Head-to-head comparison between flash and continuous glucose monitoring systems in outpatients with type 1 diabetes

被引:80
作者
Bonora, B. [1 ]
Maran, A. [1 ]
Ciciliot, S. [1 ]
Avogaro, A. [1 ]
Fadini, G. P. [1 ]
机构
[1] Univ Padua, Div Metab Dis, Dept Med, Via Giustiniani 2, I-35128 Padua, Italy
来源
JOURNAL OF ENDOCRINOLOGICAL INVESTIGATION | 2016年 / 39卷 / 12期
关键词
Sensors; Continuous glucose monitoring; Hypoglycaemia; Variability; Calibration; BLOOD-GLUCOSE; VARIABILITY; INSTABILITY; ADOLESCENTS; FREQUENCY; ACCURACY; CHILDREN; INDEX;
D O I
10.1007/s40618-016-0495-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Continuous glucose monitoring (CGM) is being increasingly used in clinical practice. The flash glucose monitoring (FGM) and CGM are different systems of interstitial glucose recording. We aimed to determine the agreement between the factory-calibrated FGM FreeStyle Libre (FSL) and the gold-standard CGM Dexcom G4 Platinum (DG4P). We analyzed data from n = 8 outpatients with type 1 diabetes, who wore the FSL and DG4P for up to 14 days during their habitual life. We aligned FSL and DG4P recordings to obtain paired glucose measures. We calculated correlation coefficients, mean absolute relative difference (MARD), percentages in Clarke error grid areas, time spent in hyperglycaemia, target glycaemia, or hypoglycaemia, as well as glucose variability with both sensors. Comparison with self-monitoring of blood glucose (SMBG) was also performed. Patients varied in terms of age, diabetes duration, and HbA1c (from 5.9 to 9.6 %). In the pooled analysis of 10,020 paired values, there was a good correlation between FSL and DG4P (r (2) = 0.76; MARD = 18.1 +/- 14.8 %) with wide variability among patients. The MARD was significantly higher during days 11-14 than in days 1-10, and during hypoglycaemia (19 %), than in normoglycaemia (16 %) or hyperglycaemia (13 %). Average glucose profiles and MARD versus SMBG were similar between the two sensors. Time spent in normo-, hyper-, or hypoglycaemia, and indexes of glucose variability was similarly estimated by the two sensors. In outpatients with type 1 diabetes, we found good agreement between the FSL and DG4P. No significant difference was detected in the estimation of clinical diagnostic parameters.
引用
收藏
页码:1391 / 1399
页数:9
相关论文
共 24 条
[1]  
[Anonymous], 2014, J DIABETES SCI TECHN
[2]   GLYCEMIC EFFECTS OF SGLT-2 INHIBITOR CANAGLIFLOZIN IN TYPE 1 DIABETES PATIENTS USING THE DEXCOM G4 PLATINUM CGM [J].
Argento, Nicholas B. ;
Nakamura, Katherine .
ENDOCRINE PRACTICE, 2016, 22 (03) :315-322
[3]   The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System [J].
Bailey, Timothy ;
Bode, Bruce W. ;
Christiansen, Mark P. ;
Klaff, Leslie J. ;
Alva, Shridhara .
DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (11) :787-794
[4]   Limitations of conventional methods of self-monitoring of blood glucose - Lessons learned from 3 days of continuous glucose sensing in pediatric patients with type 1 diabetes [J].
Boland, E ;
Monsod, T ;
Delucia, M ;
Brandt, CA ;
Fernando, S ;
Tamborlane, WV .
DIABETES CARE, 2001, 24 (11) :1858-1862
[5]   KERNEL DENSITY ESTIMATION VIA DIFFUSION [J].
Botev, Z. I. ;
Grotowski, J. F. ;
Kroese, D. P. .
ANNALS OF STATISTICS, 2010, 38 (05) :2916-2957
[6]   A New-Generation Continuous Glucose Monitoring System: Improved Accuracy and Reliability Compared with a Previous-Generation System [J].
Christiansen, Mark ;
Bailey, Timothy ;
Watkins, Elaine ;
Liljenquist, David ;
Price, David ;
Nakamura, Katherine ;
Boock, Robert ;
Peyser, Thomas .
DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (10) :881-888
[7]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[8]  
Feldman Ben, 2003, Diabetes Technol Ther, V5, P769, DOI 10.1089/152091503322526978
[9]  
Heinemann Lutz, 2015, J Diabetes Sci Technol, V9, P947, DOI 10.1177/1932296815603528
[10]   Glucose variability is associated with intensive care unit mortality [J].
Hermanides, Jeroen ;
Vriesendorp, Titia M. ;
Bosman, Robert J. ;
Zandstra, Durk F. ;
Hoekstra, Joost B. ;
DeVries, J. Hans .
CRITICAL CARE MEDICINE, 2010, 38 (03) :838-842