Catastrophic failure of femoral stem modular junction when combined with metal-on-metal bearing in comparison to ceramic-on-ceramic: A retrospective cohort study

被引:6
作者
Beauchamp, Jean-Etienne [1 ]
Vendittoli, Pascal-Andre [2 ]
Barry, Janie [2 ]
Pelet, Stephane [1 ,3 ]
Belzile, Etienne L. [1 ,3 ]
机构
[1] Univ Laval G1V 0A6, Dept Surg, Fac Med, Div Orthopaed Surg, Quebec City, PQ G1V 0A6, Canada
[2] Univ Montreal, Dept Surg, CIUSSS LEst LIle Montreal, Hop Maisonneuve Rosemont H1T 2M4, 5415 Blvd LAssompt, Montreal, PQ, Canada
[3] Univ Laval G1J IZ4, Div Orthopaed Surg, CHU Quebec, 1401,18e Rue, Quebec City, PQ, Canada
基金
加拿大健康研究院;
关键词
Total Hip Arthroplasty; Metal-on-metal; Modular neck femoral stems; Metallosis; Adverse reaction to metal debris; TOTAL HIP-ARTHROPLASTY; CHROMIUM SERUM-LEVELS; LARGE-DIAMETER METAL; ION LEVELS; NECKS ANALYSIS; CORROSION; HEAD; COBALT; INTERFACE; RELEASE;
D O I
10.1016/j.otsr.2020.102749
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Metal ion release from total hip arthroplasty's (THA) metal-on-metal (MoM) bearing surfaces or head-neck modular junction (trunnionosis) has been identified as a major cause of adverse reaction to metal debris (ARMD). No study has compared the effects of these bearing couples when combined with modular neck femoral stems (MNFS) (i.e. did a modular CoCr have a higher effect than large MoM bearings on whole blood ion concentrations and ARMD). Therefore we did a retrospective comparative analysis of prospectively collected data aiming to: (1) assess the difference in metal ion release between a group of MoM and CoC bearings implanted with a non-cemented MNFS; (2) compare the ARMD and the related revision rates between the two bearing types; (3) compare the patients' HOOS and Harris Hip scores. Hypothesis: Metal ion levels and complications rate will be higher in MoM group. Methods: Thirty hips received the same short MNFS with modular CoCr. Seventeen had CoC bearing (16 = 36 mm) and 13 large diameter head (LDH) MoM bearing (mean = 50 mm, 44-54 mm) Both groups had whole blood Cobalt (Co) and Chromium (Cr) concentrations measurements at last follow up or before revision surgery (if any) and were compared to 41 control individuals without implants. ARMD, revision rates and other complications were tensed. Functional outcome differences between operative groups were assessed with HOOS and Harris Hip scores. Results: Mean whole blood Co and Cr were 1.9 mu g/L (0.1-10.6 min-max) and 1.2 mu g/L (0.2-4.6 min-max) in the CoC group and 12.7 mu g/L (2.5-40.7 min-max) and 8.2 mu g/L (2.8-40.6 min-max) for MoM group. Controls presented whole blood Co and Cr levels of 0.2 mu g/L (0.04-0.2 min-max) and 0.6 mu g/L (0.1-0.7 min-max) respectively. CoC had significantly higher whole blood metal ion concentrations than controls, while MoM had significantly higher concentrations than both (p < 0.001). MoM had a 9/13 (69%) revision rate due to ARMD (with stem-neck junction corrosion and wear were present in all cases) while the CoC had none. Functional scores at last follow up did not statistically differ between groups (Harris: CoC = 94.1 +/- 8.4 vs MoM 91.8 +/- 14.8 (p = 0.22) and CoC HOOS 90.2 +/- 10.7 vs 75.6 +/- 21.7 (p = 0.08)). Conclusion: In our study, MNFS with CoCr modular neck released metal ion systemically, as shown in CoC THAs, but when combined with MoM LDH bearings, that elevation was significantly higher. In MNFS MoM, these high metal ion levels translated into a dramatic ARMD related revision rate where stem-neck junction corrosion and wear were present in all cases. More research is required to understand the effects of MoM bearing on neck-stem junctions, and its mechanisms of corrosion. (C) 2020 Elsevier Masson SAS. All rights reserved.
引用
收藏
页数:6
相关论文
共 33 条
[1]  
Afolaranmi G A, 2008, Open Orthop J, V2, P10, DOI 10.2174/1874325000802010010
[2]  
Barry Janie, 2017, Open Orthop J, V11, P1337, DOI 10.2174/1874325001711011337
[3]   Pseudotumour Formation Due to Tribocorrosion at the Taper Interface of Large Diameter Metal on Polymer Modular Total Hip Replacements [J].
Cook, Richard B. ;
Bolland, Benjamin J. R. F. ;
Wharton, Julian A. ;
Tilley, Simon ;
Latham, Jeremy M. ;
Wood, Robert J. K. .
JOURNAL OF ARTHROPLASTY, 2013, 28 (08) :1430-1436
[4]   Metal-on-metal bearings total hip arthroplasty: The cobalt and chromium ions release concern [J].
Delaunay, C. ;
Petit, I. ;
Learmonth, I. D. ;
Oger, P. ;
Vendittoli, P. A. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2010, 96 (08) :894-904
[5]   Modular versus Nonmodular Neck Femoral Implants in Primary Total Hip Arthroplasty: Which is Better? [J].
Duwelius, Paul J. ;
Burkhart, Bob ;
Carnahan, Clay ;
Branam, Grant ;
Ko, Laura Matsen ;
Wu, YingXing ;
Froemke, Cecily ;
Wang, Lian ;
Grunkemeier, Gary .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (04) :1240-1245
[6]   No radiological and biological sign of trunnionosis with Large Diameter Head Ceramic Bearing Total Hip Arthroplasty after 5 years [J].
Eichler, David ;
Barry, Janie ;
Lavigne, Martin ;
Masse, Vincent ;
Vendittoli, Pascal-Andre .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2021, 107 (01)
[7]  
Ellman MB, 2013, J ARTHROPLASTY, V28, P196, DOI DOI 10.1016/J.ARTH.2011.05.024
[8]  
Ellman MB, 2013, J ARTHROPLASTY, V28, pE1
[9]  
Federal Drug Administration, 2016, CLASS 2 DEV REC MOD
[10]   Adverse tissue reaction to corrosion at the neck-stem junction after modular primary total hip arthroplasty [J].
Gkagkalis, G. ;
Mettraux, P. ;
Omoumi, P. ;
Mischler, S. ;
Ruediger, H. A. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2015, 101 (01) :123-126