Proton and charged particle radiotherapy
  • [6] Multiscale joint segmentation and registration of image morphology
    Droske, Marc
    Rumpf, Martin
    [J]. IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 2007, 29 (12) : 2181 - 2194
  • [7] Respiration-correlated spiral CT: A method of measuring respiratory-induced anatomic motion for radiation treatment planning
    Ford, EC
    Mageras, GS
    Yorke, E
    Ling, CC
    [J]. MEDICAL PHYSICS, 2003, 30 (01) : 88 - 97
  • [8] Aspects of MR image distortions in radiotherapy treatment planning
    Fransson, A
    Andreo, P
    Pötter, R
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2001, 177 (02) : 59 - 73
  • [9] Gemmel A., 2009, 11th International Congress of the IUPESM. World Congress on Medical Physics and Biomedical Engineering. Radiation Oncology (WC 2009), P377, DOI 10.1007/978-3-642-03474-9_106
  • [10] Assessment of Uncertainties in Treatment Planning for Scanned Ion Beam Therapy of Moving Tumors
    Hild, Sebastian
    Durante, Marco
    Bert, Christoph
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (02): : 528 - 535