Reflections on Pediatric High-Frequency Oscillatory Ventilation From a Physiologic Perspective

被引:33
作者
Kneyber, Martin C. J. [1 ,2 ]
van Heerde, Marc [2 ]
Markhorst, Dick G. [2 ]
机构
[1] Univ Groningen, Univ Med Ctr Groningen, Beatrix Childrens Hosp, Div Pediat Intens Care,Dept Pediat, NL-9700 RB Groningen, Netherlands
[2] Vrije Univ Amsterdam Med Ctr, Dept Pediat, Div Pediat Intens Care, Amsterdam, Netherlands
关键词
HFOV; ALI/ARDS; obstructive airway disease; oxygenation; ventilation; MEAN AIRWAY PRESSURE; ACUTE LUNG INJURY; HYPOXEMIC RESPIRATORY-FAILURE; OPTIMAL DISTENDING PRESSURE; TIDAL VOLUME; ENDOTRACHEAL-TUBE; GAS-EXCHANGE; ALVEOLAR PRESSURE; DISTRESS-SYNDROME; CO2; ELIMINATION;
D O I
10.4187/respcare.01571
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Mechanical ventilation using low tidal volumes has become universally accepted to prevent ventilator-induced lung injury. High-frequency oscillatory ventilation (HFOV) allows pulmonary gas exchange using very small tidal volume (1-2 mL/kg) with concomitant decreased risk of atelectrauma. However, its use in pediatric critical care varies between only 3% and 30% of all ventilated children. This might be explained by the fact that the beneficial effect of HFOV on patient outcome has not been ascertained. Alternatively, in contrast with present recommendations, one can ask if HFOV has been employed in its most optimal fashion related especially to the indications for and timing of HFOV, as well as to using the best oscillator settings. The first was addressed in one small randomized study showing that early use of HFOV, instead of rescue use, was associated with improved survival. From a physiologic perspective, the oscillator settings could be refined. Lung volume is the main determinant of oxygenation in diffuse alveolar disease, suggesting using an open-lung strategy by recruitment maneuvers, although this is in practice not custom. Using such an approach, the patient can be oscillated on the deflation limb of the pressure-volume (P-V) curve, allowing less pressure required to maintain a certain amount of lung volume. Gas exchange is determined by the frequency and the oscillatory power setting, controlling the magnitude of the membrane displacement. Experimental work as well as preliminary human data have shown that it is possible to achieve the smallest tidal volume with concomitant adequate gas exchange when oscillating at high frequency and high fixed power setting. Future studies are needed to validate these novel approaches and to evaluate their effect on patient outcome.
引用
收藏
页码:1496 / 1504
页数:9
相关论文
共 88 条
[1]   Inability to predict outcome of acute respiratory distress syndrome in children when using high frequency oscillation [J].
Anton, N ;
Joffe, KM ;
Joffe, AR .
INTENSIVE CARE MEDICINE, 2003, 29 (10) :1763-1769
[2]   PROSPECTIVE, RANDOMIZED COMPARISON OF HIGH-FREQUENCY OSCILLATORY VENTILATION AND CONVENTIONAL MECHANICAL VENTILATION IN PEDIATRIC RESPIRATORY-FAILURE [J].
ARNOLD, JH ;
HANSON, JH ;
TOROFIGUERO, LO ;
GUTIERREZ, J ;
BERENS, RJ ;
ANGLIN, DL .
CRITICAL CARE MEDICINE, 1994, 22 (10) :1530-1539
[3]   High-frequency oscillatory ventilation in pediatric respiratory failure: A multicenter experience [J].
Arnold, JH ;
Anas, NG ;
Luckett, P ;
Cheifetz, IM ;
Reyes, G ;
Newth, CJL ;
Kocis, KC ;
Heidemann, SM ;
Hanson, JH ;
Brogan, TV ;
Bohn, DJ .
CRITICAL CARE MEDICINE, 2000, 28 (12) :3913-3919
[4]   High-frequency oscillation and surfactant treatment in an acid aspiration model [J].
Aspros, Alexander J. ;
Coto, Claudia G. ;
Lewis, James F. ;
Veldhuizen, Ruud A. W. .
CANADIAN JOURNAL OF PHYSIOLOGY AND PHARMACOLOGY, 2010, 88 (01) :14-20
[5]   The role of ventilation frequency in airway reopening [J].
Bauer, K. ;
Bruecker, Ch. .
JOURNAL OF BIOMECHANICS, 2009, 42 (08) :1108-1113
[6]   High-frequency oscillatory ventilation in pediatric patients with acute respiratory failure [J].
Ben Jaballah, Nejla ;
Khaldi, Ammar ;
Mnif, Khaled ;
Bouziri, Asma ;
Belhadj, Sarra ;
Hamdi, Asma ;
Kchaou, Wassim .
PEDIATRIC CRITICAL CARE MEDICINE, 2006, 7 (04) :362-367
[7]   High frequency oscillatory ventilation for respiratory failure due to RSV bronchiolitis [J].
Berner, Michel E. ;
Hanquinet, Sylviane ;
Rimensberger, Peter C. .
INTENSIVE CARE MEDICINE, 2008, 34 (09) :1698-1702
[8]   VENTILATION BY HIGH-FREQUENCY OSCILLATION [J].
BOHN, DJ ;
MIYASAKA, K ;
MARCHAK, BE ;
THOMPSON, WK ;
FROESE, AB ;
BRYAN, AC .
JOURNAL OF APPLIED PHYSIOLOGY, 1980, 48 (04) :710-716
[9]   VOLUME RECRUITMENT MANEUVERS ARE LESS DELETERIOUS THAN PERSISTENT LOW LUNG-VOLUMES IN THE ATELECTASIS-PRONE RABBIT LUNG DURING HIGH-FREQUENCY OSCILLATION [J].
BOND, DM ;
FROESE, AB .
CRITICAL CARE MEDICINE, 1993, 21 (03) :402-412
[10]   SUSTAINED INFLATIONS IMPROVE RESPIRATORY COMPLIANCE DURING HIGH-FREQUENCY OSCILLATORY VENTILATION BUT NOT DURING LARGE TIDAL VOLUME POSITIVE-PRESSURE VENTILATION IN RABBITS [J].
BOND, DM ;
MCALOON, J ;
FROESE, AB .
CRITICAL CARE MEDICINE, 1994, 22 (08) :1269-1277