HeartWare Ventricular Assist Device Cannula Position and Hemocompatibility-Related Adverse Events

被引:6
作者
Imamura, Teruhiko [1 ,2 ]
Narang, Nikhil [3 ]
Nitta, Daisuke [1 ]
Fujino, Takeo [1 ]
Nguyen, Ann [1 ]
Chung, Ben [1 ]
Holzhauser, Luise [1 ]
Kim, Gene [1 ]
Raikhelkar, Jayant [4 ]
Kalantari, Sara [1 ]
Smith, Bryan [1 ]
Juricek, Colleen [5 ]
Rodgers, Daniel [1 ]
Ota, Takeyoshi [5 ]
Song, Tae [5 ]
Jeevanandam, Valluvan [5 ]
Sayer, Gabriel [4 ]
Uriel, Nir [4 ]
机构
[1] Univ Chicago, Med Ctr, Dept Med, 5841 S Maryland Ave, Chicago, IL 60637 USA
[2] Univ Toyama, Dept Med 2, Toyama, Japan
[3] Advocate Christ Med Ctr, Advocate Heart Inst, Oak Lawn, IL USA
[4] Columbia Univ, Div Cardiol, Irving Med Ctr, New York, NY USA
[5] Univ Chicago, Med Ctr, Dept Surg, Chicago, IL 60637 USA
关键词
PUMP; GUIDELINES; SUPPORT; SOCIETY; RISK;
D O I
10.1016/j.athoracsur.2019.12.049
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. HeartWare ventricular assist device (HVAD) cannula position is associated with hemodynamics and heart failure readmissions. However, its impact on hemocompatibility-related adverse events (HRAEs) remains uncertain. Methods. HVAD patients were followed for 1 year after index hospitalization, when cannula coronal angle was quantified from chest x-ray film. Invasive right heart catheterization and transthoracic echocardiography were performed. One-year occurrences of each HRAE were compared between those with and without a cannula coronal angle of greater than 65 degrees. Results. Among 63 HVAD patients (median age 60 years, 63% male), 10 (16%) had a cannula coronal angle greater than 65 degrees. The wide-angle group had elevated intracardiac pressures and lower pulmonary artery pulsatility index (P <.05). They also had reduced right ventricular function by echocardiography. Freedom from HRAEs tended to be lower in the wide-angle group (24% vs 62%; P = .11). The rate of gastrointestinal bleeding was significantly higher in the greater than 65 degrees group (0.90 events/year vs 0.40 events/year; P = .013). The rates of stroke and pump thrombosis were statistically comparable irrespective of cannula angle (P >.05). Conclusions. HVAD cannula coronal angle was associated with reduced right ventricular function and HRAEs. Prospective studies evaluating surgical techniques to ensure optimal device positioning and its effects on HRAEs are warranted. (C) 2020 by The Society of Thoracic Surgeons.
引用
收藏
页码:911 / 917
页数:7
相关论文
共 50 条
  • [31] International Normalized Ratio Test Frequency in Left Ventricular Assist Device Patients Affects Anticoagulation Quality and Adverse Events
    Schloglhofer, Thomas
    Zapusek, Lydia
    Wiedemann, Dominik
    Riebandt, Julia
    Wittmann, Franziska
    Dimitrov, Kamen
    Angleitner, Philipp
    Haberl, Lisa
    Laufer, Gunther
    Moscato, Francesco
    Zimpfer, Daniel
    Schima, Heinrich
    ASAIO JOURNAL, 2021, 67 (02) : 157 - 162
  • [32] First results of HeartWare left ventricular assist device implantation with tunnelling of the outflow graft through the transverse sinus
    Hanke, Jasmin S.
    Rojas, Sebastian V.
    Cvitkovic, Tomislav
    Wiegmann, Bettina
    Horke, Alexander
    Warnecke, Gregor
    Haverich, Axel
    Schmitto, Jan D.
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2017, 25 (04) : 503 - 508
  • [33] HeartWare Left Ventricular Assist Device Exercise Hemodynamics With Speed Adjustment Based on Left Ventricular Filling Pressures
    Ali, Omaima
    Arnold, Amy C.
    Cysyk, Joshua
    Boehmer, John
    Zhu, Junjia
    Sinoway, Lawrence I.
    Eisen, Howard
    Weiss, William
    ASAIO JOURNAL, 2024, 70 (06) : e82 - e88
  • [34] HeartWare Miniature Axial-Flow Ventricular Assist Device Design and Initial Feasibility Test
    Slaughter, Mark S.
    Sobieski, Michael A., II
    Tamez, Daniel
    Horrell, Timothy
    Graham, Joel
    Pappas, Patroklos S.
    Tatooles, Antone J.
    LaRose, Jeffrey
    TEXAS HEART INSTITUTE JOURNAL, 2009, 36 (01): : 12 - 16
  • [35] Comparison of the Hemocompatibility of an Axial and a Centrifugal Left Ventricular Assist Device in an In Vitro Test Circuit
    Borchers, Patrick
    Winnersbach, Patrick
    Kraemer, Sandra
    Beckers, Christian
    Buhl, Eva Miriam
    Leonhardt, Steffen
    Rossaint, Rolf
    Walter, Marian
    Breuer, Thomas
    Bleilevens, Christian
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (12)
  • [36] Adverse events after left ventricular assist device implantation linked to psychosocial risk in women and men
    Maukel, Lisa-Marie
    Weidner, Gerdi
    Beyersmann, Jan
    Spaderna, Heike
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2023, 42 (11) : 1557 - 1568
  • [37] Clinical Experience of HeartMate II to HeartWare Left Ventricular Assist Device Exchange: A Multicenter Experience
    Agarwal, Richa
    Kyvernitakis, Andreas
    Soleimani, Behzad
    Milano, Carmelo A.
    Davis, Robert Patrick
    Kennedy, Jamie L. W.
    Yarboro, Leora
    Benza, Raymond L.
    Moraca, Robert J.
    Bailey, Stephen H.
    ANNALS OF THORACIC SURGERY, 2019, 108 (04) : 1178 - 1182
  • [38] Preclinical evaluation of the fluid dynamics and hemocompatibility of the Corheart 6 left ventricular assist device
    Fang, Peng
    Yang, Yuzhuo
    Wei, Xufeng
    Yu, Shunzhou
    ARTIFICIAL ORGANS, 2023, 47 (06) : 951 - 960
  • [39] Blood Pressure Control in Continuous Flow Left Ventricular Assist Devices: Efficacy and Impact on Adverse Events
    Lampert, Brent C.
    Eckert, Chad
    Weaver, Stephanie
    Scanlon, Ana
    Lockard, Kathy
    Allen, Carrie
    Kunz, Nicole
    Bermudez, Christian
    Bhama, Jay K.
    Shullo, Michael A.
    Kormos, Robert L.
    Dew, Mary Amanda
    Teuteberg, Jeffrey J.
    ANNALS OF THORACIC SURGERY, 2014, 97 (01) : 139 - 146
  • [40] HeartWare Ventricular Assist Device as a Bridge-to-Transplant in a Small Boy with Complicated Kawasaki Disease
    Mascio, Christopher E.
    Malankar, Dhananjay P.
    Rome, Jonathan J.
    ASAIO JOURNAL, 2018, 64 (03) : E37 - E39