A Novel Thermal Sensor for the Sensitive Measurement of Chemical Oxygen Demand

被引:18
作者
Yao, Na [1 ,2 ]
Liu, Zhuan [1 ,3 ]
Chen, Ying [4 ]
Zhou, Yikai [1 ]
Xie, Bin [5 ]
机构
[1] Huazhong Univ Sci & Technol, Tongji Med Coll, Inst Environm Med, Wuhan 430030, Peoples R China
[2] Jingzhou Cent Hosp, Jingzhou 434026, Peoples R China
[3] Yancheng Blood Ctr, Yancheng 224005, Peoples R China
[4] Hubei Ctr Dis Control & Prevent, Wuhan 430079, Peoples R China
[5] Lund Univ, Pure & Appl Biochem, SE-22100 Lund, Sweden
关键词
chemical oxygen demand; thermal sensor; flow injection analysis system; sodium hypochlorite; FLOW-INJECTION ANALYSIS; PHOTOCATALYTIC OXIDATION; ENZYME THERMISTOR; COPPER ELECTRODE; BIOSENSOR;
D O I
10.3390/s150820501
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel rapid methodology for determining the chemical oxygen demand (COD) based on a thermal sensor with a flow injection analysis system was proposed and experimentally validated. The ability of this sensor to detect and monitor COD was based on the degree of enthalpy increase when sodium hypochlorite reacted with the organic content in water samples. The measurement results were correlated with COD and were compared against the conventional method using potassium dichromate. The assay required only 5-7 min rather than the 2 h required for evaluation by potassium dichromate. The linear range was 5-1000 mg/L COD, and the limit of detection was very low, 0.74 mg/L COD. Moreover, this method exhibited high tolerance to chloride ions; 0.015 mol/L chloride ions had no influence on the response. Finally, the sensor was used to detect the COD of different water samples; the results were verified by the standard dichromate method.
引用
收藏
页码:20501 / 20510
页数:10
相关论文
共 25 条
[1]   Fast determination of antibiotics in whole blood [J].
Chen, Q. ;
Andersson, A. ;
Mecklenburg, M. ;
Xie, B. .
CLINICAL MICROBIOLOGY AND INFECTION, 2013, 19 (09) :869-874
[2]   Determination of chemical oxygen demand in fresh waters using flow injection with on-line UV-photocatalytic oxidation and spectrophotometric detection [J].
Dan, DZ ;
Sandford, RC ;
Worsfold, PJ .
ANALYST, 2005, 130 (02) :227-232
[3]   FLOW-INJECTION ANALYSIS WITH ENZYME THERMISTOR DETECTOR FOR AUTOMATED-DETERMINATION OF BETA-LACTAMS [J].
DECRISTOFORO, G ;
DANIELSSON, B .
ANALYTICAL CHEMISTRY, 1984, 56 (02) :263-268
[4]   Chemical oxygen demand using closed microwave digestion system [J].
Dharmadhikari, DM ;
Vanerkar, AP ;
Barhate, NM .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2005, 39 (16) :6198-6201
[5]   Photocatalytic sensor for the determination of chemical oxygen demand using flow injection analysis [J].
Kim, YC ;
Sasaki, S ;
Yano, K ;
Ikebukuro, K ;
Hashimoto, K ;
Karube, I .
ANALYTICA CHIMICA ACTA, 2001, 432 (01) :59-66
[6]   Photocatalytic sensor for chemical oxygen demand determination based on oxygen electrode [J].
Kim, YC ;
Lee, KH ;
Sasaki, S ;
Hashimoto, K ;
Ikebukuro, K ;
Karube, I .
ANALYTICAL CHEMISTRY, 2000, 72 (14) :3379-3382
[7]   A flow method with photocatalytic oxidation of dissolved organic matter using a solid-phase (TiO2) reactor followed by amperometric detection of consumed oxygen [J].
Kim, YC ;
Sasaki, S ;
Yano, K ;
Ikebukuro, K ;
Hashimoto, K ;
Karube, I .
ANALYTICAL CHEMISTRY, 2002, 74 (15) :3858-3864
[8]  
Lee KH, 1999, ELECTROANAL, V11, P1172, DOI 10.1002/(SICI)1521-4109(199911)11:16<1172::AID-ELAN1172>3.0.CO
[9]  
2-J
[10]   Evaluation of chemical oxygen demand (COD) based on coulometric determination of electrochemical oxygen demand (EOD) using a surface oxidized copper electrode [J].
Lee, KH ;
Ishikawa, T ;
McNiven, SJ ;
Nomura, Y ;
Hiratsuka, A ;
Sasaki, S ;
Arikawa, Y ;
Karube, I .
ANALYTICA CHIMICA ACTA, 1999, 398 (2-3) :161-171