Development of a dynamic quality assurance testing protocol for multisite clinical trial DCE-CT accreditation

被引:18
作者
Driscoll, B. [1 ]
Keller, H. [1 ,2 ]
Jaffray, D. [1 ,2 ,3 ]
Coolens, C. [1 ,2 ,3 ]
机构
[1] Princess Margaret Canc Ctr, Dept Radiat Phys, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5S 3E2, Canada
[3] Univ Hlth Network, Techna Inst, Toronto, ON M5G 1L5, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
DCE-CT; quality assurance; SQUAMOUS-CELL CARCINOMAS; PERFUSION CT; MONOCLONAL-ANTIBODY; TUMOR PERFUSION; BEVACIZUMAB; INSTITUTIONS; PARAMETERS; THERAPY; CANCER; VOLUME;
D O I
10.1118/1.4812429
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Credentialing can have an impact on whether or not a clinical trial produces useful quality data that is comparable between various institutions and scanners. With the recent increase of dynamic contrast enhanced-computed tomography (DCE-CT) usage as a companion biomarker in clinical trials, effective quality assurance, and control methods are required to ensure there is minimal deviation in the results between different scanners and protocols at various institutions. This paper attempts to address this problem by utilizing a dynamic flow imaging phantom to develop and evaluate a DCE-CT quality assurance (QA) protocol. Methods: A previously designed flow phantom, capable of producing predictable and reproducible time concentration curves from contrast injection was fully validated and then utilized to design a DCE-CT QA protocol. The QA protocol involved a set of quantitative metrics including injected and total mass error, as well as goodness of fit comparison to the known truth concentration curves. An additional region of interest (ROI) sensitivity analysis was also developed to provide additional details on intrascanner variability and determine appropriate ROI sizes for quantitative analysis. Both the QA protocol and ROI sensitivity analysis were utilized to test variations in DCE-CT results using different imaging parameters (tube voltage and current) as well as alternate reconstruction methods and imaging techniques. The developed QA protocol and ROI sensitivity analysis was then applied at three institutions that were part of clinical trial involving DCE-CT and results were compared. Results: The inherent specificity of robustness of the phantom was determined through calculation of the total intraday variability and determined to be less than 2.2 +/- 1.1% (total calculated output contrast mass error) with a goodness of fit (R-2) of greater than 0.99 +/- 0.0035 (n = 10). The DCE-CT QA protocol was capable of detecting significant deviations from the expected phantom result when scanning at low mAs and low kVp in terms of quantitative metrics (Injected Mass Error 15.4%), goodness of fit (R-2) of 0.91, and ROI sensitivity (increase in minimum input function ROI radius by 146 +/- 86%). These tests also confirmed that the ASIR reconstruction process was beneficial in reducing noise without substantially increasing partial volume effects and that vendor specific modes (e.g., axial shuttle) did not significantly affect the phantom results. The phantom and QA protocol were finally able to quickly (<90 min) and successfully validate the DCE-CT imaging protocol utilized at the three separate institutions of a multicenter clinical trial; thereby enhancing the confidence in the patient data collected. Conclusions: A DCE QA protocol was developed that, in combination with a dynamic multimodality flow phantom, allows the intrascanner variability to be separated from other sources of variability such as the impact of injection protocol and ROI selection. This provides a valuable resource that can be utilized at various clinical trial institutions to test conformance with imaging protocols and accuracy requirements as well as ensure that the scanners are performing as expected for dynamic scans. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:14
相关论文
共 36 条
[1]   Perfusion CT in Acute lschemic Stroke: A Qualitative and Quantitative Comparison of Deconvolution and Maximum Slope Approach [J].
Abels, B. ;
Klotz, E. ;
Tomandl, B. F. ;
Kloska, S. P. ;
Lell, M. M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (09) :1690-1698
[2]   Whole-tumor perfusion CT parameters and glucose metabolism measurements in head and neck squamous cell carcinomas: A pilot study using combined positron-emission tomography/CT imaging [J].
Bisdas, S. ;
Spicer, K. ;
Rumboldt, Z. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2008, 29 (07) :1376-1381
[3]   Description and implementation of a quality control program in an imaging-based clinical trial [J].
Cagnon, Christopher H. ;
Cody, Dianna D. ;
McNitt-Gray, Michael F. ;
Seibert, J. Anthony ;
Judy, Philip F. ;
Aberle, Denise R. .
ACADEMIC RADIOLOGY, 2006, 13 (11) :1431-1441
[4]   Development of a dynamic flow imaging phantom for dynamic contrast-enhanced CT [J].
Driscoll, B. ;
Keller, H. ;
Coolens, C. .
MEDICAL PHYSICS, 2011, 38 (08) :4866-4880
[5]   BW12C PERTURBS NORMAL AND TUMOR-TISSUE OXYGENATION AND BLOOD-FLOW IN MAN [J].
FALK, SJ ;
RAMSAY, JR ;
WARD, R ;
MILES, K ;
DIXON, AK ;
BLEEHEN, NM .
RADIOTHERAPY AND ONCOLOGY, 1994, 32 (03) :210-217
[6]   CT quantification of effects of thalidomide in patients with metastatic renal cell carcinoma [J].
Faria, Silvana C. ;
Ng, Chaan S. ;
Hess, Kenneth R. ;
Phongkitkarun, Sith ;
Szejnfeld, Jacob ;
Daliani, Danai ;
Charnsangavej, Chusilp .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (02) :378-385
[7]   What We Have Learned: The Impact of Quality From a Clinical Trials Perspective [J].
FitzGerald, Thomas J. .
SEMINARS IN RADIATION ONCOLOGY, 2012, 22 (01) :18-28
[8]  
Ford JM, 1996, QUANTITATIVE IMAGING IN ONCOLOGY, P1
[9]   Metastatic Renal Carcinoma: Evaluation of Antiangiogenic Therapy with Dynamic Contrast-enhanced CT [J].
Fournier, Laure S. ;
Oudard, Stephane ;
Thiam, Rokhaya ;
Trinquart, Ludovic ;
Banu, Eugeniu ;
Medioni, Jacques ;
Balvay, Daniel ;
Chatellier, Gilles ;
Frija, Guy ;
Cuenod, Charles A. .
RADIOLOGY, 2010, 256 (02) :511-518
[10]  
Gandhi D, 2006, AM J NEURORADIOL, V27, P101