Capabilities of laboratories to determine melamine in food-results of an international proficiency test

被引:15
作者
Breidbach, Andreas [1 ]
Bouten, Katrien [1 ]
Kroeger, Katy [1 ]
Ulberth, Franz [1 ]
机构
[1] Commiss European Communities, Joint Res Ctr, Inst Reference Mat & Measurements, B-2440 Geel, Belgium
关键词
Melamine; Proficiency test; Isotope dilution mass spectrometry; LIQUID-CHROMATOGRAPHY; PET FOOD; ADULTERATION; PRODUCTS; TISSUE; SALMON;
D O I
10.1007/s00216-009-3198-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A proficiency test to assess the capabilities of laboratories to determine melamine in a milk powder and a baking mix, representing starch-containing foods like bread and biscuits, was carried out in January 2009. The need for such an interlaboratory comparison arose from a health scare in China about melamine-tainted powdered milk in the second half of 2008. Laboratories in 31 countries, including Australia, China, India, Japan, New Zealand and the USA, and 21 of the 27 Member States of the European Union participated and reported back 114 results for the milk powder and 112 for the baking mix test materials. The reported results were compared to reference values determined by exact-matching double isotope dilution mass spectrometry. The so-determined assigned values were 10.0 +/- 0.6 mg/kg melamine in the milk powder and 3.18 +/- 0.17 mg/kg melamine in the baking mix. A coverage factor k of 2 was applied to calculate the expanded uncertainties. Three quarters of all reported results for both materials had associated z scores which were satisfactory (z a parts per thousand currency signaEuro parts per thousand|2|). Of the reported results, 90% was accompanied by a measurement uncertainty statement, and the majority of the measurement uncertainty ranges were reasonable. A number of laboratories were found to underestimate their measurement uncertainties. Methods that involved the use of stable-isotope-labelled melamine were shown to be clearly advantageous with regard to the accuracy of the results. However, no significant influence by other method parameters could be identified.
引用
收藏
页码:503 / 510
页数:8
相关论文
共 24 条
  • [1] Determination and confirmation of melamine residues in catfish, trout, tilapia, salmon, and shrimp by liquid chromatography with tandem mass spectrometry
    Andersen, Wendy C.
    Turnipseed, Sherri B.
    Karbiwnyk, Christine M.
    Clark, Susan B.
    Madson, Mark R.
    Gieseker, Charles A.
    Miller, Ron A.
    Rummel, Nathan G.
    Reimschuessel, Renate
    [J]. JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2008, 56 (12) : 4340 - 4347
  • [3] [Anonymous], 2009, MEDICAL TOURISM ILLI, P1
  • [4] Burns K, 2007, JAVMA-J AM VET MED A, V230, P1285
  • [5] Public-health risks of melamine in milk products
    Chan, E. Y. Y.
    Griffiths, S. M.
    Chan, C. W.
    [J]. LANCET, 2008, 372 (9648) : 1444 - 1445
  • [6] *COMM EUR COMM, 2008, OFFICIAL J EUR UNION, V273, P18
  • [7] High-performance liquid chromatographic method for the simultaneous detection of the adulteration of cereal flours with melamine and related triazine by-products ammeline, ammelide, and cyanuric acid
    Ehling, S.
    Tefera, S.
    Ho, I. P.
    [J]. FOOD ADDITIVES AND CONTAMINANTS, 2007, 24 (12): : 1319 - 1325
  • [8] The determination of melamine in muscle tissue by liquid chromatography/tandem mass spectrometry
    Filigenzi, Michael S.
    Tor, Elizabeth R.
    Poppenga, Robert H.
    Aston, Linda A.
    Puschner, Birgit
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (24) : 4027 - 4032
  • [9] Detection of melamine using commercial enzyme-linked immunosorbent assay technology
    Garber, Eric A. E.
    [J]. JOURNAL OF FOOD PROTECTION, 2008, 71 (03) : 590 - 594
  • [10] High accuracy determination of malachite green and leucomalachite green in salmon tissue by exact matching isotope dilution mass spectrometry
    Hall, Zoe
    Hopley, Chris
    O'Connor, Gavin