An increased thermal index can be achieved when performing Doppler studies in obstetric sonography

被引:26
作者
Sheiner, Eyal
Shoham-Vardi, Flana
Pombar, Xavier
Hussey, Michael J.
Strassner, Howard T.
Abramowicz, Jacques S.
机构
[1] Rush Univ, Med Ctr, Dept Obstet & Gynecol, Chicago, IL 60612 USA
[2] Soroka Univ Med Ctr, Dept Obstet & Gynecol, Beer Sheva, Israel
[3] Ben Gurion Univ Negev, Fac Hlth Sci, Beer Sheva, Israel
[4] Ben Gurion Univ Negev, Epidemiol & Hlth Serv Evaluat Dept, Beer Sheva, Israel
关键词
acoustic output; Doppler sonography; mechanical index; safety; thermal index; ultrasound;
D O I
10.7863/jum.2007.26.1.71
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective The present study was aimed at evaluating acoustic outputs during clinical ultrasound examinations, as expressed by the thermal index (TI) and the mechanical index (MI), during the second half of pregnancy and comparing acoustic outputs between B-mode and Doppler examinations. Methods. Patients with suspected fetal growth problems undergoing Doppler studies of the fetal circulation in addition to B-mode sonography were selected. Examinations took place between 21 and 40 weeks' gestation. An obstetrician collected data prospectively Sonographers were unaware of the data being sought. The analysis of variance test was applied for differences in continuous variables. Results. A total of 63 examinations were evaluated. The mean gestational age +/- SD was 31.6 +/- 5.1 weeks, The mean duration of the total examinations was 17.6 +/- 8.6 minutes, whereas the Doppler studies lasted 0.9 +/- 0.8 minutes. The TI was significantly higher in the pulsed wave Doppler studies (mean, 1.5 +/- 0.5; range, 0.9-2.8) and color flow imaging studies (mean, 0.8 +/- 0.1; range, 0.6-1.2) compared with B-mode sonography (mean, 0.3 +/- 0.1; range, 0.1-0.7; P < .01). During the examination, 190 B-mode MI variations were recorded (mean, 1.1 +/- 0.1), which were comparable with those of the 31 color flow Doppler studies (mean, 1.0 +/- 0.1; P = .09) but higher than the 190 pulsed wave Doppler MI variations (mean 0.9 +/- 0.2; P < .001). Conclusions. increased acoustic output levels, as expressed by TI levels, are reached during obstetric Doppler studies. In particular, TI levels may reach 1.5 and higher. Doppler procedures should be performed with caution and be as brief as possible during obstetric sonography.
引用
收藏
页码:71 / 76
页数:6
相关论文
共 29 条
[1]   Ultrasound in obstetrics and gynecology - Is this hot technology too hot? [J].
Abramowicz, JS .
JOURNAL OF ULTRASOUND IN MEDICINE, 2002, 21 (12) :1327-1333
[2]   Literature review by the ISUOG Bioeffects and Safety Committee [J].
Abramowicz, JS ;
Kossoff, G ;
Marsal, K ;
Ter Haar, G .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2002, 19 (03) :318-319
[3]  
*AM I ULTR MED, 2000, J ULTRAS MED, V19, P149
[4]  
American Institute of Ultrasound in Medicine, 1994, MED ULTR SAF
[5]  
Barnett S B, 2001, J Matern Fetal Med, V10, P75, DOI 10.1080/714904312
[6]   Intracranial temperature elevation from diagnostic ultrasound [J].
Barnett, SB .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (07) :883-888
[7]   IS DIAGNOSTIC ULTRASOUND SAFE - CURRENT INTERNATIONAL CONSENSUS ON THE THERMAL MECHANISM [J].
BARNETT, SB ;
KOSSOFF, G ;
EDWARDS, MJ .
MEDICAL JOURNAL OF AUSTRALIA, 1994, 160 (01) :33-37
[8]   The sensitivity of biological tissue to ultrasound [J].
Barnett, SB ;
Rott, HD ;
terHaar, GR ;
Ziskin, MC ;
Maeda, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (06) :805-812
[9]   First-trimester screening for Down syndrome with ductus venosus Doppler studies in addition to nuchal translucency and serum markers [J].
Borrell, A ;
Gonce, A ;
Martinez, JM ;
Borobio, V ;
Fortuny, A ;
Coll, O ;
Cuckle, H .
PRENATAL DIAGNOSIS, 2005, 25 (10) :901-905
[10]   HEAT GENERATED BY ULTRASOUND IN AN ABSORBING MEDIUM [J].
CAVICCHI, TJ ;
OBRIEN, WD .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1984, 76 (04) :1244-1245