Comparison between high performance liquid chromatography and capillary zone electrophoresis for the diagnosis of congenital disorders of glycosylation

被引:21
|
作者
Quintana, Ester [2 ,3 ]
Montero, Raquel [1 ,3 ]
Casado, Mercedes [1 ]
Navarro-Sastre, Aleix [2 ,3 ]
Vilaseca, Maria A. [1 ,3 ]
Briones, Paz [2 ,3 ,4 ]
Artuch, Rafael [1 ,3 ]
机构
[1] Hosp St Joan Deu, Dept Clin Biochem, Barcelona 08950, Spain
[2] Hosp Clin Barcelona, Inst Bioquim Clin, Serv Bioquim & Genet Mol, Barcelona, Spain
[3] ISCIII, CIBERER, Barcelona, Spain
[4] Consejo Super Invest Cient, Barcelona, Spain
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2009年 / 877卷 / 24期
关键词
Congenital disorders of glycosylation; High performance liquid chromatography; Capillary zone electrophoresis; Isoelectric focusing; Paediatric patients; CARBOHYDRATE-DEFICIENT TRANSFERRIN; SERUM TRANSFERRIN; HPLC;
D O I
10.1016/j.jchromb.2009.06.031
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Transferrin isoelectric focusing (IEF) is the most widely used method to screen for congenital disorders of glycosylation (CDG). Our aim was to compare high performance liquid chromatography (HPLC) and capillary zone electrophoresis (CZE) procedures for serum sialotransferrin analysis. 58 serum samples were processed both by CZE and HPLC: 35 were from paediatric controls, 18 from patients with an altered sialotransferrin IEF pattern and 5 were transferrin variant samples. HPLC analysis was performed with an anion-exchange column with spectrophotometric detection at 470 nm. CZE analysis was done using the commercial CEofix-CDT kit with spectrophotometric detection at 200nm. Passing-Bablok regression analysis showed good agreement for tri-. tetra- and penta-sialotransferrin by both procedures. But for disialotransferrin, higher values were observed by the HPLC procedure. The HPLC and CZE methods allowed reproducible separation and analysis of single transferrin glycoforms with similar peak patterns. All patients presented values outside the range established in our control population either by HPLC or by CZE, even in patients with moderate forms of COG that had been difficult to detect by IEF. In conclusion, measurement of sialotransferrin isoforms and interpretation using method-specific reference values may offer some advantages for the diagnosis of CDG as compared with the standard IEF procedure. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:2513 / 2518
页数:6
相关论文
共 50 条
  • [21] Comparison of analyses of surfactants in cosmetics using high-performance liquid chromatography and high-performance capillary electrophoresis
    Lin, WC
    Lin, ST
    Shu, SL
    JOURNAL OF SURFACTANTS AND DETERGENTS, 2000, 3 (01) : 67 - 72
  • [22] Comparative study between capillary electrophoresis and high performance liquid chromatography in 'guarana" based phytopharmaceuticals
    Sombra, LL
    Gómez, MR
    Olsina, R
    Martínez, LD
    Silva, MF
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2005, 36 (05) : 989 - 994
  • [23] Application of calixarenes for high performance liquid chromatography, capillary electrophoresis and electrochromatography
    Li, LS
    Da, SL
    Feng, YQ
    Min, L
    PROGRESS IN CHEMISTRY, 2005, 17 (03) : 523 - 530
  • [24] A survey of the potential of the high performance liquid chromatography and capillary zone electrophoresis for the determination of platinum and platinum group metals
    Sommer, L
    Vlasánková, R
    CHROMATOGRAPHIA, 2000, 52 (11-12) : 692 - 702
  • [25] Comparison of the separation of nine tryptamine standards based on gas chromatography, high performance liquid chromatography and capillary electrophoresis methods
    Wang, Man-Juing
    Liu, Ju-Tsung
    Chen, Hung-Ming
    Lin, Jian-Jhih
    Lin, Cheng-Huang
    JOURNAL OF CHROMATOGRAPHY A, 2008, 1181 (1-2) : 131 - 136
  • [26] Analysis of phenolic compounds in Spanish Albrarino and Portuguese Alvarinho and Loureiro wines by capillary zone electrophoresis and high-performance liquid chromatography
    Andrade, PB
    Oliveira, BM
    Seabra, RM
    Ferreira, MA
    Ferreres, F
    García-Viguera, C
    ELECTROPHORESIS, 2001, 22 (08) : 1568 - 1572
  • [27] Development and validation of capillary zone electrophoresis and high-performance liquid chromatography methods for the determination of oral anticoagulant edoxaban in pharmaceutical tablets
    Oliveira Rocha, Manoelly
    Bordignon Primieri, Gabriele
    Ribeiro Wingert, Nathalie
    Arguello Da Silva, Jacqueline
    Steppe, Martin
    ELECTROPHORESIS, 2022, 43 (15) : 1617 - 1625
  • [28] Comparison between transient isotachophoretic capillary zone electrophoresis and reversed-phase liquid chromatography for the determination of peptides in plasma
    Waterval, JCM
    Krabbe, H
    Teeuwsen, J
    Bult, A
    Lingeman, H
    Underberg, WJM
    ELECTROPHORESIS, 1999, 20 (14) : 2909 - 2916
  • [29] Comparative evaluation between capillary electrophoresis and high-performance liquid chromatography for the analysis of florfenicol in plasma
    Kowalski, P
    Konieczna, L
    Chmielewska, A
    Oledzka, I
    Plenis, A
    Bieniecki, M
    Lamparczyk, H
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2005, 39 (05) : 983 - 989
  • [30] Determination of Xanthii constituents by high-performance liquid chromatography and capillary electrophoresis
    Sheu, SJ
    Hsu, FL
    Tai, HM
    Sheu, MJ
    Huang, MH
    JOURNAL OF FOOD AND DRUG ANALYSIS, 2003, 11 (01) : 67 - 71