DO MAXIMUM INTENSITY PROJECTION IMAGES TRULY CAPTURE TUMOR MOTION?

被引:65
作者
Park, Kwangyoul [1 ]
Huang, Long [1 ]
Gagne, Havaleh [1 ]
Papiez, Lech [1 ]
机构
[1] Univ Texas SW Med Ctr Dallas, Dept Radiat Oncol, Dallas, TX 75390 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2009年 / 73卷 / 02期
关键词
Four-dimensional computed tomography (4D-CT); Maximum intensity projection (MIP); Stereotactic body radiotherapy; Lung cancer; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; LUNG-CANCER; TARGET VOLUME; CT; SCANS; IMPACT; TIME; MIP;
D O I
10.1016/j.ijrobp.2008.10.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: For the treatment of patients with lung cancer, internal target volume frequently is determined by using maximum intensity projection (MIP) images generated by means of four-dimensional computed tomography (4D-CT). To check the accuracy of MIPs for various target motions, especially for targets moving irregularly, we performed phantom studies using a programmable dynamic lung phantom. Methods and Materials: A custom-built programmable lung phantom was used to simulate irregular target motions along the superior-inferior direction. After scanning in helical mode using 4D-CT, reconstructed phase and MIP images were imported into the Pinnacle 8.0 treatment planning system for image analysis. Results: For all regular periodic target motions with constant amplitude and period, the measured MIP target span along the superior-inferior direction was accurate within 2-3 mm of the real target motion span. For irregular target motions with varying amplitudes and periods, the measured MIP target span systematically underrepresented the real range of target motion by more than 1 cm in some cases. The difference between measured MIP target span and real target span decreased as the target moved faster. We associate these discrepancies with the fact that current reconstruction algorithms of commercial 4D-CT are based on phase binning. Conclusions: According to our phantom measurements, MIP accurately reflects the range of target motion for regular tar-get motion. However, it generally underestimates the range of target motion when the motion is irregular in amplitude and periodicity. Clinical internal target volume determination using MIP requires caution, especially when there is breathing irregularity. (C) 2009 Elsevier Inc.
引用
收藏
页码:618 / 625
页数:8
相关论文
共 18 条
[1]   Comparison of helical, maximum intensity projection (MIP), and averaged intensity (Al) 4D CT imaging for stereotactic body radiation therapy (SBRT) planning in lung cancer [J].
Bradley, Jeffrey D. ;
Nofal, Ahmed N. ;
El Naqa, Issam M. ;
Lu, Wei ;
Liu, Jubei ;
Hubenschmidt, James ;
Low, Daniel A. ;
Drzymala, Robert E. ;
Khullar, Divya .
RADIOTHERAPY AND ONCOLOGY, 2006, 81 (03) :264-268
[2]   Estimation of error in maximal intensity projection-based internal target volume of lung tumors: A simulation and comparison study using dynamic magnetic resonance imaging [J].
Cai, Jing ;
Read, Paul W. ;
Baisden, Joseph M. ;
Larner, James M. ;
Benedict, Stanley H. ;
Sheng, Ke .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03) :895-902
[3]   Displacement-based binning of time-dependent computed tomography image data sets [J].
Fitzpatrick, MJ ;
Starkschall, G ;
Antolak, JA ;
Fu, J ;
Shukla, H ;
Keall, PJ ;
Klahr, P ;
Mohan, R .
MEDICAL PHYSICS, 2006, 33 (01) :235-246
[4]   Acquiring 4D thoracic CT scans using a multislice helical method [J].
Keall, PJ ;
Starkschall, G ;
Shukla, H ;
Forster, KM ;
Ortiz, V ;
Stevens, CW ;
Vedam, SS ;
George, R ;
Guerrero, T ;
Mohan, R .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (10) :2053-2067
[5]   The impact of temporal inaccuracies on 4DCT image quality [J].
Mutaf, Y. D. ;
Antolak, J. A. ;
Brinkmann, D. H. .
MEDICAL PHYSICS, 2007, 34 (05) :1615-1622
[6]   Improvement of the cine-CT based 4D-CT imaging [J].
Pan, Tinsu ;
Sun, Xiaojun ;
Luo, Dershan .
MEDICAL PHYSICS, 2007, 34 (11) :4499-4503
[7]   Four-dimensional computed tomography: Image formation and clinical protocol [J].
Rietzel, E ;
Pan, TS ;
Chen, GTY .
MEDICAL PHYSICS, 2005, 32 (04) :874-889
[8]   Design of 4D treatment planning target volumes [J].
Rietzel, Eike ;
Liu, Arthur K. ;
Doppke, Karen P. ;
Wolfgang, John A. ;
Chen, Aileen B. ;
Chen, George T. Y. ;
Choi, Noah C. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (01) :287-295
[9]   Impact of respiratory movement on the computed tomographic images of small lung tumors in three-dimensional (3D) radiotherapy [J].
Shimizu, S ;
Shirato, H ;
Kagei, K ;
Nishioka, T ;
Bo, X ;
Dosaka-Akita, H ;
Hashimoto, S ;
Aoyama, H ;
Tsuchiya, K ;
Miyasaka, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (05) :1127-1133
[10]   Initial evaluation of a four-dimensional computed tomography system, using a programmable motor [J].
Simon, Luc ;
Giraud, Philippe ;
Servois, Vincent ;
Rosenwald, Jean-Claude .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (04) :50-65