On the property of measurements with the PTW microLion chamber in continuous beams

被引:10
作者
Andersson, Jonas [1 ]
Johansson, Erik [2 ]
Tolli, Heikki [1 ]
机构
[1] Umea Univ, Dept Radiat Sci, SE-90185 Umea, Sweden
[2] FOI CBRN Def & Secur, Swedish Def Res Agcy, SE-90182 Umea, Sweden
关键词
general recombination; continuous beam; two-dose-rate method; liquid ionization chamber; radiation dosimetry; isooctane; tetramethylsilane; perturbation; ambient electric field; LIQUID IONIZATION CHAMBERS; GENERAL RECOMBINATION CORRECTION; COLLECTION EFFICIENCY; 2-DOSE-RATE METHOD; TETRAMETHYLSILANE; ISOOCTANE; DOSIMETRY; WATER;
D O I
10.1118/1.4736804
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The performance of liquid ionization chambers, which may prove to be useful tools in the field of radiation dosimetry, is based on several chamber and liquid specific characteristics. The present work investigates the performance of the PTW microLion liquid ionization chamber with respect to recombination losses and perturbations from ambient electric fields at various dose rates in continuous beams. Methods: In the investigation, experiments were performed using two microLion chambers, containing isooctane (C8H18) and tetramethylsilane [Si(CH3)(4)] as the sensitive media, and a NACP-02 monitor chamber. An initial activity of approximately 250 GBq F-18 was employed as the radiation source in the experiments. The initial dose rate in each measurement series was estimated to 1.0 Gy min(-1) by Monte Carlo simulations and the measurements were carried out during the decay of the radioactive source. In the investigation of general recombination losses, employing the two-dose-rate method for continuous beams, the liquid ionization chambers were operated at polarizing voltages 25, 50, 100, 150, 200, and 300 V. Furthermore, measurements were also performed at 500 V polarizing voltage in the investigation of the sensitivity of the microLion chamber to ambient electric fields. Results: The measurement results from the liquid ionization chambers, corrected for general recombination losses according to the two-dose-rate method for continuous beams, had a good agreement with the signal to dose linearity from the NACP-02 monitor chamber for general collection efficiencies above 70%. The results also displayed an agreement with the theoretical collection efficiencies according to the Greening theory, except for the liquid ionization chamber containing isooctane operated at 25 V. At lower dose rates, perturbations from ambient electric fields were found in the microLion chamber measurement results. Due to the perturbations, measurement results below an estimated dose rate of 0.2 Gy min(-1) were excluded from the present investigation of the general collection efficiency. The perturbations were found to be more pronounced when the chamber polarizing voltage was increased. Conclusions: By using the two-dose-rate method for continuous beams, comparable corrected ionization currents from experiments in low and medium energy photon beams can be achieved. However, the valid range of general collection efficiencies has been found to vary in a comparison between experiments performed in continuous beams of 120 kVp x ray, and the present investigation of 511 keV annihilation photons. At very high dose rates in continuous beams, there are presently no methods that can be used to correct for general recombination losses and at low dose rates the microLion chamber may be perturbed by ambient electric fields. Increasing the chamber polarizing voltage, which diminishes the general recombination effect, was found to increase the microLion chamber sensitivity to ambient electric fields. Prudence is thus advised when employing the microLion chamber in radiation dosimetry, as ambient electric fields of the strength observed in the present work may be found in many common situations. Due to uncertainties in the theoretical basis for recombination losses in liquids, further studies on the underlying theories for the initial and general recombination effect are needed if liquid ionization chambers are to become a viable option in high precision radiation dosimetry. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4736804]
引用
收藏
页码:4775 / 4787
页数:13
相关论文
共 22 条
[1]   Application of the two-dose-rate method for general recombination correction for liquid ionization chambers in continuous beams [J].
Andersson, Jonas ;
Tolli, Heikki .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (02) :299-314
[2]   A quantitative and comparative study of radionuclidic and chemical impurities in water samples irradiated in a niobium target with Havar vs. niobium-sputtered Havar as entrance foils [J].
Avila-Rodriguez, Miguel A. ;
Wilson, John S. ;
McQuarrie, Steve A. .
APPLIED RADIATION AND ISOTOPES, 2008, 66 (12) :1775-1780
[3]   Assessment of the relative dose distribution around an 192Ir line source using a liquid ionization chamber [J].
Bahar-Gogani, J ;
Wickman, G ;
Johansson, L ;
Johansson, BE .
MEDICAL PHYSICS, 1999, 26 (09) :1932-1942
[4]   Radionuclide impurities in proton-irradiated [18O]H2O for the production of 18F-: Activities and distribution in the [18F]FDG synthesis process [J].
Bowden, Louise ;
Vintro, Luis Leon ;
Mitchell, Peter I. ;
O'Donnell, Ruairi G. ;
Seymour, Anne Marie ;
Duffy, George J. .
APPLIED RADIATION AND ISOTOPES, 2009, 67 (02) :248-255
[5]  
Castano D. M. Gonzalez, 2010, THESIS U SANTIAGO CO
[6]  
Chu S.Y.F., 1999, Table of Radioactive Isotopes, Database Version 2.0
[7]   Dose homogeneity specification for reference dosimetry of nonstandard fields [J].
Chung, Eunah ;
Soisson, Emilie ;
Seuntjens, Jan .
MEDICAL PHYSICS, 2012, 39 (01) :407-414
[8]   Liquid ionization chamber measurements of dose distributions in small 6 MV photon beams [J].
Dasu, A ;
Lofroth, PO ;
Wickman, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (01) :21-36
[9]   First tests of a liquid ionization chamber to monitor intensity modulated radiation beams [J].
Eberle, K ;
Engler, J ;
Hartmann, G ;
Hofmann, R ;
Hörandel, JR .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (21) :3555-3564
[10]   Liquid-filled ionization chamber temperature dependence [J].
Franco, L. ;
Gomez, F. ;
Iglesias, A. ;
Pardo, J. ;
Pazos, A. ;
Pena, J. ;
Zapata, M. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 560 (02) :584-588