Stereotactic Vacuum-Assisted Biopsies on a Digital Breast 3D-Tomosynthesis System

被引:38
|
作者
Viala, Juliette [1 ]
Gignier, Pierre [1 ]
Perret, Baudouin [1 ]
Hovasse, Claudie [1 ]
Hovasse, Denis [1 ]
Chancelier-Galan, Marie-Dominique [1 ]
Bornet, Gregoire [1 ]
Hamrouni, Adel [1 ]
Lasry, Jean-Louis [1 ]
Convard, Jean-Paul [1 ]
机构
[1] Hop Prive Antony, Dept Radiol, F-92160 Antony, France
来源
BREAST JOURNAL | 2013年 / 19卷 / 01期
关键词
breast biopsy; digital breast tomosynthesis; TOMOSYNTHESIS; MAMMOGRAPHY;
D O I
10.1111/tbj.12044
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The purpose of this study was to describe our operating process and to report results of 118 stereotactic vacuum-assisted biopsies performed on a digital breast 3D-tomosynthesis system. From October 2009 to December 2010, 118 stereotactic vacuum assisted biopsies have been performed on a digital breast 3D-tomosynthesis system. Informed consent was obtained for all patients. A total of 106 patients had a lesion, six had two lesions. Sixty-one lesions were clusters of micro-calcifications, 54 were masses and three were architectural distortions. Patients were in lateral decubitus position to allow shortest skin-target approach (or sitting). Specific compression paddle, adapted on the system, performed, and graduated, allowing localization in X-Y. Tomosynthesis views define the depth of lesion. Graduated Coaxial localization kit determines the beginning of the biopsy window. Biopsies were performed with an ATEC-Suros, 9 Gauge handpiece. All biopsies, except one, have reached the lesions. Five hemorrhages were incurred in the process, but no interruption was needed. Eight breast hematomas, were all spontaneously resolved. One was an infection. About 40% of patients had a skin ecchymosis. Processing is fast, easy, and requires lower irradiation dose than with classical stereotactic biopsies. Histology analysis reported 45 benign clusters of micro-calcifications, 16 malignant clusters of micro-calcifications, 24 benign masses, and 33 malignant masses. Of 13 malignant lesions, digital 2D-mammography failed to detect eight lesions and underestimated the classification of five lesions. Digital breast 3D-tomosynthesis depicts malignant lesions not visualized on digital 2D-mammography. Development of tomosynthesis biopsy unit integrated to stereotactic system will permit histology analysis for suspicious lesions.
引用
收藏
页码:4 / 9
页数:6
相关论文
共 50 条
  • [21] Stereotactic vacuum-assisted breast biopsy under lateral decubitus position
    Lee, Sang Hyup
    Jung, Youn Joo
    Jung, Hyuk Jae
    Kim, Jee Yeon
    Choo, Ki Seok
    Nam, Kyung Jin
    Kim, Hyun Yul
    ANNALS OF SURGICAL TREATMENT AND RESEARCH, 2016, 90 (01) : 16 - 20
  • [22] Upright stereotactic vacuum-assisted needle biopsy of suspicious breast microcalcifications
    Sim, L. S. J.
    Kei, P. L.
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2008, 52 (04) : 358 - 364
  • [23] Tomosynthesis-guided Vacuum-assisted Breast Biopsy of Sonographically Occult Non-calcified Breast Lesions Detected on Tomosynthesis
    Fung, W. Y.
    Fung, E. P. Y.
    Kwok, K. M.
    Chan, S. K.
    Wong, L. K. M.
    Mak, W. S.
    Cho, D. H. Y.
    HONG KONG JOURNAL OF RADIOLOGY, 2021, 24 (04): : 287 - 294
  • [24] Stereotactic vacuum-assisted breast biopsy in 268 nonpalpable lesions [Biopsia mammaria in stereotassi, vacuum-assisted, in 268 lesioni non palpabili]
    Tonegutti M.
    Girardi V.
    La radiologia medica, 2008, 113 (1) : 65 - 75
  • [25] Ultrasonography as a Guiding Method in Breast Micro-Calcification Vacuum-Assisted Biopsies
    Keranen, A. K.
    Haapea, M.
    Rissanen, T.
    ULTRASCHALL IN DER MEDIZIN, 2016, 37 (05): : 497 - 502
  • [26] Low-cost phantoms for training of stereotactic vacuum-assisted biopsy of the breast
    Fischmann, Arne
    Siegmann, Katja C.
    CLINICAL IMAGING, 2010, 34 (02) : 97 - 99
  • [27] Immediate Migration of Biopsy Clip Markers After Upright Digital Breast Tomosynthesis-Guided Vacuum-Assisted Core Biopsy
    Teichgraeber, Davis C.
    Martaindale, Sarah
    Omofoye, Toma S.
    Hess, Kenneth R.
    Parikh, Jay R.
    Whitman, Gary J.
    ACADEMIC RADIOLOGY, 2020, 27 (02) : 204 - 209
  • [28] Accuracy and outcomes of stereotactic vacuum-assisted breast biopsy for diagnosis and management of nonpalpable breast lesions
    Tsai, Huei-Yi
    Chao, Min-Fang
    Fu Ou-Yang
    Kan, Jung-Yu
    Hsu, Jui-Sheng
    Hou, Ming-Feng
    Chiu, Herng-Chia
    KAOHSIUNG JOURNAL OF MEDICAL SCIENCES, 2019, 35 (10): : 640 - 645
  • [29] Vacuum-assisted breast biopsy for breast cancer
    Park, Hai-Lin
    Hong, Jisun
    GLAND SURGERY, 2014, 3 (02) : 120 - 127
  • [30] Vacuum-assisted breast biopsies - Experience at the Antoine Lacassagne Cancer Center (Nice, France)
    Chapellier, C
    Balu-Maestro, C
    Amoretti, N
    Chauvel, C
    Ben-Taarit, I
    Birtwisle-Peyrottes, I
    CLINICAL IMAGING, 2006, 30 (02) : 99 - 107