Experimental validation of finite element model for proximal composite femur using optical measurements

被引:65
作者
Grassi, Lorenzo [1 ,2 ]
Vaananen, Sami P. [3 ]
Yavari, Saber Amin [4 ]
Weinans, Harrie [4 ,5 ]
Jurvelin, Jukka S. [3 ]
Zadpoor, Amir A. [4 ]
Isaksson, Hanna [1 ,2 ]
机构
[1] Lund Univ, Div Solid Mech, Lund, Sweden
[2] Lund Univ, Dept Orthoped, Lund, Sweden
[3] Univ Eastern Finland, Dept Appl Phys, Kuopio, Finland
[4] Delft Univ Technol, Fac Mech Maritime & Mat Engn, Delft, Netherlands
[5] Erasmus MC, Dept Orthoped, Rotterdam, Netherlands
基金
瑞典研究理事会; 芬兰科学院;
关键词
Digital image correlation; Composite femur; Finite element; Bone fracture; DIGITAL-IMAGE-CORRELATION; STRAIN DISTRIBUTION; HIP-FRACTURES; BONE; OSTEOPOROSIS; ACCURACY; STRENGTH; MECHANICS; DENSITY; WOMEN;
D O I
10.1016/j.jmbbm.2013.02.006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Patient-specific finite element models have been used to predict femur strength and fracture risk in individuals. Validation of the adopted finite element modelling procedure against mechanical testing data is a crucial step when aiming for clinical applications. The majority of the works available in literature used data from strain gages to validate the model, thus having up to 15 experimental measurements. Optical techniques, such as digital image correlation, can help to improve the models by providing a continuous field of deformation data over a femoral surface. The main objective of this study was to validate finite element models of six composite femora against strain data from digital image correlation, obtained during fracture tests performed in quasi-axial loading configuration. The finite element models were obtained from CT scans, by means of a semi-automatic segmentation. The principal strains both during the elastic phase and close to the fracture were compared, and showed a correlation coefficient close to 0.9. In the linear region, the slope and intercept were close to zero and unity, while for the case when fracture load was simulated, the slope decreased somewhat. The accuracy of the obtained results is comparable with the state-of-the-art literature, with the significant improvement of having around 50,000 data points for each femur. This large number of measurements allows a more comprehensive validation of the predictions by the finite element models, since thousand of points are tracked along the femoral neck and trochanter region, i.e., the sites that are most critical for femur fracture. Moreover, strain measurement biases due to the strain gage reinforcement effect, were avoided. The combined experimental numerical approach proved to be ready for application to in-vitro tests of human cadaver femurs, thus helping to develop a suitable mechanistic fracture risk criterion. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:86 / 94
页数:9
相关论文
共 54 条
  • [1] Ajovalasit A., 2010, EPJ WEB C, P6
  • [2] Direct in vivo strain measurements in human bone-A systematic literature review
    Al Nazer, R.
    Lanovaz, J.
    Kawalilak, C.
    Johnston, J. D.
    Kontulainen, S.
    [J]. JOURNAL OF BIOMECHANICS, 2012, 45 (01) : 27 - 40
  • [3] Epidemiology of hip fractures in Okinawa, Japan
    Arakaki, Harumi
    Owan, Ichiro
    Kudoh, Hirohisa
    Horizono, Hidehiro
    Arakaki, Kaoru
    Ikema, Yasunari
    Shinjo, Hirotaka
    Hayashi, Kaori
    Kanaya, Fuminori
    [J]. JOURNAL OF BONE AND MINERAL METABOLISM, 2011, 29 (03) : 309 - 314
  • [4] Pathophysiology of Osteoporosis New Mechanistic Insights
    Armas, Laura A. G.
    Recker, Robert R.
    [J]. ENDOCRINOLOGY AND METABOLISM CLINICS OF NORTH AMERICA, 2012, 41 (03) : 475 - +
  • [5] TEXTURE CORRELATION - A METHOD FOR THE MEASUREMENT OF DETAILED STRAIN DISTRIBUTIONS WITHIN TRABECULAR BONE
    BAY, BK
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 1995, 13 (02) : 258 - 267
  • [6] NUMERICAL-ANALYSIS OF THE REINFORCEMENT EFFECT OF A STRAIN-GAUGE APPLIED TO A SOFT MATERIAL
    BEATTY, MF
    CHEWNING, SW
    [J]. INTERNATIONAL JOURNAL OF ENGINEERING SCIENCE, 1979, 17 (07) : 907 - 915
  • [7] Digital image correlation shows localized deformation bands in inelastic loading of fibrolamellar bone
    Benecke, Gunthard
    Kerschnitzki, Michael
    Fratzl, Peter
    Gupta, Himadri S.
    [J]. JOURNAL OF MATERIALS RESEARCH, 2009, 24 (02) : 421 - 429
  • [8] Falls: Epidemiology, pathophysiology, and relationship to fracture
    Berry S.D.
    Miller R.R.
    [J]. Current Osteoporosis Reports, 2008, 6 (4) : 149 - 154
  • [9] Prediction of strength and strain of the proximal femur by a CT-based finite element method
    Bessho, Masahiko
    Ohnishi, Isao
    Matsuyama, Juntaro
    Matsumoto, Takuya
    Imai, Kazuhiro
    Nakamura, Kozo
    [J]. JOURNAL OF BIOMECHANICS, 2007, 40 (08) : 1745 - 1753
  • [10] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204