Absolute x-ray calibration of a gated x-ray framing camera for the Laser MegaJoule facility in the 0.1 keV-1 keV spectral range

被引:2
|
作者
Hubert, S. [1 ]
Boubault, F. [1 ]
机构
[1] CEA CESTA, F-33116 Le Barp, France
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2021年 / 92卷 / 03期
关键词
Plasma diagnostics - Laser diagnostics - Signal to noise ratio - Cameras;
D O I
10.1063/5.0004105
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
X-ray framing cameras (XRFCs) are routinely used at the Laser MegaJoule facility in x-ray imaging plasma diagnostics around the target chamber. Most of these diagnostics are based on multilayer x-ray toroidal mirrors under grazing incidence. The absolute calibration of the XRFCs is expressly expected both to optimize the signal-to-noise ratio for the dynamic range for specific experiments and to quantitatively process the data. The purpose of this paper is to describe our technique to routinely calibrate these instruments in the sub-keV spectral range. The calibration presented in this work was carried out using the XRFC enclosed in a sealed "airbox" structure. This calibration relies on a Manson source recently upgraded to operate at high emission current (5 mA) with 10 kV accelerating voltage to work with a 1-m grazing-incidence Rowland circle monochromator. The framing camera sensitivity was absolutely determined over the 0.1-1.2 spectral range with an average uncertainty of 2.4% rms while operating in DC mode. Finally, we compare the results with a synchrotron source calibration previously obtained and a theoretical model.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Absolute calibration of the spectral sensitivity of an x-ray streak camera over the 0.1-10 keV spectral range equipped with CsI photocathode
    Hubert, S.
    Boubault, F.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2023, 94 (11):
  • [2] Overview of the ARGOS X-ray framing camera for Laser MegaJoule
    Trosseille, C.
    Aubert, D.
    Auger, L.
    Bazzoli, S.
    Beck, T.
    Brunel, P.
    Burillo, M.
    Chollet, C.
    Gazave, J.
    Jasmin, S.
    Maruenda, P.
    Moreau, I.
    Oudot, G.
    Raimbourg, J.
    Soullie, G.
    Stemmler, P.
    Zuber, C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (11):
  • [3] X-ray calibration facility for plasma diagnostics of the MegaJoule laser
    Hubert, S.
    Prevot, V.
    IFSA 2011 - SEVENTH INTERNATIONAL CONFERENCE ON INERTIAL FUSION SCIENCES AND APPLICATIONS, 2013, 59
  • [4] SPECTRUM OF COSMIC X-RAY AND GAMMA-RAY BACKGROUND IN ENERGY RANGE 1 KEV-1 MEV
    KASTURIRANGAN, K
    RAO, UR
    ASTROPHYSICS AND SPACE SCIENCE, 1972, 15 (01) : 161 - +
  • [5] The x-ray calibration facility of the laser integration line in the 0.9-10 keV range: The high energy x-ray source and some applications
    Hubert, S.
    Dubois, J. L.
    Gontier, D.
    Lidove, G.
    Reverdin, C.
    Soullie, G.
    Stemmler, P.
    Villette, B.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (05):
  • [6] The MPE X-ray test facility PANTER: Calibration of hard X-ray (15–50 kev) optics
    M. J. Freyberg
    H. Bräuninger
    W. Burkert
    G. D. Hartner
    O. Citterio
    F. Mazzoleni
    G. Pareschi
    D. Spiga
    S. Romaine
    P. Gorenstein
    B. D. Ramsey
    Experimental Astronomy, 2005, 20 : 405 - 412
  • [7] First set of gated x-ray imaging diagnostics for the Laser Megajoule facility
    Rosch, R.
    Trosseille, C.
    Caillaud, T.
    Allouche, V.
    Bourgade, J. L.
    Briat, M.
    Brunel, P.
    Burillo, M.
    Casner, A.
    Depierreux, S.
    Gontier, D.
    Jadaud, J. P.
    Le Breton, J. P.
    Llavador, P.
    Loupias, B.
    Miquel, J. L.
    Oudot, G.
    Perez, S.
    Raimbourg, J.
    Rousseau, A.
    Rousseaux, C.
    Rubbelynck, C.
    Stemmler, P.
    Troussel, P.
    Ulmer, J. L.
    Wrobel, R.
    Beauvais, P.
    Pallet, M.
    Prevot, V.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (03):
  • [8] The MPE X-ray test facility PANTER:: Calibration of hard X-ray (15-50 kev) optics
    Freyberg, M. J.
    Braeuninger, H.
    Burkert, W.
    Hartner, G. D.
    Citterio, O.
    Mazzoleni, F.
    Pareschi, G.
    Spiga, D.
    Romaine, S.
    Gorenstein, P.
    Ramsey, B. D.
    EXPERIMENTAL ASTRONOMY, 2005, 20 (1-3) : 407 - 414
  • [9] Performance of Laser Megajoule's x-ray streak camera
    Zuber, C.
    Bazzoli, S.
    Brunel, P.
    Burillo, M.
    Fronty, J. P.
    Gontier, D.
    Goulmy, C.
    Moreau, I.
    Oudot, G.
    Rubbelynck, C.
    Soullie, G.
    Stemmler, P.
    Trosseille, C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (11):
  • [10] Instrumental aspects of x-ray microbeams in the range above 1 keV
    Dhez, P
    Chevallier, P
    Lucatorto, TB
    Tarrio, C
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (04): : 1907 - 1920