Effects of Screw Configuration on the Preload Force of Implant-Abutment Screws

被引:19
|
作者
Zipprich, Holger [1 ]
Rathe, Florian [2 ]
Pinz, Soeren [3 ]
Schlotmann, Luca [3 ]
Lauer, Hans-Christoph [3 ]
Ratka, Christoph [3 ]
机构
[1] Goethe Univ Frankfurt, Dept Prosthodont, Mat Sci Unit, Frankfurt, Germany
[2] DANUBE Private Univ, Dept Prosthodont, Krems An Der Donau, Austria
[3] Goethe Univ Frankfurt, Dept Prosthodont, Theodor Stern Kai 7,House 29, DE-60596 Frankfurt, Germany
关键词
abutment; abutment screw; abutment screw configuration; dental implant; preload force; thread pitch; FINITE-ELEMENT-ANALYSIS; BONE;
D O I
10.11607/jomi.5837
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The aim of this study was to investigate the effects of tightening torque, screw head angle, and thread number on the preload force of abutment screws. Materials and Methods: The test specimens consisted of three self-manufactured components (ie, a thread sleeve serving as an implant analog, an abutment analog, and an abutment screw). The abutment screws were fabricated with metric M1.6 external threads. The thread number varied between one and seven threads. The screw head angles were produced in eight varying angles (30 to 180 degrees). A sensor unit simultaneously measured the preload force of the screw and the torsion moment inside the screw shank. The tightening of the screw with the torque wrench was performed in five steps (15 to 35 Ncm). The torque wrench was calibrated before each step. Results: Only the tightening torque and screw head angle affected the resulting preload force of the implant-abutment connection. The thread number had no effect. There was an approximately linear correlation between tightening torque and preload force. Conclusion: The tightening torque and screw head angle were the only study parameters that affected the resulting preload force of the abutment screw. The results obtained from this experiment are valid only for a single torque condition. Further investigations are needed that analyze other parameters that affect preload force. Once these parameters are known, it will add value for a strong, but detachable connection between the implant and abutment. Short implants and flat-to-flat connections especially will benefit significantly from this knowledge.
引用
收藏
页码:E25 / E32
页数:8
相关论文
共 50 条
  • [31] Removal capability, implant-abutment connection damage and thermal effect using ultrasonic and drilling techniques for the extraction of fractured abutment screws: an in vitro study
    María Bufalá Pérez
    Álvaro Zubizarreta-Macho
    Javier Borrajo Sánchez
    Jorge Hernández Rodríguez
    Jorge Alonso Pérez-Barquero
    Elena Riad Deglow
    Sofía Hernández Montero
    BMC Oral Health, 22
  • [32] Silver coating on dental implant-abutment connection screws as potential strategy to prevent loosening and minimizing bacteria adhesion
    Velasco-Ortega, Eugenio
    Jimenez-Guerra, Alvaro
    Ortiz-Garcia, Ivan
    Nunez-Marquez, Enrique
    Moreno-Munoz, Jesus
    Gil, Javier
    Delgado, Luis M.
    Rondon-Romero, Jose Luis
    Monsalve-Guil, Loreto
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 11
  • [33] Analysis of the Mechanical Behavior and Effect of Cyclic Fatigue on the Implant-Abutment Interface
    German Sandoval, Ramon
    Ortiz Magdaleno, Marine
    Sanchez Robles, Paula
    Zavala Alonso, Norma
    Romo Ramirez, Gabriel Fernando
    ODOVTOS INTERNATIONAL JOURNAL OF DENTAL SCIENCES, 2021, 23 (02) : 104 - 114
  • [34] Implant-abutment interface design affects fatigue and fracture strength of implants
    Steinebrunner, Lars
    Wolfart, Stefan
    Ludwig, Klaus
    Kern, Matthias
    CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (12) : 1276 - 1284
  • [35] Cross-sectional analysis of the implant-abutment interface
    Coelho, A. L.
    Suzuki, M.
    Dibart, S.
    Da Silva, N.
    Coelho, P. G.
    JOURNAL OF ORAL REHABILITATION, 2007, 34 (07): : 508 - 516
  • [36] Does Ferrule Effect Affect Implant-Abutment Stability?
    Mohajerfar, Maryam
    Beyabanaki, Elaheh
    Geramy, Allahyar
    Siadat, Hakimeh
    Alikhasi, Marzieh
    EUROPEAN JOURNAL OF PROSTHODONTICS AND RESTORATIVE DENTISTRY, 2016, 24 (04) : 186 - 191
  • [37] Fatigue induced changes in conical implant-abutment connections
    Blum, Kai
    Wiest, Wolfram
    Fella, Christian
    Balles, Andreas
    Dittmann, Jonas
    Rack, Alexander
    Maier, Dominik
    Thomann, Ralf
    Spies, Benedikt Christopher
    Kohal, Ralf Joachim
    Zabler, Simon
    Nelson, Katja
    DENTAL MATERIALS, 2015, 31 (11) : 1415 - 1426
  • [38] Effects of diameters of implant and abutment screw on stress distribution within dental implant and alveolar bone: A threedimensional finite element analysis
    Shen, Yen -Wen
    Huang, Heng-Li
    Hsu, Jui-Ting
    Fuh, Lih-Jyh
    JOURNAL OF DENTAL SCIENCES, 2024, 19 (02) : 1126 - 1134
  • [39] The behavior of a zirconia or metal abutment on the implant-abutment interface during cyclic loading
    Queiroz, Daher Antonio
    Hagee, Natalia
    Lee, Damian J.
    Zheng, Fengyuan
    JOURNAL OF PROSTHETIC DENTISTRY, 2020, 124 (02) : 211 - 216
  • [40] Load Distribution in Tooth and Implant-Abutment Identical Cases
    Ulusoy, Muhammed Mutahhar
    Arinc, Hakan
    Ozden, Asiye Nehir
    CYPRUS JOURNAL OF MEDICAL SCIENCES, 2020, 5 (03): : 211 - 216