Reduction of radiation exposure by using storage phosphor radiography on pelvis and lumbar spine

被引:6
作者
Heyne, JR [1 ]
Sehner, J [1 ]
Neumann, R [1 ]
Werner, B [1 ]
Adler, R [1 ]
Freesmeyer, M [1 ]
Kaiser, WA [1 ]
机构
[1] Univ Jena, Inst Diagnost & Intervent Radiol, D-07740 Jena, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2002年 / 174卷 / 01期
关键词
storage phosphor radiography; radiation dose; dose reduction; pelvis; lumbar spine;
D O I
10.1055/s-2002-19543
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To examine the reduction of radiation dose for radiography of the pelvis and lumbar spine (LS) by using storage phosphor radiography (SPR)? Material and Methods: Each pelvis and LS of 2 phantoms (3 M) and 2 corpses were X-raxed digitally in an anterior-posterior orientation using SPR (ADC compact, Agfa). The tube current time product (mAs) was diminished gradually while keeping the voltage constant. The surface entrance dose was measured by a Dosimax (Wellhofer) sensor. Five investigators evaluated the images for characteristics and critical features and pathological findings. Results: The images of the 4 various pelvises/LS could be evaluated sufficiently down to an average dose of 34 to 40 %/18 to 35 %. For the assessment of hip joint, region typical corticalis and calcifications (pelvis/LS) 9 to 18 %/8 to 23 % of the initial dose (initial dose is equivalent to a conventional screen film system with a speed class of 400) are enough, spongiosa typical for the region and the sacrum 30 to 56 %/19 to 36 % on average among the readers. Conclusion: For the sufficient exclusion of a fracture the dose can be reduced on pelvis/LS to at least 39 %/28 % (diameter 23 %/17 %), for the assessment of sacroiliitis/ osteolyses/demineralisation to at least 78 %/44 % (diameter 34 to 36 %/ 22 to 26 %), to position check after fracture to at least 15 %/25 % (diameter 9 %/10 %). Question-related tube current time products of the pelvis and LS can be deduced for various objects.
引用
收藏
页码:104 / 111
页数:8
相关论文
共 23 条
[1]  
BEDNAR W, 1980, STRAHLENTHER ONKOL, V156, P776
[2]  
Bernhardt JU, 1995, Z MED PHYS, V5, P33
[3]  
BRAUNSCHWEIG R, 1994, UNFALLCHIRURG, V97, P337
[4]   PHOSPHOR PLATE COMPUTED RADIOGRAPHY - RESPONSE TO VARIATION IN MAS AT FIXED KVP IN AN ANIMAL-MODEL - POTENTIAL ROLE IN NEONATAL IMAGING [J].
BRODERICK, NJ ;
LONG, B ;
DREESEN, RG ;
COHEN, MD ;
CORY, DA ;
KATZ, BP ;
KALASINSKI, LA .
CLINICAL RADIOLOGY, 1993, 47 (01) :39-45
[5]  
Busch H P, 1991, Rontgenpraxis, V44, P329
[6]  
Busch H P, 1997, Aktuelle Radiol, V7, P56
[7]   Digital vs conventional radiography: cost and revenue analysis [J].
Dalla Palma, L ;
Grisi, G ;
Cuttin, R ;
Rimondini, A .
EUROPEAN RADIOLOGY, 1999, 9 (08) :1682-1692
[8]   A comparison of digital luminescence radiography and conventional film/screen techniques in the diagnosis of fractures [J].
Fink, U ;
Schwerdtfeger, J ;
Fink, BK ;
Schatzl, M ;
Reiser, M .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1996, 164 (04) :275-280
[9]  
Gocke C, 1996, Aktuelle Radiol, V6, P78
[10]   Reduction of radiation dose by using digital luminescence radiography compared to conventional screen film system with grid cassette. [J].
Heyne, JP ;
Merbold, H ;
Sehner, J ;
Neumann, R ;
Freesmeyer, M ;
Jonetz-Mentzel, L ;
Kaiser, WA .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1999, 171 (01) :54-59